Operating Instructions — C-VLR 62 OXY vacuum pump (BA 882-31-OXY, ed. 1.0 27.09.2022, EN)

Technical Reference Summary

Covered Models & Document Scope

Covers the contactless operating claw vacuum pumps C-VLR 62 OXY and C-VLR 122 OXY with active cooling, including the XD variant with corrosion-protective coating for water-damp gases. Document BA 882-31-OXY-EN, edition 1.0, 27.09.2022.

Section-by-Section Technical Breakdown

pp. 14-16
Product overview and functioning

Describes the double-shaft rotary-piston claw vacuum pump with non-contacting, dry-running claw rotors synchronised by an oil-lubricated gear pair. The compression chamber is separated from the gearbox by special seals, and active cooling is provided by a drum fan drawing fresh air between compressor and insulation hood. The inlet has a flange with integrated non-return valve and mesh filter; the outlet has an exhaust silencer. OXY pumps are blue painted.

  • Vacuum connection G 1 1/2, exhaust air outlet G 1 1/2
  • Integrated non-return valve with mesh filter prevents ventilation of the evacuated system after stop
  • Depending on mechanical variant, a vacuum relief valve is integrated
  • Gearbox and bearings oil-lubricated; oil conveying device supplies all permissible speeds
  • Hot surfaces > 70 °C are marked
  • Condensate drain provided
pp. 15
Data plate

The data plate encodes model/size, serial number, year of construction, data matrix barcode, article number, speed 50/60 Hz, motor output 50/60 Hz, operating mode, suction capacity and final pressure (abs.). Example plate for C-VLR 62 OXY shows 100.00/100.00 mbar (abs), 62.0/74.0 m³/h, 1.50/1.80 kW, 1450/1745 rpm, S1 100%.

  • Barcode encodes material number (MA), production order (PR) and serial number (SC)
  • Data plate must not be removed even if machine is resold
  • Always quote serial number for queries
pp. 16
Fields of application

The claw vacuum pumps are suitable for sucking air or other gases with increased oxygen content (by volume above 21 %). They evacuate closed systems or provide continuous vacuum within an intake pressure range of 100–1000 mbar (abs). Maximum suction capacity with unrestricted suction is 62 m³/h or 120 m³/h at 50 Hz. The XD model has a corrosion-protective coating and is suitable for water-damp gases.

  • Intake pressure range for continuous vacuum: 100–1000 mbar (abs)
  • Max suction capacity 62 m³/h (C-VLR 62) or 120 m³/h (C-VLR 122) at 50 Hz
  • XD model suitable for conveying water-damp gases (water vapour)
  • Frequent switching (approx. 10 times per hour) or high ambient/inlet temperatures may exceed motor winding and bearing temperature limits — contact manufacturer
  • Observe motor protection class from motor data plate
  • If installed outdoors, protect from environmental influences (e.g. protective roof)
pp. 17-18
Installation

Installation requires authorised specialist personnel trained in oxygen plant safety (BGR 500). The pump is not gas-tight; sufficient ventilation must keep room oxygen content by volume ≤ 21 %. The machine must be set up horizontally with max inclination 8° in all directions, secured against tipping, on a level floor able to carry its weight. Cooling air inlets/outlets must be at least 30 cm from adjacent walls, and at least 50 cm space around the machine for maintenance.

  • Ambient temperature min. +5 °C, max. +40 °C
  • Max inclination 8° in all directions
  • Cooling air inlets/outlets at least 30 cm from walls; exhaust air must not be re-sucked
  • At least 50 cm space around machine for maintenance
  • Oil filling point, oil sight glass and oil drain must be easily accessible
  • Output reduction noticeable above 1000 m above sea level — contact manufacturer
  • If a non-return valve is integrated, it may not be removed
pp. 18-19
Connection of pipes

Pipes must be cleaned and free from organic substances before connection; the user must install an appropriate filter on the inlet side to protect the OXY pump. The exhaust air opening must not be closed or restricted, and maximum permissible back pressure is +50 mbar. Pipes must be screwed in by hand only; flexible connections or compensators are recommended to avoid distorting the piping system.

  • Maximum permissible back pressure: +50 mbar
  • Do not install blocking devices in exhaust line
  • Prevent liquids from accumulating in the exhaust line
  • Remove blind plugs from vacuum connection and exhaust air outlet before connecting
  • Only screw in pipes by hand; use flexible connections if necessary
  • Too narrow and/or too long inlet pipe reduces suction capacity
  • Regularly check connected exhaust lines for pollution
pp. 19, 21, 26-27
Lubricating oil

Only oil type OXY-GEAR-LUBE 160 is permitted for the OXY vacuum pumps. The oil level is checked through the oil sight glass and topped up if necessary. Using wrong lubricating oil can cause an acute danger of fire.

