Agilent Technologies Cary 3500
Agilent Technologies Cary 3500 — Molecular spectroscopy.
The Cary 3500 Multicell UV-Vis is a modular spectrophotometer built from a Xenon-flash dual-monochromator engine and an eight-position Multicell module (seven samples plus reference). It is the multicell modification of the Cary 3500 series, distinguished by simultaneous multi-cuvette measurement; Peltier variants add four independently controlled temperature zones from -5 to 110 °C. It serves high-throughput absorbance and thermal-melting workflows in laboratory molecular spectroscopy.
Agilent Technologies Cary Series — Model Comparison & Selection
| Model | Wavelength range | Focused beam diameter, mm | Integrating sphere diameter, mm | Temperature range, °C |
|---|---|---|---|---|
| Cary 3500 Current | — | 1.5 | — | 5...110 |
| Cary 3500 Compact UV-Vis Spectrophotometer | — | 1.5 | — | 5...110 |
| Cary 3500 Flexible UV-Vis, Large Sample Compartment | — | 1.5 | — | — |
| Cary 4000 UV-Vis | 175...900 | — | 150 | — |
| Cary 5000 UV-Vis-NIR spectrophotometer | 175...900 | — | — | — |
| Cary 60 | 190...1,100 | — | — | — |
| Cary 6000i UV-Vis-NIR | 175...900 | — | 150 | — |
Full specifications
| Focused Beam Diameter | 1.5 mm |
| Memory | 16 RAM |
| Operating system | 10 bit |
| Temperature range | 5...110 °C |
| Number of cell positions mfr | eight positions positions |
| Maximum samples plus reference mfr | seven samples and a reference reference |
| Microcuvette measurement capacity mfr | eight microcuvettes microcuvettes |
| Flow cell channels mfr | three channels channels |
| Flow cells filled and rinsed simultaneously mfr | three three |
| Measurement category mfr | 61010 I |
| Clearance for Peltier models mfr | 20 cm |
| Rear clearance for Peltier models mfr | 10 cm |
| Clearance for non-Peltier models mfr | 10 cm |
| Rear clearance for non-Peltier models mfr | 10 cm |
| Processor mfr | 5 cache |
| Storage mfr | 512 Drive |
| Graphics mfr | 770 |
| Communications mfr | 2 NIC |
| Nitrogen water vapor mfr | 2570 ppm |
| Compressed air purity mfr | 8573 |
| Purge kit maximum pressure mfr | 689 kpa |
| Purge gas flow rate mfr | 15 LPM |
| Drain vessel minimum volume mfr | 2 liters |
| Purge tubing dimensions mfr | 6 ID |
| Magnet safe distance mfr | 145 mm |
| Cell positions mfr | eight eight |
| Samples plus reference mfr | seven samples and a reference reference |
| Flow cell pump channels mfr | three three |
| Nitrogen purge gas water vapor mfr | 2570 ppm |
| Air purge gas quality mfr | 8573 |
| Minimum drain vessel volume mfr | 2 liters |
| data acquisition rate mfr | 250 second |
| wavelength drive speed mfr | 2500 second |
| spectral bandwidth mfr | 0.1...5 nm |
| G9862AA mfr | 3500 Vis |
| G9866AA mfr | 3500 Vis |
| G9867AA mfr | 3500 Vis |
| Lamp module mfr | 9864 |
| Magnetic stirrer bar PTFE star type mfr | 7418000400 |
| Flow meter mfr | 8003 |
| Purge tubing mfr | 3710043100 |
| G9894A mfr | Cary Compact Peltier UV-Vis Module Module |
| G9889A mfr | Cary Multicell Peltier UV-Vis Module Module |
| Peltier side clearance mfr | 20 cm |
| Peltier rear clearance mfr | 10 cm |
| Non-Peltier side clearance mfr | 10 cm |
| Non-Peltier rear clearance mfr | 10 cm |
| Drain vessel minimum capacity mfr | 2 liters |
| Drain outlet tubing OD mfr | 6 OD |
| Workbench height mfr | 90 cm |
| Drain tubing outer diameter mfr | 6 OD |
| path length options mfr | 2 cm |
| G9871A mfr | 3500 A |
| G9873A mfr | 3500 A |
| G9875A mfr | 3500 A |
| Pollution degree mfr | 61010 |
| Cuvette cooling time mfr | five minutes minutes |
| lamp replacement warranty mfr | 10 year |
Description
Architecture and series positioning
- The system separates a Cary engine (Xenon flash lamp, compact dual out-of-plane monochromator) from interchangeable sample modules; the Multicell module carries eight cell positions, usable as seven samples plus a reference or as eight microcuvettes.
- Within the series, the Compact module offers two channels (sample and reference/blank) in one zone, the Flexible module provides a large compartment for liquid and solid samples, and the Multicell/Multizone Peltier modules provide the multi-station, multi-zone temperature capability described here.
- Peltier Multicell configurations control the eight stations in four zones (positions 1/2, 3/4, etc.), each zone variable over -5 to 110 °C; ambient Multicell configurations have no thermal control.
