Agilent Technologies 571 Bayard-Alpert Gauge Tube
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 571 Bayard-Alpert Gauge Tube

Agilent Technologies 571 Bayard-Alpert Gauge Tube — Vacuum technologies with Max pressure 1.33×10⁻⁴ megapascal.

The 571 is a Bayard-Alpert ionization gauge tube with a yttria-coated iridium filament, covering 1×10⁻³ to 2×10⁻¹⁰ Torr. It is available in glass, Kovar, CF, and KF flange variants, all sharing the same electrical ratings. It serves as the high-vacuum sensor in systems where a controller provides emission, degassing, and ground integrity.

Max pressure
1.33×10⁻⁴ megapascal

Agilent Technologies Vacuum Transducers Series — Model Comparison & Selection

Model Measurement rangeAtomic mass unit, gramsElectron charge, coulombMolar volume, litersMax pressure, megapascalSensitivity
571 Bayard-Alpert Gauge Tube Current -3...21.66160222.411.33×10⁻⁴17
531 Thermocouple Gauge, Rough Vacuum Pressure Measurement —1.66160222.41——
572 Tungsten Bayard-Alpert Gauge -3...21.66160222.41-3...2—
Cold Cathode Gauge Tube 525, High Vacuum (HV) Gauge —1.66160222.41——
Hot Filament Gauge MBA2-201, Ionization BA Gauge —1.66160222.41——
IMG-100 Inverted Magnetron Gauge —1.66160222.41——
IMG-300 UHV Inverted Magnetron Gauge —1.66160222.41——
UHV-24/UHV-24p 11.66160222.41—25

Full specifications

Core Performance & Functional Specifications
Max pressure 1 Torr
Measurement range -3...2 Torr
Sensitivity 17
Degassing 21
Electron Charge 1.602 coulomb
Internal Volume 220 cc
Maximum operating pressure 1 Torr
Molar Volume 22.41 liters
Operation 17
Pumping Speed, Ionic 0.06 liters/sec
X-ray Limit 2 Torr
Acceleration due to gravity mfr 9.8066
Angstrom unit mfr 10 m
Avogadro's number mfr 6.02214076
Boltzmann's constant mfr 1.38064852
Equivalent of heat mfr 4.185
Natural log base mfr 2.7183
Velocity of light mfr 2.9979
Velocity of sound mfr 343
Magnetic flux density mfr 1
Pumping speed conversion mfr 1 L/s
Throughput conversion mfr 1 cc/sec
Outgassing rate mfr 2
Envelope mfr Nonex Nonex
Grid mfr 0.025 diameter
Collector mfr 0.010 diameter
Base Leads mfr 0.060 diameter
Collector Lead mfr 0.040 diameter
Shield mfr Internally connected to filament filament
Electron Bombardment mfr 50 maximum
Gas Sensitivity (Helium) mfr 1.5 Torr
Gas Sensitivity (Nitrogen) mfr 10 Torr
Gas Sensitivity (Air) mfr 10 Torr
Gas Sensitivity Constant S mfr 10 Torr
Description of the 571 Ionization Gauge mfr 15
Electrical Connections mfr 16
Technical Specifications mfr 18
Use of The Gauge Tube mfr 21
Gas Correction Factors mfr 23
Accessories and spare parts mfr 25
Helium mfr 1.5 Torr
Nitrogen mfr 10 Torr
Air mfr 10 Torr
L7430301 mfr Glass ion gauge connector kit kit
R32443010 mfr -600 ft
R32443025 mfr -600 ft
R32443050 mfr -600 ft
NW25 KF mfr 40.1
NW40 KF mfr 55.1
base pressure (ion/turbomolecular pumps, quantum) mfr 10...9 mbar
base pressure (ion/turbomolecular pumps, accelerator) mfr 10...12 mbar
system pressure range (accelerator UHV) mfr 10...9 mbar
base pressure (NEG ion pumps) mfr 10...11 mbar
base pressure (turbo pumping systems) mfr 1 mbar
gauge measurement range mfr 1 mbar
scroll pump maintenance interval mfr 20...30 minutes
scroll pump maintenance period mfr 2...3 years
Dimensions, Weight & Mechanical Design
Atomic Mass Unit 1.6605 grams
Mounting 20
Shipping Weight 3 lbs.
Electrical, Power & Drive Specifications
Emission Setting For 0.01 Amp/torr Sensitivity 6.7 mA
Filament Current, Absolute Maximum 6.0 Amps
Filament Voltage mfr 4.0 VAC
Grid Voltage mfr 6.3 VAC
Grid Current mfr 8.7 Amps
Grid to Filament Voltage mfr 700 VDC
Filament voltage mfr 30 VDC
Grid to filament voltage for electron bombardment mfr 700 VDC
Grid current for electron bombardment mfr 100 DC
Electron bombardment power absolute maximum mfr 50 W
Grid voltage for resistance heating mfr 6.3 VAC
Grid current for resistance heating mfr 8.7 Amps
Interfaces, Connections & Control
Flange A dimension (2.75 in DN 40 ConFlat flange) mfr 69.3 mm
Flange B dimension (2.75 in DN 40 ConFlat flange) mfr 25.4 mm
Flange A dimension (NW25 KF) mfr 40.1 mm
Flange B dimension (NW25 KF) mfr 25.4 mm
Flange A dimension (NW40 KF) mfr 55.1 mm
Flange B dimension (NW40 KF) mfr 25.4 mm
Flange B dimension (1.00 in tube) mfr 25.4 mm
Flange B dimension (0.75 in tube) mfr 19.3 mm
Flange A dimension mfr 55.1 mm
Flange B dimension mfr 19.3 mm
Materials, Media & Process Compatibility
Filament Temperature, Absolute Maximum 1400 °C
Shield Coating mfr Platinum, internally connected to filament filam
Standards, Safety & General Data
Standard pressure mfr 101325 Pa

