BINDER CB 260
BINDER ·Made in Germany · In Active Production

BINDER CB 260

BINDER CB 260 — Incubators with Temperature 5…50 °C.

The BINDER CB 260 is a 267 L CO2 incubator with a 200–230 V single-phase supply and a temperature range of +5 to 50 °C, distinguished within the CB series by its larger interior volume compared to the CB 170 and CB 56. It provides stable culture conditions with 0.1 K fluctuation and 0.4 K uniformity, and is intended for cell and tissue culture in research and production laboratories.

Temperature
5…50 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER CB Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmHumidity range, % rhInterior volume, LPermitted load, kg
CB 260 Current 5–5078.510290...9526740
CB 170 4–5071.58790 ...9517030
CB 56 4–5054.567.490...955330

Full specifications

Core Performance & Functional Specifications
Temperature 5...50 °C
CO2 recovery time 6 min (all models)
Interior volume 267 L (all models)
Load per rack 10 kg (all models)
Nominal power 1.5 kW (all models)
Permitted load 40 kg (all models)
Temperature fluctuation 0.1 K
Temperature uniformity 0.4 K
Back wall clearance 100 mm (all models)
Side wall clearance 50 mm (all models)
Article Number mfr 9040
Recovery time after door was opened for 30 s at 37°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 785 mm (all models)
Height 1020 mm (all models)
Width 740 mm (all models)
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (all models)
Rated voltage 200...230 V (CB260-230V / CB260-230V-O)
Unit fuse 10.0 A (CB260-230V / CB260-230V-O)
Operating Conditions, Environment & Protection
Humidity range 90...95 % RH

Description

Construction and series positioning

Behavior under load and in operation

Installation, connection, and control

Integration & Operating Notes

Mounting
Maintain 100 mm rear and 50 mm side clearance

Insufficient clearance restricts heat dissipation from the chamber walls and can cause overheating, reduced temperature uniformity, or premature component failure. Ensure the bench or floor stand can support the chamber weight and that the rear is accessible for gas and power connections.

Operating Manual CB/CBF (E7.1), Section 4 Installation and Connections
Electrical
Use a 10 A fused supply; do not exceed rated voltage

The unit fuse is 10.0 A for 200–230 V models. Connecting to a supply outside the rated voltage range or using an undersized fuse will compromise safety and may damage the controller. Verify that the supply can deliver the nominal 1.5 kW without significant voltage drop.

Data Sheet Model CB 260 en.pdf; Operating Manual CB/CBF (E7.1), Section 4.6 Electrical Connection
Process Limits
CO2 control is inactive during sterilization and above 80 °C

During hot-air sterilization, CO2 and O2 control are automatically deactivated. They remain off until the chamber temperature falls below 80 °C and the user confirms the 'Successful sterilization' message. Do not expect gas control to function during this period; plan cell culture work accordingly.

Operating Manual CB/CBF (E7.1), Section 21.3.5 Hot-air sterilization
Control & I/O
Tolerance alarms activate only after setpoint is reached

The tolerance range alarm (e.g., ±2 °C around 37 °C) does not trigger until the setpoint has been reached at least once. This prevents false alarms during warm-up. Configure the alarm delay to avoid nuisance alarms during door openings or transient deviations.

Operating Manual CB/CBF (E7.1), Section 11.5 Tolerance range and alarm delay
Maintenance
Replace door gasket only when chamber is cold

Replacing the door gasket while the chamber is warm can deform or damage the gasket, leading to leaks and loss of temperature/humidity control. Allow the chamber to cool to ambient temperature before replacement. Also, after replacing the CO2 sensor, perform a CO2 control adjustment.

Operating Manual CB/CBF (E7.1), Section 22.5 Maintenance and Service
Maintenance
Annual electrical safety test and maintenance required

Perform electrical safety tests according to EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial startup and after any maintenance or repair. Schedule at least annual maintenance; a maintenance message appears after 8760 operating hours or two years and cannot be permanently cleared by the user.

Operating Manual CB/CBF (E7.1), Sections 22.1 and 22.5

Selection & Integration Q&A

When should I choose the CB 260 over the CB 170 or CB 56?

Choose the CB 260 when you need a larger working volume (267 L) for high-throughput culture or multiple users. The CB 170 (170 L) and CB 56 (56 L) are better suited for smaller labs or lower sample numbers. All share the same temperature range and controller features.

What are the requirements for connecting CO2 and O2 gases to the CB 260?

Use the gas cylinder connection kits (accessory) for CO2, O2, and N2. For standard CO2 control, only CO2 is required. For -O variants with O2 control, both O2 and N2 are needed. Ensure the gas supply pressure matches the specifications in the manual and that the connections are leak-tested.

What maintenance items and accessories are not included with the base CB 260?

The base delivery does not include gas cylinder connection kits, the BINDER-Water-Set for humidity, additional shelves, or the divided inner door. Also, annual electrical safety testing and sensor replacement (after 8760 h or 2 years) are not included. Order these accessories and plan service accordingly.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model CB 260 en
Datasheet ·443,8 КБ ·9 pp.
View PDF
PDF
CB+CBF (E7.1) 01 2026 en
Manual ·19,4 МБ ·201 pp.
View PDF