BINDER KB ECO 240
BINDER ·Made in Germany · In Active Production

BINDER KB ECO 240

BINDER KB ECO 240 — Incubators with Temperature 0…70 °C.

The KB ECO 240 is a 247 L thermoelectric (Peltier) cooling incubator with a 0–70 °C working range and ±0.1 K uniformity, built for microbiological cultivation and long-term storage. Within the KB ECO series it is the mid-size 230 V modification (200–230 V, 0.9 kW nominal), distinguished from the 400/720/1020 sizes by volume and from the UL variant by supply voltage. It serves as a stand-alone temperature-controlled chamber in laboratory workflows requiring stable, contamination-free incubation.

Temperature
0…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB ECO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB ECO 240 Current 0–7079.6146.124765030
KB ECO 1020 0–70114.9192.5102097345
KB ECO 400 0–7079.6194.640065030
KB ECO 720 0–7088.5192.570097345

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 247 L
Load per rack 30 kg
Max. heat compensation 350 W
Nominal power 0.9 kW
Permitted load 100 kg (both variants)
Temperature fluctuation 0.1 K
Temperature uniformity 0.1 K
Back wall clearance 100 mm (both variants)
Side wall clearance 100 mm (both variants)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 796 mm (both variants)
Height 1461 mm (both variants)
Interior width 650 mm (both variants)
Width 925 mm (both variants)
Net weight of the unit (empty) mfr 152 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 200...230 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior and working range

Control, installation and operation

Integration & Operating Notes

Electrical
Size supply and fusing for the 230 V variant's nominal current

The 230 V model draws a nominal current that must be accommodated by the branch circuit and the 16 A unit fuse. Undersized supply wiring or shared circuits will trip under Peltier run-up; verify voltage stays within 200–230 V at 50 Hz or 208–230 V at 60 Hz.

Instruction Manual KB ECO (E6.2), chap. 1.6 Type Plate
Process Limits
Do not load corrosive or explosive materials into the chamber

Loading material must not contain acids or halides that attack stainless steel, aluminum or copper components, and the chamber has no explosion protection. Corrosive damage from such ingredients is excluded from BINDER liability, so verify sample chemistry before loading.

Instruction Manual KB ECO (E6.2), chap. 1.9 Intended Use
Maintenance
Plan annual maintenance and electrical repeat inspection

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, and an electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial start-up and after any repair. Non-authorized maintenance voids the warranty.

Instruction Manual KB ECO (E6.2), chap. 24.1 and 24.2 Maintenance and Service
Cooling & Media
Clean Peltier fan grid several times a year in dusty rooms

With increased dust in ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year. A clogged grid reduces heat rejection and can prevent the chamber from reaching low set-points such as 4 °C.

Instruction Manual KB ECO (E6.2), chap. 24.2 Maintenance
Control & I/O
Use service reminder display to track maintenance due date

Holding the OK button for 5 seconds shows remaining days until maintenance is due; after the one-year interval the message 'Service due!' appears and reappears weekly. This lets operators schedule service without external tracking.

Instruction Manual KB ECO (E6.2), chap. 24.3 Service Reminder

Selection & Integration Q&A

When should I choose the KB ECO 240 over the 400, 720 or 1020 siblings?

Pick the 240 when 247 L interior (650 mm wide) fits your batch and you want the smallest 230 V KB ECO. Step up to 400/720/1020 only for larger loads; note the 400 additionally requires the flexible tilt protection kit, which the 240 does not.

What supply and fusing does the 230 V KB ECO 240 need at the wall?

The 230 V variant requires 200–230 V / 50 Hz or 208–230 V / 60 Hz, 1 N~, with a 16 A unit fuse. Size the branch circuit for the nominal current stated on the type plate and keep voltage within tolerance to avoid Peltier controller faults.

Can I install the KB ECO 240 in a corner or under a shelf?

Only with 100 mm clearance at the rear and both sides and never in an unventilated recess. The Peltier fan grid rejects heat to ambient; blocked airflow causes overheating and loss of cooling at low set-points. IP 20 also means keep it splash-proof.

What maintenance and accessories are needed beyond the base chamber?

Annual maintenance by BINDER-authorized personnel plus electrical repeat inspection per EN 50678/VDE 0701 are mandatory. For networking up to 100 chambers, APT-COM™ 4 software is a separate accessory; USB stick and Ethernet cabling are not included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB ECO 240 en
Datasheet ·527,3 КБ ·7 pp.
View PDF
PDF
KB ECO (E6.2) 07 2026 en
Manual ·9,3 МБ ·91 pp.
View PDF