BINDER CB-S 170
BINDER CB-S 170 — Incubators with Temperature 6…50 °C.
The CB-S 170 is a 170 L CO2 incubator with IR CO2 sensing and hot-air sterilization, configured for 200–230 V single-phase supply. It holds +6 to 50 °C with 0.1 K fluctuation and 0.3 K uniformity, and recovers to setpoint in 6 min after a 30 s door opening at 37 °C. It serves as the standard-capacity chamber for cell culture where 170 L and six shelves suffice.
Direct drop-in replacements
BINDER CB-S Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Permitted load, kg | Power frequency, Hz |
|---|---|---|---|---|---|---|
| CB-S 170 Current | 6–50 | 71.5 | 87 | 170 | 30 | 50–60 |
| CB-S 260 | 6–50 | 78.5 | 102 | 267 | 40 | 50 |
Full specifications
| Temperature | 6...50 °C |
| CO2 measuring technology | IR (all variants) |
| CO2 recovery time | 6 min (all variants) |
| Interior volume | 170 L (all variants) |
| Load per rack | 10 kg (all variants) |
| Nominal power | 1.3 kW (all variants) |
| Permitted load | 30 kg (all variants) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Back wall clearance | 100 mm (all variants) |
| Side wall clearance | 50 mm (all variants) |
| Article Number mfr | 9040 |
| Recovery time after door was opened for 30 s at 37°C mfr | 6 min |
| Depth | 715 mm (all variants) |
| Height | 870 mm (all variants) |
| Width | 680 mm (all variants) |
| Power frequency | 50/60 Hz (all variants) |
| Rated voltage | 200...230 V (CBS170-230V / CBS170-230V-L) |
| Unit fuse | 10.0 A (CBS170-230V / CBS170-230V-L) |
| Humidity range | 90...95 % RH |
Description
Construction and series positioning
- The CB-S 170 (Art. 9040-0219) is the 170 L member of the CB-S series; the CB-S 260 is the larger sibling with identical +6 to 50 °C range but eight shelves instead of six.
- This model is the 230 V standard-hinge variant; left-hinged (CBS170-230V-L) and 100–120 V UL variants (CBS170UL-120V, -L) share the same chamber but differ in voltage, plug, and door swing.
- The inner chamber accepts a maximum of six shelves; the CO2 sensor is an IR type rated to 60 °C maximum.
Behavior under load and in operation
- Temperature control holds 0.1 K fluctuation and 0.3 K uniformity across the +6 to 50 °C working range.
- Humidity runs 90–95 % RH; recovery after a 30 s door opening at 37 °C is 6 min.
- Load limits: 10 kg per rack and 30 kg total permitted load; nominal power is 1.3 kW.
- Hot-air sterilization is effective only if run at least 4 hours; aborting earlier leaves the cycle ineffective.
Installation, control, and operation
- The RD4 controller manages temperature and CO2 setpoints, alarms, tolerance ranges, and Ethernet networking; a zero-voltage relay alarm output and optional analog outputs for temperature and CO2 are available.
- Electrical supply is 200–230 V, 50/60 Hz, 1 N~, with a 10.0 A unit fuse; nominal current is 5.7 A at 230 V.
- Wall clearance is 100 mm at the rear and 50 mm at the sides; IP20 requires dry indoor installation.
- The CO2 gas connection uses a cylinder and hose kit (accessory); the gas inlet fine filter (0.45 µm, 99.99 %) is checked at each maintenance interval and replaced at least annually.
Integration & Operating Notes
The 230 V model draws 5.7 A nominal and is protected by a 10.0 A unit fuse; connecting to a 100–120 V supply or a circuit rated below the fuse will prevent startup or cause repeated fuse trips. Verify voltage and frequency within ±10 % before first power-on.
The chamber requires 100 mm clearance at the back and 50 mm at each side; insufficient clearance restricts heat dissipation and can raise internal temperatures or trip the over-temperature protective device. IP20 rating also mandates a dry indoor location.
The CB-S 170 inner chamber accepts a maximum of six shelves, with 10 kg per rack and 30 kg total permitted load. Exceeding these limits can deform shelves or compromise temperature uniformity; the CB-S 260 is the alternative when more capacity is needed.
The aseptic gas inlet fine filter (0.45 µm, 99.99 % efficiency) must be checked at each maintenance interval and replaced at least annually to prevent dirt accumulation in gas inlet valves and tubes. Schedule this with the annual electrical safety test per EN 50678:2020 and EN 50699:2020.
The RD4 controller provides a zero-voltage relay alarm output (chap. 15.5) that can be wired to an external alarm device or building monitoring system. Plan the contact rating and wiring during installation; optional analog outputs for temperature and CO2 are separate accessories.
Premature termination of the hot-air sterilization cycle after less than 4 hours is ineffective; only after more than 4 hours (during the cooling-down phase) is sterilization successful. The CO2 sensor must not exceed 60 °C during any procedure, so verify sensor protection before starting a cycle.
Selection & Integration Q&A
Pick the CB-S 170 when 170 L and six shelves cover your workload; the CB-S 260 adds volume and two more shelves at the same temperature range. Choose the -L variant if door swing must be left-hand, and the UL variant only for 100–120 V supplies.
It needs 200–230 V, 50/60 Hz, 1 N~ with a 10.0 A fuse (5.7 A nominal). CO2 is supplied from a gas cylinder via the gas hose and connection kit (accessory); the inlet fine filter is 0.45 µm. Ethernet and zero-voltage relay are available for monitoring.
Install indoors (IP20) with 100 mm rear and 50 mm side clearance. Load max 10 kg per rack and 30 kg total across six shelves. Hot-air sterilization must run at least 4 hours; the CO2 sensor must stay below 60 °C.
Base delivery excludes the CO2 gas cylinder connection kit, silicone access ports (8012-0558/0559/0560), base on castors, stacking frame, APT-COM 4 software, and analog outputs. Plan annual gas inlet filter replacement and electrical safety testing per EN 50678/50699.