Operating Instructions — C-VLR 62 vacuum pump (BA 882-PC, ed. 1.0 28.03.2019, EN)
Technical Reference Summary
Covered Models & Document Scope
Covers the contactless claw vacuum pumps C-VLR 62 and C-VLR 122 with passive cooling, including the standard model (CD) and the corrosion-protective coated model (XD). The document is the operating instructions BA 882-PC-GB, edition 1.0 dated 28.03.2019.
Section-by-Section Technical Breakdown
Describes the construction and operating principle of the C-VLR 62 and C-VLR 122 claw vacuum pumps. The pumps are double-shaft rotary piston machines with contact-free dry-running claw rotors synchronised by an oil-lubricated gear pair. The document identifies components, connections, data plate fields, fields of application and available accessories.
- C-VLR 62: vacuum connection G 1 1/2, exhaust air outlet, inlet flange with non-return valve, cooling air inlet/outlet, oil filling point, oil sight glass, oil discharge point, exhaust silencer, vacuum relief valve (ZUV).
- C-VLR 122: same as C-VLR 62 plus exhaust air outlet G 1 and condensate drain X.
- Data plate includes type/size, serial number, year of construction, data matrix barcode, item number, speed 50/60 Hz, motor output 50/60 Hz, recommended oil type, operating mode, pumping capacity 50/60 Hz, final pressure (abs.) 50/60 Hz.
- Fields of application: evacuation of closed systems or continuous vacuum within inlet pressure range 400–1000 mbar (abs.); maximum suction capacity 62 m3/h or 120 m3/h at 50 Hz.
- XD model has corrosion-protective coating and is suitable for water-damp gases (water vapour).
- Accessories: intake filter (paper or polyester) and vacuum relief valve (ZUV) set at factory for lowest permissible final pressure.
Provides requirements for preparing the installation site, mechanical installation, pipe connection, oil level check and electrical connection. Includes limits for inclination, cooling air clearance, back pressure and electrical tolerances.
- Machine must be set up horizontally; maximum inclination 8° in all directions.
- Cooling air inlets and outlets must be at least 30 cm away from adjacent walls.
- Maximum permissible back pressure: +50 mbar.
- Only screw in pipes by hand; use flexible connections if necessary.
- Oil level must be checked through the oil sight glass and topped up to the middle of the sight glass.
- Electrical installation by qualified electrician per EN 60204; motor protection circuit breaker required.
- Permissible voltage deviation ±5%, frequency deviation ±2%.
- Motors comply with DIN EN 60034, protection class IP 55, insulation class F.
Covers start-up checks, rotation direction verification, after-run procedure, decommissioning steps and recommissioning. Includes safety warnings for hot surfaces, suction hazards and noise.
- Check rotation direction with phase sequence indicator; when looking at motor blower, it must rotate counter-clockwise.
- Start motor briefly (max. two seconds) to check direction of rotation.
- Before putting out of operation for more than 2 hours, run pump at -700 mbar (abs.) for at least 15–30 minutes with dry air.
- Drain condensate from silencer (VLR 122) before decommissioning.
- Decommissioning: switch off, allow to cool, close cut-off devices, disconnect power, depressurise slowly, remove pipes, drain condensate, seal connections.
Defines maintenance intervals and procedures for oil change, air filter cleaning/replacement and coupling inspection. Includes safety instructions and spare parts ordering information.
- Monthly: check pipes and screws for leaks, clean ventilation slots, check oil level, check/clean/replace air filters, drain condensate (VLR 122).
- Oil change every 20,000 operating hours; extreme load can reduce oil life to 5,000 hours.
- Oil type: Elmo Rietschle GEAR-LUBE 150 synthetic oil, viscosity ISO VG 150 per DIN 3448.
- Check couplings for wear at least once per year.
- General overhaul at 40,000 hours by Elmo Rietschle Service.
- Mesh filter: clean or replace monthly by washing or blowing off.
- Intake filter cartridge: clean monthly, replace every six months.
