BD BF ED FD FED (E3.1) 07 2025
Technical Reference Summary
Covered Models & Document Scope
This operating manual (Issue 07/2025, Art. No. 7001-0292) covers the BINDER Avantgarde.Line E3.1 series: BD incubators with natural convection, BF incubators with forced convection, ED drying and heating ovens with natural convection, FD drying and heating ovens with forced convection, and FED drying and heating ovens with forced convection and enhanced timer functions. Models are listed in sizes 56, 115, 260 and 720, in 230 V, UL 120 V/208 V/240 V and 400 V variants with the article numbers given on page 1.
Section-by-Section Technical Breakdown
Describes the chamber overview, the triangular instrument panel and the main power switch on the rear of ED, FD and FED size 720 chambers. It identifies the controller and interface elements used during operation.
- Triangular instrument panel layout
- Main power switch on chamber rear for ED, FD, FED 720
- Chamber overview of interior and connections
Covers mounting of tilt protection holders on chambers with window, electrical connection, optional connection to an exhaust/ventilation system, and insertion of racks. Electrical connection requirements and supply voltage checks are referenced.
- Mounting tilt protection holders (chambers with window)
- Electrical connection and correct supply voltage
- Optional exhaust/ventilation system connection
- Inserting the racks
Describes chamber behavior when opening the door, performance during and after power failures, and loading of the chamber. These behaviors affect temperature recovery and operating procedure.
- Behavior when opening the door
- Performance during and after power failures
- Loading the chamber
Covers the R4 controller overview, normal display, menu language, date and time, temperature unit, set-point entry for temperature and fan speed, air flap position, and password changes. Includes special set-point entry for two-door chambers ED, FD and FED 720.
- Controller overview and normal display
- Setting menu language, date and time, temperature unit
- Set-point entry for temperature and fan speed
- Set-point entry for two-door chambers ED, FD, FED 720
- Adjusting air flap position
- Changing user and general controller passwords
Describes the overtemperature protective device class 1 and the safety controller, including setting the safety controller set-point, alarm messages and proceeding in case of an alarm, function check, and the optional disconnectable audible over-temperature alarm.
- Overtemperature protective device class 1
- Safety controller set-point setting
- Alarm message and proceeding in case of an alarm
- Function check
- Disconnectable audible over-temperature alarm (option)
Covers selection of timer functions including Delayed Off, Temperature dependent Delayed Off (BF, FED), and Delayed On (BF, FED), with entry and activation of run-time and fan setting, changing settings, and a temperature programming example for BF and FED.
- Selecting the timer function
- Delayed Off timer function
- Temperature dependent Delayed Off (BF, FED)
- Delayed On (BF, FED)
- Temperature programming example (BF, FED)
Describes general information on the ramp function, setting and displaying it, displaying the effective ramp set-point and changing the target ramp set-point, and turning off the ramp function.
- Setting and displaying the ramp function
- Displaying effective ramp set-point and changing target ramp set-point
- Turning off the ramp function
Covers starting and terminating data recording through the USB interface.
- Starting data recording
- Terminating data recording
Describes network configuration for chambers equipped with an Ethernet interface. Ethernet is standard with FED and optional with BD, BF, ED and FD.
- Ethernet interface configuration
- Ethernet standard with FED, optional with BD, BF, ED, FD
Lists options and accessories including APT-COM 4 Multi Management Software, object temperature display with additional Pt100 sensor, analog output for temperature, HEPA fresh air filter for FD and FED, mostly gas-tight version for BF, FD and FED, and inert gas connection with mostly gas-tight version.
- APT-COM 4 Multi Management Software
- Object temperature display with additional Pt100 temperature sensor
- Analog output for temperature
- HEPA fresh air filter (FD, FED)
- Mostly gas-tight version (BF, FD, FED)
- Inert gas connection with mostly gas-tight version (BF, FD, FED)
Covers cleaning and decontamination / chemical disinfection procedures for the chamber.
- Cleaning
- Decontamination / chemical disinfection
Covers general information and personnel qualification, maintenance intervals and service, simple troubleshooting, and sending the chamber back to BINDER GmbH. Includes annual maintenance and electrical testing requirements.
- Maintenance at least once a year
- Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702
- Simple troubleshooting by operating personnel
- Repair only by BINDER Service or qualified service partners
- Chamber must be tested after maintenance before resuming operation
Covers factory calibration and adjustment, definition of usable volume, over current protection, and technical data for BD, BF, ED, FD and FED models. Also includes equipment, options, accessories, spare parts and dimensional drawings for sizes 56, 115, 260 and 720.
