BINDER FED 400
BINDER FED 400 — Drying ovens with Temperature 5…300 °C.
The FED 400 is a forced-convection drying and heating oven with a 400 L stainless-steel interior (V2A 1.4301/AISI 304) and APT.line™ preheating chamber, rated +5 to 300 °C. Within the FED series it is the 400 L modification with 3.4 kW heating and 18 x/h air change, distinguished from smaller siblings by its 1,000 × 800 × 765 mm working volume and 90 kg total load capacity. It serves batch drying, heating and thermal testing where uniform temperature distribution (3.8 K) and short recovery after door opening are required.
BINDER FED Series — Model Comparison & Selection
Full specifications
| Temperature | 5...300 °C |
| Air change rate | 18 x/h |
| Heating time to 150°C | 29 min |
| Interior volume | 400 L |
| Load per rack | 35 kg (FED400-400V / FED400UL-208V) |
| Nominal power | 3.4 kW |
| Permitted load | 90 kg (FED400-400V / FED400UL-208V) |
| Temperature fluctuation | 0.7 K |
| Temperature uniformity | 3.8 K |
| Rear wall clearance | 160 mm (FED400-400V / FED400UL-208V) |
| Side wall clearance | 100 mm (FED400-400V / FED400UL-208V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 6 min |
| Depth | 765 mm (FED400-400V / FED400UL-208V) |
| Height | 1025 mm (FED400-400V / FED400UL-208V) |
| Interior height | 800 mm (FED400-400V / FED400UL-208V) |
| Interior width | 1000 mm (FED400-400V / FED400UL-208V) |
| Width | 1235 mm (FED400-400V / FED400UL-208V) |
| Net weight of the unit (empty) mfr | 145 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
Description
Construction and series positioning
- Electrically heated oven with fan-assisted forced-air circulation, electronic PID controller with 1 °C display resolution, and APT.line™ preheating chamber.
- Inner chamber, preheating chamber and door interior in stainless steel V2A (1.4301 / AISI 304); housing powder-coated RAL 7035 or RAL 9003.
- This 400 L model sits between the FED 260 and FED 720 siblings; unlike the smaller FED 56/115/260 (+10 °C start) it is rated from +5 °C above ambient, and it provides 400 L interior volume with 1,000 mm width, 800 mm height and 765 mm depth.
- Two electrical variants exist: FED400-400V (400 V 3N~ 50/60 Hz, CEE 5-pole) and FED400UL-208V (208 V 3~ 60 Hz, NEMA L21-20P).
Behavior under load and in operation
- Temperature range +5 to 300 °C; nominal power 3.4 kW, nominal current 7.4 A, unit fuse 16 A.
- Heating-up time to 150 °C is 29 min; recovery after a 30 s door opening at 150 °C is 6 min.
- Temperature fluctuation 0.7 K and temperature uniformity 3.8 K define the working tolerance across the 400 L volume.
- Air change rate 18 x/h supports moisture and vapor removal during drying; the ventilation set-point is entered via the controller.
- Load limits: 35 kg per rack and 90 kg total permitted load; exceeding these risks rack deformation and impaired uniformity.
- Above 150 °C, natural oxidation may discolor stainless surfaces yellowish-brown or blue; this is harmless and does not affect function.
Installation, control and operation
- Indoor installation only; unventilated recesses are prohibited. Minimum wall clearances: 160 mm rear and 100 mm sidewise for heat dispersal.
- Start-up by pressing the switch until the display lights up; the red heating control light indicates active heating.
- Temperature and ventilation set-points are entered with or without a temperature ramp; time functions offer Continuous and Timer operation via the time management key.
- Standard temperature safety device class 2 per DIN 12880:2007 with visual and audible alarm; class 3.1 and a disconnectable audible over-temperature alarm are options.
- RS 422 serial interface enables computer communication, e.g. via APT-COM™ 4 Multi Management Software.
- IP20 protection: no water protection; the chamber must not become wet during operation, cleaning or maintenance.
Integration & Operating Notes
The type plate specifies 400 V 3N~ with a ±10% tolerance at the indicated power frequency. Operation outside this band can cause controller faults, contactor chatter or heating underperformance; verify site voltage and frequency before connecting the 5-pole CEE plug.
The chamber dissipates heat through its surfaces; insufficient clearance in recesses or against walls causes overheating and possible damage. Installation in unventilated recesses is prohibited, and the unit must not be placed in damp areas, puddles, hazardous locations or explosive atmospheres.
The chamber has no explosion protection; combustible or explosive substances at working temperature must not be introduced, and the operating temperature must remain below the flash point or sublimation point of the charge. Acids, halides and other corrosive ingredients are excluded and can damage chamber components.
If the temperature safety device (class 2 standard) responds, the chamber becomes inactive with no heating and the display shows the alarm value. This is expected behavior; investigate and correct the over-temperature cause before resetting, otherwise the trip will recur.
Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER. Before any maintenance, switch off at the main power switch and disconnect the power plug; the rear panel must not be removed. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required, plus tests before initial start-up and after repair.
Replacing the door gasket while the chamber is hot may damage the new gasket and compromise the seal. Allow the chamber to cool to ambient before replacement, and never clean the chamber while it is switched on (listed as foreseeable misuse).
Natural oxidation of stainless steel surfaces above 150 °C produces yellowish-brown or blue tints. These are harmless and do not impair function or quality; no corrective action or warranty claim is warranted for this appearance.
Selection & Integration Q&A
Choose the FED 400 for 400 L batches needing +5 °C start and 90 kg total load; the FED 260 offers less volume, while the FED 720 gives more capacity but starts at +12 °C. Select by load size and lowest required temperature.
The FED400-400V needs 400 V 3N~ 50/60 Hz within ±10%, 7.4 A nominal, protected at 16 A, with a 5-pole CEE receptacle. The UL variant needs 208 V 3~ 60 Hz and a NEMA L21-20P receptacle.
Indoor use only, no unventilated recesses, damp areas or explosive atmospheres. Maintain 160 mm rear and 100 mm side clearance. The chamber controls from +5 °C above ambient to 300 °C; IP20 means no water exposure.
Annual service by BINDER-authorized electricians and annual electrical inspection per EN 50678/50699 are mandatory. RS 422 cable and APT-COM™ 4 software, HEPA filter, class 3.1 safety device, analog output and interior lighting are options not included.