BINDER KT 170
BINDER ·Made in Germany · In Active Production

BINDER KT 170

BINDER KT 170 — Incubators with Temperature 4…100 °C.

The BINDER KT 170 is a 163 L Peltier-cooled incubator with a 4–100 °C working range, 0.4 K uniformity and 0.1 K fluctuation, positioned between the KT 53 and KT 115 as the largest standard chamber in the series. Its 100 W max heat compensation and 30 kg per-rack load capacity suit refrigerated incubation and alternating 37 °C / 4 °C protocols in microbiology and long-term storage.

Temperature
4…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KT Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KT 170 Current 4–10065.5102.516360030
KT 115 4–10065.571.510260030
KT 53 4–1006363.55340015

Full specifications

Core Performance & Functional Specifications
Temperature 4...100 °C
Interior volume 163 L
Load per rack 30 kg
Max. heat compensation 100 W
Nominal power 0.8 kW
Permitted load 120 kg (KT170-230V / KT170UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.4 K
Back wall clearance 100 mm (KT170-230V / KT170UL-120V)
Side wall clearance 240 mm (KT170-230V / KT170UL-120V)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 655 mm (KT170-230V / KT170UL-120V)
Height 1025 mm (KT170-230V / KT170UL-120V)
Interior width 600 mm (KT170-230V / KT170UL-120V)
Width 860 mm (KT170-230V / KT170UL-120V)
Net weight of the unit (empty) mfr 102 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 200...240 V
Unit fuse 10.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

The KT 170 uses Peltier refrigerating technology with HIT insulation and the T4.12 program controller. Within the KT series it is the largest standard model (163 L interior, 600 mm interior width, 860 mm net width, 102 kg empty), above the KT 115 and KT 53, which share the same 4–100 °C range. The 230 V variant (KT170-230V, Art. 9020-0289) is rated 200–240 V 1~ 50/60 Hz with a CEE 7/7 plug; the UL variant (KT170UL-120V, Art. 9020-0310) is 100–120 V 1~ with a NEMA 5-20 plug and UL 61010-1 certification. Nominal power is 0.8 kW, nominal current 3.0 A, unit fuse 10.0 A.

Behavior under load and in different modes

The intended application range is 4–37 °C for cultivation of microorganisms with narrow temperature optimum, including long-term storage at 4 °C, refrigerated incubation at 20–25 °C, incubation at 37 °C and alternating 37 °C / 4 °C. Temperature fluctuation is 0.1 K and uniformity 0.4 K; recovery time after a 30 s door opening at 37 °C is 3 min. Maximum heat compensation is 100 W, so internal heat loads from the load must be assessed against this limit. Permitted load is 120 kg total, with 30 kg per rack. The chamber is protected to IP 20 and carries a temperature safety device class 3.1 per DIN 12880:2007.

Installation, control and operation

The T4.12 controller supports fixed-value mode, time programs and week programs with set-point ramps/steps, tolerance ranges and optional zero-voltage relay outputs. Data can be exported/imported via USB. Optional equipment includes door heating, zero-voltage relay alarm/control outputs, object temperature display with flexible Pt 100 sensor and APT-COM 4 software. The chamber must not be installed in unventilated recesses; prescribed wall clearances are 100 mm at the rear and 240 mm sidewise to allow heat dispersal. The type plate is on the left chamber side, bottom right.

Integration & Operating Notes

Mounting
Rear and side clearances are mandatory for heat dispersal

The Peltier system rejects heat to the ambient air, so the chamber requires 100 mm rear clearance and 240 mm sidewise clearance. Installing it in an unventilated recess or closer to walls risks overheating and chamber damage; verify the final bench layout against these distances before positioning the 860 mm wide, 655 mm deep unit.

Data Sheet Model KT 170 en.pdf; Instruction Manual KT (E6.1), chap. 3.4 Installation
Maintenance
Annual maintenance by BINDER-authorized personnel is mandatory

Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER; warranty becomes void if maintenance is conducted by non-authorized personnel. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, and the chamber must be tested before initial startup and before resuming operation after maintenance or repair.

Instruction Manual KT (E6.1), chap. 21.1/21.2, p.108
Maintenance
Clean the Peltier fan grid several times a year in dusty air

With increased dust in the ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year to maintain cooling performance. Before any maintenance work, turn off the chamber at the main power switch and disconnect the power plug; do not remove the rear panel. Replace the door gasket only when the chamber is cold to avoid damaging it.

Instruction Manual KT (E6.1), chap. 21.2, p.108

Selection & Integration Q&A

When should I choose the KT 170 over the KT 115 or KT 53?

Choose the KT 170 when the 163 L interior and 600 mm width are needed for larger batches or long-term 4 °C storage; the KT 115 and KT 53 share the same 4–100 °C range but offer smaller interiors. All three use Peltier technology and the T4.12 controller, so selection is driven by chamber volume and load, not temperature capability.

What supply and protection must be provided for the KT 170?

The 230 V model requires 200–240 V 1~ 50/60 Hz within ±10%, 0.8 kW nominal power, 3.0 A nominal current and a 10.0 A unit fuse; the UL model requires 100–120 V 1~ with a NEMA 5-20 plug. Confirm the type plate voltage class before ordering, since the two variants are not interchangeable on the same supply.

What ambient and installation limits apply to the KT 170?

Install with 100 mm rear and 240 mm sidewise clearance and never in an unventilated recess, because the Peltier system disperses heat to the ambient air. The chamber is IP 20 and must be kept splash-proof. In dusty air, clean the Peltier fan grid several times a year to preserve cooling performance.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance by BINDER-authorized personnel and annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required; warranty is void if non-authorized personnel perform maintenance. Optional items not in base delivery include door heating, zero-voltage relay alarm/control outputs, Pt 100 object temperature sensor and APT-COM 4 software.