Agilent Technologies Cary 3500 Flexible UV-Vis, Large Sample Compartment
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Cary 3500 Flexible UV-Vis, Large Sample Compartment

Agilent Technologies Cary 3500 Flexible UV-Vis, Large Sample Compartment — Molecular spectroscopy.

The Cary 3500 Flexible UV-Vis is a large-sample-compartment spectrophotometer for liquid and solid samples, distinct from the Compact and Multicell siblings. Its eight-position cell changer holds seven samples plus a reference, and the 1.5 mm focused beam suits 3.5 mL cuvettes. It serves as the series' open-architecture platform for variable-path-length and solid-sample work.

Agilent Technologies Cary Series — Model Comparison & Selection

Model Wavelength rangeFocused beam diameter, mmIntegrating sphere diameter, mmTemperature range, °C
Cary 3500 Flexible UV-Vis, Large Sample Compartment Current —1.5——
Cary 3500 —1.5—5...110
Cary 3500 Compact UV-Vis Spectrophotometer —1.5—5...110
Cary 4000 UV-Vis 175...900—150—
Cary 5000 UV-Vis-NIR spectrophotometer 175...900———
Cary 60 190...1,100———
Cary 6000i UV-Vis-NIR 175...900—150—

Full specifications

Core Performance & Functional Specifications
Cuvette Volume 3.5 mL
Focused Beam Diameter 1.5 mm
Number of cell positions mfr eight positions positions
Maximum samples plus reference mfr seven samples and a reference reference
Microcuvette measurement capacity mfr eight microcuvettes microcuvettes
Flow cell channels mfr three channels channels
Flow cells filled and rinsed simultaneously mfr three flow cells cells
Wavelength range shown mfr 250...600 nm
Dimensions, Weight & Mechanical Design
Path length mfr 1 cm

Description

Construction and series positioning

Measurement behavior

Installation and operation

Integration & Operating Notes

Mounting
Variable-path-length holder is a separate accessory

The large sample compartment is empty by default; path-length work requires the dedicated variable-path-length cuvette holder, and solid-sample work requires the corresponding solid-sample accessory. Confirm the holder is on the order before planning 1, 2, 5 or 10 cm measurements.

Technical Note: Cary 3500 UV-Vis Variable Path Length Cell Holder 5994-5781EN
Process Limits
Cuvette volume must match the 1.5 mm focused beam

The 1.5 mm beam diameter is matched to 3.5 mL rectangular cuvettes; larger 7, 17.5 and 35 mL cells and cylindrical cells are listed accessories. Using a cell whose window does not fully intercept the beam will bias absorbance readings.

Technical Note: Cary 3500 UV-Vis Variable Path Length Cell Holder 5994-5781EN
Control & I/O
Eight-position changer runs seven samples plus reference

Batch sequences can be queued across seven sample positions with one reference position for baseline correction, or eight microcuvettes. Plan sample batches around this eight-position limit; larger batches require reloading mid-sequence.

Agilent Cary 3500 UV-Vis Brochure 5994-0335EN
Maintenance
Flow-cell lines need rinse between three-channel runs

The three-channel flow configuration fills and rinses three flow cells simultaneously; residual sample in the lines after a run will contaminate the next measurement. Include a rinse cycle in the method and verify line cleanliness before switching media.

Agilent Cary 3500 UV-Vis Brochure 5994-0335EN

Selection & Integration Q&A

When should I choose the Cary 3500 Flexible UV-Vis instead of the Compact or Multicell variants?

Choose the Flexible when you need the large open compartment for variable-path-length cuvettes (1–10 cm) or solid samples. The Compact suits routine fixed-path liquid work in a small footprint; the Multicell adds multiple temperature zones for parallel thermal ramps. The Flexible trades enclosure for accessory breadth.

What accessories must be ordered alongside the Flexible to run 5 cm path-length measurements?

The variable-path-length cuvette holder is a separate accessory and is required for 1, 2, 5 or 10 cm work; the base instrument does not include it. Cuvettes are also not included: rectangular cells with PTFE lids are listed at 3.5, 7, 17.5 and 35 mL for 1, 2, 5 and 10 cm paths.

Can I run a three-channel flow application and a batch cuvette sequence on the same instrument?

The three-channel flow configuration fills and rinses three flow cells simultaneously, while the eight-position changer handles seven samples plus a reference or eight microcuvettes. These are separate sample-handling setups; switching between them requires changing the compartment configuration, so plan campaigns accordingly.

What maintenance is needed between flow-cell runs to avoid carryover?

The three-channel flow system fills and rinses three flow cells simultaneously; a rinse cycle must be included in the method after each sample set. Residual liquid in the lines will contaminate the next measurement, so verify line cleanliness before switching media or samples.