BINDER CB-S 260
BINDER ·Made in Germany · In Active Production

BINDER CB-S 260

BINDER CB-S 260 — Incubators with Temperature 6…50 °C.

The CB-S 260 is a 267 L CO₂ incubator with IR CO₂ sensing and hot-air sterilization, configured for 200–230 V single-phase supply (CBS260-230V). It extends the CB-S series to 8 shelves and 40 kg total load, with 0.1 K fluctuation and 0.4 K uniformity at +6 to 50 °C. It serves as the controlled incubation stage for cell culture where contamination control and CO₂ recovery (5 min) matter.

Temperature
6…50 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER CB-S Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LPermitted load, kgPower frequency, Hz
CB-S 260 Current 6–5078.51022674050
CB-S 170 6–5071.5871703050–60

Full specifications

Core Performance & Functional Specifications
Temperature 6...50 °C
CO2 measuring technology IR
Interior volume 267 L (CBS260-230V / CBS260UL-120V)
Load per rack 10 kg (both variants)
Nominal power 1.5 kW
Permitted load 40 kg (both variants)
Temperature fluctuation 0.1 K
Temperature uniformity 0.4 K
Back wall clearance 100 mm (both variants)
Side wall clearance 50 mm (both variants)
Article Number mfr 9040
Recovery time after door was opened for 30 s at 37°C mfr 8 min
CO2 recovery time after door was opened for 30 s at 5 vol. % CO2 mfr 5 min
Dimensions, Weight & Mechanical Design
Depth 785 mm (both variants)
Height 1020 mm (both variants)
Width 740 mm (both variants)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 200...230 V
Unit fuse 10.0 A
Phase (Nominal voltage) mfr 1
Operating Conditions, Environment & Protection
Humidity range 90...95 % RH

Description

Construction and series positioning

Behavior under load and in operation

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must be 200–230 V 1~ with 10 A fusing

The CBS260-230V draws 1.5 kW nominal and requires a 200–230 V single-phase 50/60 Hz supply protected by a 10.0 A fuse. Verify voltage and frequency tolerance before connection; the 100–120 V UL variant is a separate model (CBS260UL-120V) and cannot be fed from the same circuit.

Data Sheet Model CB-S 260 en.pdf; Operating Manual CB-S (E7.1), chap. 1.6
Process Limits
Keep CO₂ sensor below 60 °C during sterilization

The IR CO₂ sensor is temperature resistant only up to 60 °C / 140 °F. During hot-air sterilization and any cleaning procedure, ensure the sensor is not exposed above this limit; exceeding it risks sensor damage and loss of CO₂ control.

Operating Manual CB-S (E7.1), chap. 24 / page 90
Process Limits
Sterilization cycle must run at least 4 hours

Hot-air sterilization is ineffective if terminated before 4 hours have elapsed. If aborted after more than 4 hours (during the cooling-down phase), sterilization is still considered successful. Plan the duty cycle so the chamber is not needed for incubation during this window.

Operating Manual CB-S (E7.1), chap. 23.4 / pages 87–88
Maintenance
Replace gas inlet fine filter at least annually

The aseptic gas inlet fine filter (0.45 µm, 99.99 % efficiency) is checked at each maintenance interval and replaced at least once a year. Skipping replacement allows dirt to accumulate in gas inlet valves and tubes, compromising CO₂ supply and sterility.

Operating Manual CB-S (E7.1), chap. 24.4 / page 92
Maintenance
Annual electrical safety test required before reuse

An annual repeat electrical inspection is required, plus a test before initial startup and before resuming operation after maintenance or repair, recommended under EN 50678:2020 and EN 50699:2020. Schedule these tests so the chamber is not returned to service without a valid safety check.

Operating Manual CB-S (E7.1), chap. 24.1 / page 90
Control & I/O
Zero-voltage relay can drive external alarm devices

The RD4 controller provides a zero-voltage relay alarm output (chap. 15.5) that can be wired to external alarm devices or a monitoring system. Use it to propagate door-open, temperature, CO₂, or power-failure alarms to a central station; verify contact rating before connecting loads.

Operating Manual CB-S (E7.1), chap. 15.5 / page 62
Control & I/O
Ethernet allows remote monitoring via APT-COM 4

Network settings include MAC address, IP address, subnet mask, gateway, and DNS; IP assignment can be automatic or manual. Integrating the chamber into the lab network enables remote monitoring and data logging with APT-COM 4 software, which is an accessory and not included in base delivery.

Operating Manual CB-S (E7.1), chap. 16 / pages 63–68

Selection & Integration Q&A

When should I choose the CB-S 260 over the CB-S 170?

Pick the CB-S 260 when you need 267 L interior volume and up to 8 shelves (40 kg total load) rather than the CB-S 170's 6 shelves. Both share the +6 to 50 °C range and IR CO₂ sensing; the 260 suits larger culture batches, while the 170 fits smaller benches.

What gas supply and connection parts are required for the CB-S 260?

The chamber needs a CO₂ gas cylinder and gas hose (gas cylinder connection kit is an accessory). The gas inlet fine filter (0.45 µm) is built in and must be replaced annually. Verify inlet pressure using the bar–psi conversion table in the manual before connecting.

What maintenance and accessories are not included in the base delivery?

Base delivery excludes APT-COM 4 software, analog outputs, silicone access ports (e.g., 8012-0558 rear), base on castors, and stacking frame. Plan for annual gas inlet fine filter replacement and annual electrical safety testing per EN 50678/50699.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model CB S 260 en
Datasheet ·495,9 КБ ·7 pp.
View PDF
PDF
CB S (E7.1) 02 2026 en
Manual ·10,7 МБ ·119 pp.
View PDF