  • Only OXY-GEAR-LUBE 160 permitted
  • Check oil level through oil sight glass (Fig. 2/I)
  • If lubricating oil could be mixed with organic substances, it must be replaced immediately
pp. 19-20
Connect to the energy supply network

Electrical installation must be carried out by a qualified electrician observing EN 60204. The user must provide a main switch and a motor protection switch. Motors comply with DIN EN 60034, protection class IP 55 and insulation class F. Permissible tolerances are ±5 % voltage deviation and ±2 % frequency deviation. The direction of rotation must match the arrow on the motor flange.

  • Motor protection switch must be installed by operating company
  • Permissible voltage deviation: ±5 % of nominal value
  • Permissible frequency deviation: ±2 %
  • Motor protection class IP 55, insulation class F
  • Connection diagram located in terminal box (not present for plug-connection version)
  • Recommended: motor protection circuit breakers with delayed switch-off for temporary excess current at cold start
  • When using a frequency converter, observe its manufacturer's operating instructions
pp. 20
Pressure monitoring

The pump must not be operated without a pressure monitoring system. The pressure monitoring device is not included in the scope of delivery and must be installed by the user to ensure the minimum permissible pressure is not undershot. If the permissible pressure is undershot, the pump must be switched off to protect it from consequential damage.

  • Pressure monitoring system not included in scope of delivery
  • User must install a pressure monitoring device (e.g. pressure switch)
  • Permissible pressure for each pump is stated on the data plate
  • Pump must be switched off if permissible pressure is undershot
pp. 21-22
Start-up and operation

Before start-up, an installation check must confirm no transport damage, safe horizontal standing, correct and leak-tight pipe connections, tight screw/flange connections, compliant electrical installation, adequate room ventilation, correct oil level, and cleaned pump and pipelines. Direction of rotation must be checked with a phase sequence indicator; when looking at the motor blower it must rotate counter-clockwise. Condensate must be drained regularly from the silencer via the condensate drain.

  • Brief start max. two seconds to check rotation direction
  • Keep 1 m distance to outlet and inlet connections during rotation check
  • Before shutdown for more than 2 hours, run pump at 700 mbar (abs) for at least 15–30 minutes with dry air
  • Do not discharge condensate manually; allow machine to cool first
  • Wear protective gloves when draining condensate
  • Operation only allowed with sufficient cooling air quantity
pp. 23
Decommissioning and recommissioning

Decommissioning requires disconnecting the machine from power supply via main switch or plug and securing against unexpected restart. Electrical work must be done by an electrician only. Hot surfaces above 70 °C must be avoided. Condensate must be drained before decommissioning.

  • Disconnect from power supply and secure against unexpected restart
  • Avoid touching hot surfaces marked by warning signs
  • Drain condensate before decommissioning
  • Works on electrical installation only by electrician
pp. 24-30
Maintenance and repair

Maintenance includes cleaning the vacuum pump, checking and changing lubricating oil, replacing the air filter, and servicing motor and coupling. A maintenance table specifies intervals. Repairs are only allowed by authorised workshops using original spare parts.

  • Maintenance table defines required intervals
  • Oil change and refilling procedures specified
  • Air filter must be replaced when polluted
  • Repairs only by authorised workshops using original spare parts
  • Spare parts list E 882-31-OXY available
pp. 15, 33
Technical data

Technical data section refers to data sheet D 882-31-OXY for technical data and characteristic curves. Data plate example gives final pressure 100 mbar (abs), suction capacity 62.0/74.0 m³/h, motor output 1.50/1.80 kW, speed 1450/1745 rpm, operating mode S1 100%.

  • Data sheet D 882-31-OXY contains technical data and characteristic curves
  • Example data plate: final pressure 100.00/100.00 mbar (abs)
  • Example data plate: suction capacity 62.0/74.0 m³/h
  • Example data plate: motor output 1.50/1.80 kW
  • Example data plate: speed 1450/1745 rpm
  • Operating mode S1 100%