Description

Construction and series positioning

The 571 is a wide-range, linear Bayard-Alpert gauge with a burn-out-resistant yttria-coated iridium filament. It is calibrated against McLeod gauge standards. Within the Agilent vacuum transducer series, it is the high-vacuum ionization gauge, distinct from the 531 thermocouple gauge (rough vacuum), the 572 tungsten-filament Bayard-Alpert gauge, and the 525 cold-cathode gauge. The 571 is offered in multiple mechanical configurations: 1 in glass (Nonex) and Kovar tubes, 0.75 in glass and Kovar tubes, 2.75 in DN 40 ConFlat, NW25 KF, NW40 KF, and a sealed version at 10⁻⁶ Torr. All variants use the same filament and electrical ratings.

Behavior under load and in different media

The gauge measures equivalent nitrogen pressure from 10⁻¹ Torr down to the X-ray limit of about 2×10⁻¹⁰ Torr. Maximum operating pressure is 1×10⁻³ Torr; continuous operation in argon at 1×10⁻¹ Torr does not affect filament life, but tube life in air may be shortened above 1×10⁻³ Torr. Sensitivity is composition-dependent: 10/Torr for nitrogen and air at 1.0 mA emission, and 1.5/Torr for helium at 6.7 mA emission. Ionic pumping speed is 0.06 L/s for N₂ at 1 mA, which can cause gauge pumping at low pressures. Near the X-ray limit, readings are not unequivocal and degassing effects can be mistaken for the limit.

Installation, connection, and operation

The tube may be mounted vertically or horizontally and operated/degassed in either position. It should be mounted as close as possible to the measurement location through a large tubulation (1 inch OD recommended) to avoid gauge pumping. The vacuum system must be grounded to a common ground with the controller, and the ground must be tested to support at least 10 A before operation and after each maintenance procedure. Normal operating ratings are 4.0 VAC at 3.5 A filament, +30 VDC filament, +180 VDC grid, and 1 mA grid current. Degassing can be by resistance heating (6.3 VAC, 8.7 A grid, 1200 °C grid temperature) or electron bombardment (up to +700 VDC grid-to-filament, 100 mA DC, 50 W maximum). High-temperature baking should not be attempted above 10⁻⁵ Torr. The gauge is not intrinsically safe and must not be operated with combustible gases present.

Integration & Operating Notes

Electrical
Ground must support 10 A and be verified after service

The vacuum system and controller must share a common ground tested to carry at least 10 A. Inadequate grounding can expose lethal voltages up to 1000 V. Verify ground integrity after every maintenance or service procedure before applying power.