- Factory calibration and adjustment
- Definition of usable volume
- Over current protection and chamber fuse
- Technical data BD, BF, ED, FD, FED
- Equipment, options, accessories and spare parts
- Dimensions size 56, 115, 260, 720
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal voltage tolerance is ±10% at the indicated power frequency (e.g. 230 V / 50 Hz or 230 V / 60 Hz). p.10 ↗ | Supply voltage must be checked against the type plate before connection; wrong voltage is a listed cause of chamber malfunction. | All BD, BF, ED, FD, FED models |
| FED 115-230V type plate example: nominal temperature 300 °C, nominal power 1.30 kW, nominal current 5.7 A, 230 V / 50 Hz or 230 V / 60 Hz, IP protection 20, safety device DIN 12880 class 2.0. p.9 ↗ | Electrical supply and overcurrent protection must be sized for 5.7 A at 230 V for this model. | FED 115-230V |
| FED 115-UL-120V type plate example: nominal temperature 300 °C, nominal power 1.40 kW, nominal current 11.7 A, 120 V / 50 Hz or 120 V / 60 Hz, IP protection 20, safety device DIN 12880 class 2.0. p.9 ↗ | UL 120 V variant draws 11.7 A and requires a correspondingly rated 120 V supply. | FED 115-UL-120V |
| The chamber must be maintained at least once a year. p.69 ↗ | Plan annual maintenance by licensed electricians or BINDER-authorized experts; warranty becomes void if maintenance is performed by non-authorized personnel. | All models |
| An annual repeat electrical inspection and a test prior to initial startup and prior to resuming operation after maintenance or repair are required; EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended. p.69 ↗ | Electrical safety testing must be scheduled annually and after any maintenance or repair. | All models |
| The door gasket must be replaced only when cold. p.69 ↗ | Replacing the door gasket while hot may damage it. | All models |
| ED, FD and FED size 720 chambers have a main power switch on the chamber rear. p.18 ↗ | For size 720, check that the main power switch is turned on when the chamber has no function. | ED 720, FD 720, FED 720 |
| If the overtemperature protective device class 1 has turned off the chamber, disconnect the chamber from the power supply for at least 10 seconds and let it cool down. p.70 ↗ | This is the specified reset procedure before contacting BINDER service if the device responds again. | All models |
| The chambers are not classified as medical devices under Regulation (EU) No 2017/745 and are not qualified to sterilize medical devices. p.14 ↗ | Do not use these chambers for medical device sterilization. | All models |
| Ethernet interface is standard with FED and optional with BD, BF, ED and FD. p.13 ↗ | Network configuration applies to FED as standard and to other series only when the Ethernet option is fitted. | FED standard; BD, BF, ED, FD optional |
| Connect only external devices to the Ethernet and USB interfaces that comply with EN 61010-1:2010 or EN 60950-1:2006 mod. p.13 ↗ | External equipment connected to the chamber interfaces must meet these safety standards. | All models with Ethernet or USB |
| The chambers are designed for indoor installation inside a building and must not be installed in unventilated recesses; prescribed minimum distances apply. p.11 ↗ | Ensure sufficient ventilation and observe wall clearances during installation. | All models |
| The chambers have no explosion protection measures; no combustible or explosive substance, dust or air-solvent mixture may be introduced, and the operating temperature must lie below the flash point or sublimation point of the charging material. p.12 ↗ | Process suitability must be verified against the material flash point before loading. | All models |
| The loading material must not contain corrosive ingredients such as acids and halides that may damage stainless steel, aluminum and copper components. p.13 ↗ | Corrosive damage from such ingredients is excluded from BINDER liability. | All models |
| The incubators BD and BF are especially useful for incubation of cultures at a standard temperature of 37 °C / 98.6 °F. p.13 ↗ | BD and BF are the intended series for culture incubation at 37 °C. | BD, BF |
| The chamber fuse is checked and replaced if appropriate per chapter 16.3; if it responds again, contact BINDER service. p.70 ↗ | A repeatedly blowing chamber fuse indicates a fault requiring BINDER service. | All models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber without function. p.70 ↗ | No power supply. | Check connection to power supply. ED, FD, FED size 720: Check whether the chamber is turned on at the main power switch. | Chapter 14.3 |
| Chamber without function. p.70 ↗ | Wrong voltage. | Check power supply for correct voltage (chap. 4.2). | Chapter 14.3 |
| Chamber without function. p.70 ↗ | Chamber fuse has responded. | Check chamber fuse and replace it if appropriate (chap. 16.3). If it responds again, contact BINDER service. | Chapter 14.3 |
| Chamber without function. p.70 ↗ | Defective controller. | Contact BINDER service. | Chapter 14.3 |
| Chamber without function. p.70 ↗ | Overtemperature protective device class 1 has turned off the chamber. | Disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, contact BINDER service. | Chapter 14.3 |
| Chamber without function, the standby icon is displayed. p.70 ↗ | Chamber in standby mode. | Press down the standby button until the display lights up. | Chapter 14.3 |
| Set-point temperature is not reached after specified time. p.70 ↗ | Chamber door not properly closed. | Completely close chamber door. | Chapter 14.3 |
| Set-point temperature is not reached after specified time. p.70 ↗ | Defective door gasket. | Replace door gasket. | Chapter 14.3 |
| Set-point temperature is not reached after specified time. p.70 ↗ | Controller not adjusted. | Calibrate and adjust controller. | Chapter 14.3 |
| Set-point temperature is not reached after specified time. p.70 ↗ | Wrong voltage. | Check the power supply for correct voltage (chap. 4.2). | Chapter 14.3 |
| Chamber heating permanently, set-point not held. p.70 ↗ | Defective controller. | Contact BINDER service. | Chapter 14.3 |
| Chamber heating permanently, set-point not held. p.70 ↗ | Defective Pt 100 sensor. | Contact BINDER service. | Chapter 14.3 |
| Chamber heating permanently, set-point not held. p.70 ↗ | Defective semiconductor relay. | Contact BINDER service. | Chapter 14.3 |
| Chamber heating permanently, set-point not held. p.70 ↗ | Controller not adjusted. | Calibrate and adjust controller. | Chapter 14.3 |