Troubleshooting table listing malfunctions, probable causes and corrective actions. Covers motor protection switch trips, insufficient suction capacity, failure to reach final pressure, overheating and unusual noise.
- Machine switched off by motor protection switch: check mains voltage/frequency, motor terminal connection, motor protection switch setting.
- Suction capacity insufficient: clean/replace air filter, check inlet pipe length/diameter, check for leaks.
- Final pressure not reached: check for leaks, clean/replace air filter.
- Machine gets too hot: check ambient/inlet temperatures, cooling air supply, clean ventilation slots.
- Strange noise: clean working space and rotary piston; contact Elmo Rietschle Service.
Lists technical specifications for C-VLR 62 and C-VLR 122, including sound pressure level, weight, dimensions, connections, oil volume, voltage, current and motor output.
- C-VLR 62: sound pressure level 79 dB(A) at 50 Hz, 82 dB(A) at 60 Hz; sound power level 85 dB(A).
- C-VLR 62: weight 70 kg, length 707 mm, width 439 mm, height 277 mm.
- C-VLR 122: weight 90 kg, length 776 mm, width 378 mm, height 277 mm.
- Vacuum connection G 1 1/2; oil filled volume 0.43 l.
- Rated voltage 3~: 230/400 V ±10% at 50 Hz; 265/460 V ±10% at 60 Hz.
- Current consumption C-VLR 62: 4.89/2.81 A at 50 Hz; 4.84/2.79 A at 60 Hz.
- Motor output C-VLR 62: 1.25 kW at 50 Hz; 1.5 kW at 60 Hz.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Maximum inclination for operation: 8° in all directions. p.16 ↗ | Machine must be installed horizontally; ensure floor is plane and even. | C-VLR 62 and C-VLR 122 |
| Cooling air inlets and outlets must be at least 30 cm away from adjacent walls. p.16 ↗ | Ensure sufficient clearance for cooling air flow to prevent overheating. | C-VLR 62 and C-VLR 122 |
| Maximum permissible back pressure: +50 mbar. p.17 ↗ | Do not close or restrict exhaust air openings; prevent liquid accumulation in exhaust line. | C-VLR 62 and C-VLR 122 |
| Permissible voltage deviation: ±5%; frequency deviation: ±2%. p.18 ↗ | Verify mains supply matches motor data plate before connection. | C-VLR 62 and C-VLR 122 |
| Motor protection: IP 55, insulation class F, per DIN EN 60034. p.18 ↗ | Use motor protection circuit breaker with delayed switch off to accommodate temporary excess current at cold start. | C-VLR 62 and C-VLR 122 |
| Oil change interval: 20,000 operating hours; extreme load can reduce to 5,000 hours. p.23 ↗ | Use Elmo Rietschle approved oil (GEAR-LUBE 150, ISO VG 150) to maintain interval. | C-VLR 62 and C-VLR 122 |
| Oil filled volume: 0.43 l. p.31 ↗ | Fill oil to middle of sight glass; if changing oil type, empty tank completely. | C-VLR 62 and C-VLR 122 |
| Recommended oil: GEAR-LUBE 150 synthetic oil, viscosity ISO VG 150 per DIN 3448. p.25 ↗ | Use only approved oil to maintain warranty and performance. | C-VLR 62 and C-VLR 122 |
| After-run procedure: run at -700 mbar (abs.) for at least 15–30 minutes with dry air before shutdown >2 hours. p.20 ↗ | Prevents condensate and impurity deposits that can damage machine. | C-VLR 62 and C-VLR 122 |
| Rotation direction check: start motor briefly (max. 2 seconds); when looking at motor blower, it must rotate counter-clockwise. p.20 ↗ | Wrong rotation can damage machine and cause injury; use phase sequence indicator. | C-VLR 62 and C-VLR 122 |
| Maximum suction capacity: 62 m3/h or 120 m3/h at 50 Hz. p.15 ↗ | Select model based on required pumping speed; inlet pressure range 400–1000 mbar (abs.). | C-VLR 62 (62 m3/h) and C-VLR 122 (120 m3/h) |