Behavior and measurement modes
- The focused beam diameter is 1.5 mm, matching microcuvette work; the eight-position format allows sequential or simultaneous multi-sample acquisition against a single reference.
- Thermal workflows are supported on Peltier variants: the documented melting-temperature example gives Tm 86.5...87.1 °C, and the temperature ramp field in the source data is incomplete, so no ramp rate is stated here.
- Flow measurements use three flow-cell channels, with three flow cells filled and rinsed simultaneously.
Installation, connection and operation
- The system is designed for customer installation: software installation, hardware installation (engine and module), then System Health tests, following the Cary UV-Vis Installation Video and Help and Learning Center.
- Peltier modules require purge gas to prevent condensation below ambient and to protect optics: 5 LPM for Compact Peltier, 15 LPM for Multicell Peltier and Multizone Peltier, through the G9891A purge kit (maximum 689 kPa, 0-30 LPM flow meter, 5 m of 6 mm OD/4 mm ID tubing).
- Multicell and Compact modules drain spills through a 6 mm OD outlet into a user-supplied chemically inert vessel of at least 2 liters; the Flexible module instead directs spills onto the bench.
Integration & Operating Notes
Multicell Peltier and Multizone Peltier systems require 15 LPM purge flow, three times the Compact Peltier figure, so the regulator and flow meter must be sized for that rate. Purging prevents condensation in the sample compartment at temperatures below ambient and keeps vapors and particles off optical surfaces.
Gas must pass through a pressure regulator; the G9891A purge kit is rated to a maximum of 689 kPa (100 PSIG). Use clean flexible plastic (PVC or equivalent) 6 mm OD/4 mm ID tubing, never rubber, because rubber tubing may be internally treated with talc that can be blown into the optical system. Nitrogen must have water vapor below 2570 ppm; compressed air must meet ISO 8573-1:2010 CLASS 1.3.1.
Peltier models require 20 cm on both sides and about 10 cm at the rear for electrical and gas connections and for air flow; insufficient clearance impairs air-cooled Peltier control, especially at low temperatures. Non-Peltier models need 10 cm on both sides and 10 cm at the rear. Recommended ambient is 20 to 25 °C held within ±2 °C through the working day.
The sample compartment and cuvettes may become extremely hot during Peltier operation and remain hot after the setpoint is lowered; allow at least five minutes to cool before removing cuvettes. Plan access to the eight positions around this cool-down so that multi-zone runs are not interrupted by handling hot cells.
Spills in the Multicell sample compartment drain through a 6 mm OD outlet into a user-supplied chemically inert collection vessel of at least 2 liters (4 pints); glass or narrow-necked vessels are not suitable. Check the tubing and vessel regularly for kinks, blockages or a full vessel, and select materials compatible with the solvents in use.
A 3-wire outlet grounded at the grounding pin must be provided, with a separate receptacle for each unit; extension cords and outlet adaptors are not recommended. The instrument is rated IEC 61010 measurement category I and must not be connected to measurement categories II, III or IV circuits. Disconnect mains before fuse replacement and use only the rating marked on the instrument.
The controlling computer requires an Intel i5 3.0 GHz with 18 MB cache, 16 GB RAM, 512 GB SSD, Intel UHD Graphics 770 and Windows 10/11 Professional or Enterprise 64-bit. Agilent representatives install Workstation Plus or Networked Workstation software; the engine and module connect through the communications port, and the module front LED indicates status by color and flashing pattern.
The lamp module operates at high voltage; switch off and disconnect the engine from mains before replacement, and never disassemble the module. All other screwed panels require Agilent service engineers. Troubleshooting and maintenance procedures are hosted in the Cary UV Help and Learning Center, and spare parts are ordered per the manual's spare-parts section.
Selection & Integration Q&A
Choose Multicell for eight cuvette stations (seven samples plus reference, or eight microcuvettes) and, in Peltier form, four zones each settable from -5 to 110 °C. The Compact module gives only two channels in one zone; the Flexible module is for large liquid and solid samples, not multi-station throughput.
A regulated purge supply at 15 LPM (nitrogen below 2570 ppm water vapor, or ISO 8573-1:2010 CLASS 1.3.1 air) through the G9891A kit at no more than 689 kPa, 6 mm OD/4 mm ID plastic tubing, a 2-liter inert drain vessel, 20 cm side and 10 cm rear clearance, and a dedicated earth-grounded outlet per unit.
Yes; the Cary system is designed to be completely customer-installable: install the software, install the engine and module hardware, then run System Health tests. Site it at 20 to 25 °C held within ±2 °C, observing the Peltier clearance of 20 cm at the sides and 10 cm at the rear, and follow the installation video and Help and Learning Center.
Users supply purge tubing and regulator, the 2-liter drain vessel and drain tubing, and the control PC meeting the i5/16 GB/512 GB specification. Fuse replacement is user-performed with the marked rating; lamp module replacement is the only other user-accessible service, and all further maintenance is directed to the Cary UV Help and Learning Center or Agilent engineers.