Agilent Technologies Cary 4000 UV-Vis
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Cary 4000 UV-Vis

Agilent Technologies Cary 4000 UV-Vis — Molecular spectroscopy.

The Cary 4000 UV-Vis is a double-beam spectrophotometer with an out-of-plane double Littrow monochromator and silica-overcoated optics on a floating solid aluminium casting, covering 175–900 nm. It is the UV-Vis-only member of the Cary 4000/5000/6000i series, distinguished from the 5000 (to 3300 nm) and 6000i (to 1800 nm, InGaAs NIR) by its fixed slits and absence of NIR capability. It serves routine UV-Vis absorbance, transmittance and validation measurements in analytical laboratories.

Agilent Technologies Cary Series — Model Comparison & Selection

Model Wavelength rangeFocused beam diameter, mmIntegrating sphere diameter, mmTemperature range, °C
Cary 4000 UV-Vis Current 175...900—150—
Cary 3500 —1.5—5...110
Cary 3500 Compact UV-Vis Spectrophotometer —1.5—5...110
Cary 3500 Flexible UV-Vis, Large Sample Compartment —1.5——
Cary 5000 UV-Vis-NIR spectrophotometer 175...900———
Cary 60 190...1,100———
Cary 6000i UV-Vis-NIR 175...900—150—

Full specifications

Core Performance & Functional Specifications
Wavelength range 175...900 nm
Integrating Sphere Diameter 150 mm
Minimum Side Clearance 50 mm
Warm-up Time 2 hours
Optimized NIR region mfr 1200...1800 nm
NIR detector mfr InGaAs InGaAs
Optical design mfr PbSmart PbSmart
Monochromator design mfr out-of-plane double Littrow Littrow
Optics mfr silica overcoated overcoated
Optical isolation mfr floating solid aluminium casting casting
Lamp replacement mfr pre-aligned, plug-and-go go
Slits mfr fixed, in NIR as well as UV-Vis Vis
Minimum rear clearance mfr 150 mm
Nitrogen operating pressure mfr 83...172 kPa
Nitrogen supply tubing internal diameter mfr 1 inch
Verifying Safe State mfr 5
Ultraviolet radiation mfr 6
Electrical hazards mfr 6
Other precautions mfr 7
Information symbols mfr 8
Color coding mfr 9
Starting the software mfr 13
Software Status Bulletin mfr 14
Setting up the instrument mfr 15
Turning on the system mfr 16
Instrument performance tests mfr 17
Sample holders mfr 17
Installing the base plate mfr 25
Installing other accessories mfr 25
Nitrogen purging mfr 26
Removing the sample compartment base mfr 28
Cleaning mfr 31
Source lamps mfr 32
Fuses mfr 32
Spare Parts mfr 33
Rear clearance mfr 150 mm
NIR grating mfr 600 lines/mm
internal diffuse reflectance accessory range mfr 110 mm
external diffuse reflectance accessory range mfr 150 mm
Electrical, Power & Drive Specifications
Sample compartment high voltage pin mfr 125 VDC
Interfaces, Connections & Control
GPIB communications mfr 14

Description

Construction and series positioning

Operation and media

Installation, connection and control

Integration & Operating Notes

Mounting
Bench clearance is required for air circulation

Allow at least 50 mm (2 in.) on the sides and 150 mm (6 in.) at the rear of the system for free air circulation. Insufficient clearance can cause overheating and impaired performance, so bench space must be planned before installation.

User Manual, Setting up the instrument, page 15
Electrical
Sample compartment connector carries high voltage

The 15-pin D-range connector in the sample compartment has high-voltage pins rated −1000 VDC and +125 VDC on the Cary 4000/5000/7000. Accessory compatibility and safety planning must account for these voltages, and one connector is reserved for Agilent Internal and External DRAs only.

User Manual, Electrical Connections, page 12; Installing other accessories, page 25
Process Limits
Purging requires regulated dry, oil-free nitrogen

Nitrogen purging must be held at 83–172 kPa (12–25 psig) with instrument flow 0–20 L/min and sample compartment flow 0–10 L/min. Use clean flexible plastic (PVC or equivalent) tubing of 6 mm (1/4 in.) internal diameter; never use rubber tubing, and never use compressed nitrogen from a supplier who uses oil or water in the compression process.

User Manual, Nitrogen purging, pages 26–28
Control & I/O
GPIB converter must be installed after Cary WinUV

The computer interfaces through a National Instruments USB-GPIB-HS+ converter connected between instrument and computer USB port. The procedure requires Cary WinUV software to be installed first; the rear IEEE 488 (GPIB) connection is used with a shielded IEEE-488 cable.

User Manual, GPIB Communications; Setting Up the Instrument
Maintenance
Lamp changes require power isolation and cool-down

Before changing the deuterium or visible lamps, switch off the instrument, remove the power supply cable and allow the lamps to cool. Never touch the glass surfaces of new lamps; handle a lamp by its base using a soft cloth, since touching the envelope reduces efficiency. Lamp replacement is pre-aligned and plug-and-go.

User Manual, Source lamps, page 32; Factory Specification Sheet
Maintenance
Repeated fuse failure indicates a service fault

Two operator-accessible fuses are located at the rear; replace only with the type and rating indicated in the Site Preparation Guide or on the rear of the instrument. If a fuse repeatedly blows, it may indicate other problems and a service call may be required rather than repeated replacement.

User Manual, Fuses, pages 32–33
Process Limits
Removing the sample compartment floor loses alignment

The floor of the Cary 4000/5000/6000i/7000 sample compartment is completely removable for large apparatus, but once removed, optical alignment with respect to the floor is lost. Realign the optics before performing any microcell analyses.

User Manual, Removing the sample compartment base, page 28

Selection & Integration Q&A

When should I choose the Cary 4000 instead of the Cary 5000 or Cary 6000i?

Choose the Cary 4000 for UV-Vis work limited to 175–900 nm with fixed slits. The Cary 5000 extends to 3300 nm and the Cary 6000i to 1800 nm with an InGaAs NIR detector and PbSmart optics; select those only if NIR measurement is required.

What are the requirements for connecting the Cary 4000 to a computer and to accessories in the sample compartment?

Use the rear IEEE 488 (GPIB) port with a shielded IEEE-488 cable and a National Instruments USB-GPIB-HS+ converter, installing Cary WinUV first. The sample compartment 15-pin D-range connector carries −1000 VDC and +125 VDC; the DRA connector must only be used with Agilent Internal and External DRAs.

What mounting and duty-cycle conditions must be met before running measurements?

Allow 50 mm side and 150 mm rear clearance for air circulation, keep the sample compartment empty with its lid closed at power-on, and warm up for 2 hours for optimum performance. The Universal Power Supply adapts to mains voltage, but the actual supply voltage must still be verified.

What maintenance items and consumables are not included in the base delivery?

Nitrogen purge tubing (6 mm ID PVC or equivalent, never rubber), a pressure regulator and flowmeter, the USB-GPIB-HS+ converter and shielded IEEE-488 cable, replacement lamps, and the two rear fuses are not part of the base instrument. Lamp changes require power isolation and cool-down.