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Maximum permissible back pressure: +50 mbar p.8 ↗ Exhaust line must not be restricted or blocked; higher process back pressures require consultation with manufacturer. C-VLR 62 OXY and C-VLR 122 OXY
Ambient and inlet temperature range: 5 to 40 °C p.8 ↗ Operation outside this range requires manufacturer consultation; cooling and motor limits depend on it. C-VLR 62 OXY and C-VLR 122 OXY
Only oil type OXY-GEAR-LUBE 160 is permitted p.19 ↗ Wrong oil can cause acute fire risk; oil must be replaced immediately if contamination with organic substances is suspected. C-VLR 62 OXY and C-VLR 122 OXY
Permissible voltage deviation: ±5 % of nominal value; frequency deviation: ±2 % p.20 ↗ Mains supply must match motor data plate within these tolerances to avoid loss of power or damage. C-VLR 62 OXY and C-VLR 122 OXY
Motor protection class IP 55, insulation class F, compliant with DIN EN 60034 p.19 ↗ Suitable for dusty/damp industrial environments; electrical installation must observe EN 60204. C-VLR 62 OXY and C-VLR 122 OXY
Max inclination during operation: 8° in all directions p.17 ↗ Machine must be set up horizontally and secured against tipping; floor must be level and able to carry weight. C-VLR 62 OXY and C-VLR 122 OXY
Cooling air inlets and outlets must be at least 30 cm from adjacent walls p.18 ↗ Prevents overheating; exhaust air must not be re-sucked into cooling system. C-VLR 62 OXY and C-VLR 122 OXY
At least 50 cm space around machine for maintenance p.17 ↗ Installation layout must allow access to oil filling point, oil sight glass, oil drain and maintenance areas. C-VLR 62 OXY and C-VLR 122 OXY
Intake pressure range for continuous vacuum: 100–1000 mbar (abs) p.16 ↗ Pump is suitable for evacuation of closed systems or continuous vacuum within this range. C-VLR 62 OXY and C-VLR 122 OXY
Maximum suction capacity with unrestricted suction: 62 m³/h (C-VLR 62) or 120 m³/h (C-VLR 122) at 50 Hz p.16 ↗ Selection must account for inlet pipe pressure drop; too narrow or too long inlet pipe reduces suction capacity. C-VLR 62 OXY and C-VLR 122 OXY
Example data plate C-VLR 62 OXY: final pressure 100.00/100.00 mbar (abs), suction capacity 62.0/74.0 m³/h, motor output 1.50/1.80 kW, speed 1450/1745 rpm, S1 100% p.15 ↗ Reference values for the 62 size at 50/60 Hz; actual values per data plate and data sheet D 882-31-OXY. C-VLR 62 OXY
Direction of rotation check: when looking at the motor blower, it must rotate counter-clockwise; brief start max. two seconds p.22 ↗ Wrong rotation can damage machine and cause injury; use phase sequence indicator and keep 1 m distance from connections. C-VLR 62 OXY and C-VLR 122 OXY
Before shutdown for more than 2 hours, run pump at 700 mbar (abs) for at least 15–30 minutes with dry air p.22 ↗ Removes moisture and pollution to prevent deposits that can block restart. C-VLR 62 OXY and C-VLR 122 OXY
Pressure monitoring system must be installed by user; pump must not be operated without it p.20 ↗ Minimum permissible pressure must not be undershot; pump must be switched off if it is. C-VLR 62 OXY and C-VLR 122 OXY
Storage temperature: -10 °C to +60 °C; relative air humidity max. 80 %; storage not more than 6 months p.13 ↗ Store dust-free, dry, vibration-free, protected from sun radiation, with openings closed air-tight. C-VLR 62 OXY and C-VLR 122 OXY
Output reduction noticeable when installed at more than 1000 m above sea level p.18 ↗ Contact manufacturer for high-altitude installations. C-VLR 62 OXY and C-VLR 122 OXY
Oxygen content by volume in room air may not exceed 21 % p.17 ↗ Installation room must be sufficiently ventilated because the OXY pump is not gas-tight and can enrich room air with oxygen. C-VLR 62 OXY and C-VLR 122 OXY
Vacuum connection and exhaust air outlet: G 1 1/2 p.14 ↗ Pipes must be screwed in by hand only; flexible connections recommended to avoid excessive forces and torques. C-VLR 62 OXY and C-VLR 122 OXY

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Machine is switched off by the motor protection switch p.31 ↗ Mains voltage/ Frequency does not correspond with the motor data Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.31 ↗ Connection to motor terminal board is not correct Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.31 ↗ Motor protection switch is not set correctly Check by qualified electrician Chapter 5.5
Motor protection switch is triggered too quickly p.31 ↗ Short excess current at start up Use a motor protection switch with an overload-dependent turn-off delay that considers the short excess current at start up (version with short circuit and overload trigger as per VDE 0660 Part 102 or IEC 60947-4-1)
Suction capacity is insufficient p.31 ↗ Air filter is polluted Replace air filter Chapter 7.6
Suction capacity is insufficient p.31 ↗ The inlet pipe is too long or too narrow Check the hose and/or the pipe Chapter 5.3
Suction capacity is insufficient p.31 ↗ Machine or system leaking Check the pipework and screw connections for leaks and check for tight fit Chapter 5.3
Final pressure (max. vacuum) is not reached p.31 ↗ Machine or system leaking Check the pipework and screw connections for leaks and check for tight fit Chapter 5.3
Final pressure (max. vacuum) is not reached p.31 ↗ Air filter is polluted Replace air filter Chapter 7.6
Machine gets too hot p.31 ↗ Ambient or inlet temperatures too high Ensure proper use Chapter 2.4
Machine gets too hot p.31 ↗ Cooling air supply is obstructed Check ambient conditions Chapter 5.1
Machine gets too hot p.31 ↗ Cooling air supply is obstructed Clean ventilation slots Chapter 7.4
The machine makes a strange noise p.31 ↗ Deposits on the rotary piston Clean the working space and the rotary piston Elmo Rietschle Service