BINDER FED 115
BINDER ·Made in Germany · In Active Production

BINDER FED 115

BINDER FED 115 — Drying ovens with Temperature 10…300 °C.

The FED 115 is a forced-convection drying and heating oven with enhanced timer functions, 116 L interior, and a 10–300 °C working range. Within the FED series it is the mid-size 230 V single-phase model (1.3 kW, 5.7 A), distinguished from the 720 by supply and from smaller sizes by usable volume. It serves precise tempering, drying, and heat treatment of non-flammable solids and bulk materials in laboratory and production processes.

Temperature
10…300 °C

BINDER FED Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
FED 115 Current 10–300326051870.5116
FED 260 10–30097601994259
FED 400 5–3001876529102.5400
FED 56 10–300805651562.560
FED 720 12–300587025159741

Full specifications

Core Performance & Functional Specifications
Temperature 10...300 °C
Air change rate 32 x/h
Heating time to 150°C 18 min
Interior volume 116 L
Load per rack 30 kg (FED115-230V / FED115UL-120V)
Nominal power 1.3 kW
Permitted load 150 kg (FED115-230V / FED115UL-120V)
Temperature fluctuation 0.3 K
Temperature uniformity 1.2 K
Rear wall clearance 160 mm (FED115-230V / FED115UL-120V)
Side wall clearance 100 mm (FED115-230V / FED115UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 605 mm (FED115-230V / FED115UL-120V)
Height 705 mm (FED115-230V / FED115UL-120V)
Width 710 mm (FED115-230V / FED115UL-120V)
Net weight of the unit (empty) mfr 54 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, control and operation

Integration & Operating Notes

Mounting
Keep 160 mm rear and 100 mm side clearance; install indoors only

The chamber must not be set in unventilated recesses. Rear clearance 160 mm and side clearance 100 mm are required for heat dissipation and airflow. Ambient must stay 18–40 °C and ≤70 % r.h. non-condensing; altitude ≤2000 m. IP 20 means no water protection—do not spill liquids.

Operating Manual 07/2025, Installation location p.20 and Quick Start Guide p.18
Process Limits
Load ≤150 kg total, ≤30 kg per rack, and ≤50 % usable volume

Exceeding the load or filling more than half the usable volume prevents sufficient airflow and homogeneous temperature. Maintain at least 10 % of interior dimensions as wall clearance. Charging materials must be non-flammable and operating temperature below flash point; corrosive acids/halides are excluded.

Operating Manual 07/2025, Intended use p.12-13 and Placing charging material p.19
Maintenance
Annual maintenance and electrical safety test are mandatory

The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized personnel; warranty is void if non-authorized personnel perform maintenance. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required, plus a test before initial startup and after any repair. Replace door gasket only when cold.

Operating Manual 07/2025, Maintenance intervals p.69
Cooling & Media
Hot-air sterilization possible at 190 °C for ≥30 min

For decontamination, the FED series can be hot-air sterilized at 190 °C for at least 30 minutes; all flammable goods must be removed first. Recommended cleaning agents are neutral cleaner art. 1002-0016 and disinfectant spray art. 1002-0022; do not use neutral cleaner on zinc-coated hinge parts or rear wall.

Quick Start Guide 03/2023, Cleaning and decontamination p.23

Selection & Integration Q&A

When should I choose the FED 115 over the FED 260 or FED 720?

Choose the FED 115 for 116 L usable volume and 230 V single-phase supply (1.3 kW, 5.7 A). The FED 260 offers larger volume but also 230 V; the FED 720 requires 400 V three-phase, has a two-door design, and its max set-point depends on ambient temperature. Select the 115 when bench space and single-phase power are constraints.

What are the power supply and fusing requirements for the FED 115?

The FED115-230V requires 230 V 1~ 50/60 Hz, nominal power 1.3 kW, current 5.7 A, with an external 6.3 A time-lag (T) 5×20 mm / 250 V fuse. Supply fluctuation must not exceed ±10 %. The UL variant needs 120 V 1~ 60 Hz, 11.7 A. Verify the type plate before connection.

What ambient and installation limits apply to the FED 115?

Install indoors on a non-flammable surface with ambient 18–40 °C, max 70 % r.h. non-condensing, altitude ≤2000 m. Maintain 160 mm rear and 100 mm side clearance; do not install in unventilated recesses. IP 20 protects against solid objects >12.5 mm but not water. Supply tolerance ±10 %.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance by authorized personnel and annual electrical safety test per EN 50678/VDE 0701 and EN 50699/VDE 0702 are mandatory. Recommended consumables: neutral cleaner art. 1002-0016, disinfectant spray art. 1002-0022. Optional accessories include APT-COM 4 software, Pt100 sensor, analog output, HEPA filter, and inert gas connection. Ethernet is standard; USB for data recording.

Dimensional Drawings & Technical Documentation