Agilent Technologies Refinery gas analyzer, four channel multi-dimensional system

Agilent Technologies Refinery gas analyzer, four channel multi-dimensional system — Gas chromatography.

The Agilent four-channel refinery gas analyzer is a multi-dimensional Micro GC configuration with three backflush channels (10 m CP-Molesieve 5 Å, 10 m CP-PoraPLOT U, 10 m CP-AL2O3/KCl) and one straight 8 m CP-Sil 5CB channel. It targets fixed gases, C1–C6 hydrocarbons and sulfur compounds in refinery gas streams, using argon carrier gas at 80 °C column temperature and 300 kPa column head pressure.

Agilent Technologies Energy Chemical GC Analyzers Series — Model Comparison & Selection

Model Typical quantitation limit, mol%Carrier gasAnalysis time, min
Refinery gas analyzer, four channel multi-dimensional system Current —Argon—
Analyzers for biodiesel, biofuels 0.01——
Gasoline analyzer, benzene analysis 0.01——
GCxGC-based analyzers, aviation turbine fuels —Hydrogen—
natural gas analyzer, fast analysis times 0.01—20
NGA Analyzer, natural gas analysis 0.01—20

Full specifications

Core Performance & Functional Specifications
Carrier gas Argon
Column temperature 80 °C
Channel 1 column mfr 10 backflush
Channel 1 compounds analyzed mfr 2
Channel 2 column mfr 10 backflush
Channel 2 compounds analyzed mfr 2 S
Channel 3 column mfr 10 backflush
Channel 3 compounds analyzed mfr 3 olefins
Channel 4 column mfr 8 straight
Channel 4 compounds analyzed mfr 6 hydrocarbons
Injection time mfr 40 ms
Column head pressure mfr 200 kPa
Dimensions, Weight & Mechanical Design
Depth 32.94 cm
Height 28.28 cm
Weight 7.3 kg
Width 14.5 cm
Operating Conditions, Environment & Protection
Operating temperature 0 °C

Description

Construction and positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Argon carrier is mandatory for the specified channel set

The four-channel configuration is specified with argon carrier gas. Substituting another carrier changes response on the Molesieve and PoraPLOT U channels and breaks comparability with the published RGA configuration; confirm argon supply purity and delivery pressure before commissioning.

Factory Specification Sheet; Application Note 5994-1043EN
Mounting
Site preparation follows the 990 Micro GC checklist

Bench space, power supply components and environmental conditions must be prepared per document G3588-90007 Rev 1.0 (January 2022), which covers the 990 Micro GC, Mobile Micro GC Quad channel and power supply. The checklist is the controlling document for utilities and clearance before installation.

Agilent 990 Micro GC Site Preparation Checklist, G3588-90007 Rev 1.0
Control & I/O
Injection timing is fixed at 40 ms per channel

The method uses a 40 ms injection time. The injection event must be synchronized across the four channels so that backflush and straight-column runs start together; any change to injection duration alters peak areas and must be re-validated against the RGA configuration.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose this four-channel RGA configuration instead of the three-channel RGA configuration on the same platform?

Pick the four-channel version when the 8 m CP-Sil 5CB straight column is needed for the six-hydrocarbon separation alongside the three backflush channels. The three-channel RGA configuration uses a backflush-to-detector AL2O3/KCl channel and omits the straight Sil 5CB channel, so it covers fewer hydrocarbon compounds.

What carrier gas and inlet conditions must the adjacent gas supply system provide for this analyzer?

Supply argon carrier regulated to 300 kPa column head pressure. The method runs at 80 °C column temperature and 110 °C injector temperature with a 40 ms injection; the gas panel must hold pressure stable at the specified setpoint, otherwise retention times on all four channels drift.

What site and environmental conditions must be prepared before installing this unit?

Follow the 990 Micro GC Site Preparation Checklist G3588-90007 Rev 1.0 (January 2022), which defines bench space, power supply components and environmental conditions for the 990 Micro GC and Mobile Micro GC Quad channel. The unit operates from 0 °C ambient and weighs 7.3 kg at 14.5 cm width.

What maintenance and consumables are not part of the base delivery and must be planned separately?

The base unit ships without the argon carrier supply, pressure regulation to 300 kPa, and the site-preparation items listed in G3588-90007 (power supply components, bench utilities). Column sets are configuration-specific: the four-channel RGA set comprises three 10 m backflush columns and one 8 m straight column, ordered as a defined configuration.