Shimadzu Nexera Mikros
Shimadzu Nexera Mikros — Mass spectrometers.
The Nexera Mikros is a triple quadrupole LC-MS/MS configured as a microflow LC/MS system, operating at 1–500 µL/min with columns of 0.1–1 mm I.D. This modification is distinguished within the series by its microflow range, enabling sensitive analysis with reduced solvent consumption. It serves as the detection and quantification engine in micro-LC workflows, coupling narrow-bore separations to tandem mass spectrometry.
Direct drop-in replacements
Full specifications
| Nebulizing gas flow | 3 L/min |
| Flow rate range mfr | 1 to 500 μL/min (Nexera Mikros) |
| Flow rate (microflow analysis) mfr | 5 μL/min (microflow analysis) |
| Flow rate (semi-microflow analysis) mfr | 225 μL/min (semi-microflow analysis) |
| Column (LC-Mikros pump evaluation) mfr | Shim-pack MC C18 (0.3 mm I.D. × ...) (LC-Mikros pump evaluation) |
| LC flow rate range (micro-LC/MS operation) mfr | a few to a few tens μL/min (micro-LC/MS operation) |
| Test flow rate (low) (sensitivity/robustness evaluation) mfr | 2 μL/min (sensitivity/robustness evaluation) |
| Test flow rate (high) (sensitivity/robustness evaluation) mfr | 400 μL/min (sensitivity/robustness evaluation) |
| Ion intensity (500 pg/mL) (2 μL/min) mfr | 5,000 cps (2 μL/min) |
| Ion intensity (500 pg/mL) (400 μL/min) mfr | 1,500 cps (400 μL/min) |
| Ion intensity (50 pg/mL) (2 μL/min) mfr | 10,000 cps (2 μL/min) |
| Ion intensity (50 pg/mL) (400 μL/min) mfr | 500 cps (400 μL/min) |
| Retention time window (500 pg/mL / 50 pg/mL runs) mfr | 12.75–16.5 min (500 pg/mL / 50 pg/mL runs) |
| Retention time window (LTD4, 50 pg/mL, 5 μL injection) mfr | 11.5–13.5 min (LTD4, 50 pg/mL, 5 μL injection) |
| Ionization (negative ionization) mfr | Micro-ESI (negative ionization) |
| Drying gas mfr | Off |
| Dwell time mfr | 115 ms |
| MRM transition (quantifier), aldosterone (negative) mfr | 359.20 > 189.30 m/z (negative) |
| MRM transition (qualifier), aldosterone (negative) mfr | 359.20 > 331.35 m/z (negative) |
| MRM transition (quantifier), d4-aldosterone mfr | 363.00 > 190.20 m/z |
| Trapping column mfr | CERI C8 5 μm 5*0.3 mm |
| Trapping flow rate mfr | 100 μL/min |
| Microbore column mfr | Shimadzu PLONAS C18 2.7 μm 100*0.2 mm |
| Analytical flow rate mfr | 4 μL/min |
| Serum sample volume mfr | 10 μL |
| Calibration range mfr | 1 to 2 500 pg/mL |
| Lower limit of quantitation (LOQ) (QC, n=3 replicates) mfr | 1 pg/mL (QC, n=3 replicates) |
| Intra-day precision (%RSD) (real samples) mfr | < 8 % (real samples) |
| LOQ accuracy (QC LOQ 1 pg/mL) mfr | 103 (mean), 94.4 (min), 114 (max) % (QC LOQ 1 pg/mL) |
| LOQ precision (%RSD) (QC LOQ 1 pg/mL) mfr | 5.2 intra-day, 4.9 inter-day % (QC LOQ 1 pg/mL) |
| Matrix effect recovery mfr | 99 (QC), 103 (solution) % |
| Column inner diameter range mfr | 0.1 to 1 mm (Nexera Mikros) |
| Mass spectrometer system mfr | LCMS-8060 |
| Probe voltage (negative ionization) mfr | -2.4 kV (negative ionization) |
| Signal-to-noise ratio improvement (vitamin D metabolites, microflow vs semi-microflow) mfr | three- to five-fold (vitamin D metabolites, microflow vs semi-microflow) |
| Signal-to-noise ratio improvement (Verapamil, microflow vs semi-microflow) mfr | five-fold (Verapamil, microflow vs semi-microflow) |
| Signal-to-noise ratio improvement (Nor-Verapamil, microflow vs semi-microflow) mfr | four-fold (Nor-Verapamil, microflow vs semi-microflow) |
Description
Construction and positioning
- The Nexera Mikros is a triple quadrupole mass spectrometer integrated into a microflow LC/MS system. - It is defined by its compatibility with microflow rates (1–500 µL/min) and micro-bore columns (0.1–1 mm I.D.), positioning it for applications requiring high sensitivity and low sample consumption. ### Behavior under load and in different modes
- The system operates across a wide flow range, from 1 to 500 µL/min, accommodating both capillary and micro-LC scales. - Column internal diameters from 0.1 to 1 mm are supported, allowing flexibility in separation efficiency and loadability. - No digitized curve points or mode-specific performance data are available; operation should follow the manufacturer's specifications for flow and column dimensions. - The system requires appropriate LC plumbing and electrical connections as per standard triple quadrupole LC-MS/MS practice. - Users should refer to the manufacturer's manual for precise installation, connection, and control protocols.
Integration & Operating Notes
Install the system in a stable, temperature-controlled laboratory environment free from excessive vibration and dust. Adequate bench space and ventilation are required to accommodate the instrument and its accessories.
Selection & Integration Q&A
Choose the Nexera Mikros when your application requires microflow LC (1–500 µL/min) and narrow columns (0.1–1 mm I.D.) for high sensitivity and low sample consumption. For conventional flow rates, consider standard-flow models.
The LC system must deliver flow rates within 1–500 µL/min and use columns with 0.1–1 mm I.D. Ensure the plumbing and connections are compatible with microflow rates to avoid dead volumes and peak broadening.
Regular maintenance includes cleaning the ion source and checking the LC connections. Accessories such as columns, fittings, and calibration solutions are not included in the base delivery and must be ordered separately. Refer to the manual for the maintenance schedule.