| Sound pressure level: 79 dB(A) at 50 Hz, 82 dB(A) at 60 Hz (C-VLR 62); tolerance ±3 dB(A). p.31 ↗ | Use ear protection when spending long time near running machine. | C-VLR 62 |
| Weight: C-VLR 62 = 70 kg; C-VLR 122 = 90 kg. p.31 ↗ | Ensure bearing surface can carry weight; secure machine against tipping. | C-VLR 62 and C-VLR 122 |
| Rated voltage 3~: 230/400 V ±10% at 50 Hz; 265/460 V ±10% at 60 Hz. p.31 ↗ | Verify motor data plate matches mains network. | C-VLR 62 |
| Motor output: 1.25 kW at 50 Hz; 1.5 kW at 60 Hz (C-VLR 62). p.31 ↗ | Size motor protection and supply accordingly. | C-VLR 62 |
| Ambient and inlet temperature range: 5 to 40 °C. p.7 ↗ | Contact manufacturer for temperatures outside this range. | C-VLR 62 and C-VLR 122 |
| Storage conditions: relative humidity 0–80%, temperature -10 to +60 °C, max. 6 months. p.11 ↗ | Store in dry, dust-free, vibration-free environment. | C-VLR 62 and C-VLR 122 |
| Coupling sprocket must be checked for wear at least once per year. p.28 ↗ | Frequent starting and high ambient temperature reduce sprocket lifetime. | C-VLR 62 and C-VLR 122 |
| Intake filter cartridge should be replaced every six months. p.27 ↗ | Clean monthly by blowing off from inside to outside; separation efficiency deteriorates with cleaning. | C-VLR 62 and C-VLR 122 with intake filter accessory |
| General overhaul at 40,000 hours by Elmo Rietschle Service. p.23 ↗ | Plan for service intervention at this interval. | C-VLR 62 and C-VLR 122 |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| The machine is switched off by the motor protection switch p.30 ↗ | Mains voltage/ Frequency does not correspond with the motor data | Get it checked by an electrician | Chapter 5.5 |
| The machine is switched off by the motor protection switch p.30 ↗ | Connection to motor terminal board is not correct | Get it checked by an electrician | Chapter 5.5 |
| The machine is switched off by the motor protection switch p.30 ↗ | Motor protection switch is not set correctly | Get it checked by an electrician | Chapter 5.5 |
| Motor protection switch is triggered too quickly p.30 ↗ | Motor protection switch is not set correctly | Use a motor protection switch with an overload-dependent turn-off delay that considers the short excess current at start up (version with short circuit and overload trigger as per VDE 0660 Part 102 or IEC 60947-4-1) | |
| Suction capacity is insufficient p.30 ↗ | Air filter is polluted | Clean / replace the air filter | Chapter 7.2.2, Chapter 7.4 |
| Suction capacity is insufficient p.30 ↗ | The inlet pipe is too long or too narrow | Check the hose and/or the pipe | Chapter 5.3 |
| Suction capacity is insufficient p.30 ↗ | Machine or system leaking | Check the pipework and screw connections for leaks and check for tight fit | Chapter 7.2 |
| Final pressure (max. vacuum) is not reached p.30 ↗ | Machine or system leaking | Check the pipework and screw connections for leaks and check for tight fit | Chapter 7.2 |
| Final pressure (max. vacuum) is not reached p.30 ↗ | Air filter is polluted | Clean / replace the air filter | Chapter 7.2.2, Chapter 7.4 |
| Machine gets too hot p.30 ↗ | Ambient or inlet temperatures too high | Ensure proper use | Chapter 2.3 |
| Machine gets too hot p.30 ↗ | Cooling air supply is obstructed | Check ambient conditions | Chapter 5.1 |
| Machine gets too hot p.30 ↗ | Cooling air supply is obstructed | Clean ventilation slots | Chapter 7.2 |
| The machine makes a strange noise p.30 ↗ | Deposits on the rotary piston | Clean the working space and the rotary piston | Elmo Rietschle Service |