User Manual, Electrical Connections, p.16
Process Limits
Do not exceed 1×10⁻³ Torr operating pressure

The gauge is rated for a maximum operating pressure of 1×10⁻³ Torr. Prolonged exposure above this in air shortens tube life; continuous operation in argon at 1×10⁻¹ Torr is permitted without affecting filament life. Select the gauge only for high-vacuum stages.

User Manual, Technical Specifications, p.18; Operation, p.17
Process Limits
Baking above 10⁻⁵ Torr risks oxidation

High-temperature baking of the gauge should not be attempted at pressures above 10⁻⁵ Torr. Baking at higher pressures can oxidize gauge elements and make ultrahigh vacuum readings difficult to attain. Plan bakeout only after the system is below this threshold.

User Manual, Use of The Gauge Tube, p.21
Control & I/O
Controller must provide resistance degassing within ratings

Use only commercial controls equipped for resistance degassing and rated for 6.3 VAC, 8.7 A grid current, and 1200 °C grid temperature. Exceeding these absolute maximums can cause grid sagging. Electron bombardment degassing requires up to 700 VDC and must not exceed 50 W.

User Manual, Degassing Ratings, p.19
Mounting
Mount close to measurement point via 1-inch tubulation

Low conductance between the gauge and the system causes gauge pumping, reading lower than actual system pressure, particularly below 10⁻⁸ Torr. Mount the tube as close as possible to the measurement location through a large tubulation, 1 inch OD recommended.

User Manual, Mounting, p.21
Maintenance
Degas overnight for pressures below 10⁻⁸ Torr

Resistance heating degassing typically requires one-half hour or more in the 10⁻⁹ Torr range. For measurements below 10⁻⁸ Torr, allow the gauge to degas overnight. Near the X-ray limit of 2×10⁻¹⁰ Torr, distinguish degassing effects from the true X-ray limit before trusting readings.

User Manual, Degassing, p.22; X-ray Limit, p.22
Process Limits
Not intrinsically safe; no combustible gases

The 571 is not intrinsically safe. Do not turn on the gauge when combustible gases or mixtures are present. The hot filament and high voltage can act as ignition sources. Ensure the system is free of combustibles before operating.

User Manual, Use with Combustibles and Mixtures, p.11
Maintenance
Use Apiezon L grease and powder-free gloves

Do not use silicone oil or silicone grease; silicone contamination degrades vacuum performance. Use powder-free butyl or polycarbonate gloves. If grease is necessary, use Apiezon L grease (Agilent Part Number 695400004).

User Manual, Vacuum Equipment Cleanliness, p.11

Selection & Integration Q&A

When should I choose the 571 instead of the 572 tungsten-filament gauge or the 525 cold-cathode gauge?

Choose the 571 for high-vacuum measurement from 1×10⁻³ to 2×10⁻¹⁰ Torr with a burn-out-resistant yttria-coated iridium filament that tolerates accidental atmosphere exposure. The 572 uses a tungsten filament; the 525 cold-cathode gauge suits different pressure ranges. Select the 571 when filament robustness and linear Bayard-Alpert output are required.

What controller and grounding are required to operate the 571?

Use a commercial controller equipped for resistance degassing and rated for 4.0 VAC/3.5 A filament, +30 VDC filament, +180 VDC grid, and 1 mA grid current. The vacuum system must share a common ground with the controller, tested to support at least 10 A before operation and after each service. Verify controller compatibility with these voltages.

Can the 571 be mounted in any orientation, and what environment limits apply?

The tube may be mounted and operated/degassed vertically or horizontally for prolonged periods. Mount close to the measurement point via a 1-inch OD tubulation to avoid gauge pumping below 10⁻⁸ Torr. Do not bake above 10⁻⁵ Torr, and do not operate with combustible gases present because the gauge is not intrinsically safe.

What maintenance and accessories are needed beyond the base gauge tube?

Plan for degassing: resistance heating typically takes one-half hour or more at 10⁻⁹ Torr, and overnight degassing is usually needed below 10⁻⁸ Torr. Accessories include a glass ion gauge connector kit and controller cables in 10 ft, 25 ft, and 50 ft lengths. Use Apiezon L grease (P/N 695400004) if grease is necessary; silicone grease is prohibited.