BINDER KB ECO 720
BINDER ·Made in Germany · In Active Production

BINDER KB ECO 720

BINDER KB ECO 720 — Incubators with Temperature 0…70 °C.

The KB ECO 720 is a 700 L thermoelectric cooling incubator (Peltier technology) with a 0–70 °C working range and 0.1 K uniformity, positioned as the mid-large chamber between the 400 and 1020 siblings. Its 500 W heat compensation and 45 kg per-rack load capacity suit high-density incubation and long-term storage at 4–37 °C.

Temperature
0…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB ECO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB ECO 720 Current 0–7088.5192.570097345
KB ECO 1020 0–70114.9192.5102097345
KB ECO 240 0–7079.6146.124765030
KB ECO 400 0–7079.6194.640065030

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 700 L
Load per rack 45 kg
Max. heat compensation 500 W
Nominal power 1.3 kW
Permitted load 150 kg (both variants)
Temperature fluctuation 0.1 K
Temperature uniformity 0.1 K
Back wall clearance 100 mm (both variants)
Side wall clearance 100 mm (both variants)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 885 mm (both variants)
Height 1925 mm (both variants)
Interior width 973 mm (both variants)
Width 1250 mm (both variants)
Net weight of the unit (empty) mfr 272 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 200...230 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in different modes

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must be sized for 5.8 A at 200–230 V

The 230 V variant draws a nominal current of 5.8 A and 1.3 kW nominal power; the unit fuse is 16 A. Ensure the branch circuit and upstream protection are rated for this continuous load and that the CEE 7/7 plug is compatible with the site outlet. UL variants (100–120 V, 7.5 A) are separate models.

Data Sheet Model KB ECO 720 en.pdf; Instruction Manual KB ECO (E6.2), chap. 1.6 Type plate
Mounting
Keep 100 mm clearance at back and sides

The chamber must not be installed in unventilated recesses; prescribed minimum distances of 100 mm at the rear and sides must be observed to allow heat dispersal from the Peltier cooling system. Failure to do so risks overheating damage and invalidates performance.

Instruction Manual KB ECO (E6.2), chap. 1.8 and chap. 3.4
Process Limits
Do not load corrosive substances or operate in explosive areas

Loading material must not contain acids or halides that attack stainless steel, aluminum or copper. The chamber has no explosion protection and must not be used with combustible dust, solvents or explosive atmospheres. It is not a medical device per EU 2017/745 and must not be used for sterilization.

Instruction Manual KB ECO (E6.2), chap. 1.8 and chap. 1.9
Maintenance
Annual maintenance and electrical repeat inspection required

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts; warranty is void if non-authorized personnel perform it. An electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial start-up and after any maintenance or repair. The service reminder displays remaining days when OK is held for 5 s.

Instruction Manual KB ECO (E6.2), chap. 24.1–24.3
Cooling & Media
Clean Peltier fan grid several times a year in dusty air

With increased dust in the ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year to maintain cooling performance. The door gasket must only be replaced when cold; replacing it warm may damage it.

Instruction Manual KB ECO (E6.2), chap. 24.2

Selection & Integration Q&A

When should I choose the KB ECO 720 instead of the 400 or 1020 models?

Choose the 720 when you need 700 L interior volume and up to 150 kg total load with 45 kg per rack, between the 400 and 1020. All share 0–70 °C and 0.1 K uniformity; the 720 balances capacity and footprint (1,250 mm wide).

What are the power supply and fusing requirements for the 230 V variant?

The 230 V model requires 200–230 V, 50/60 Hz, 1-phase supply with nominal current 5.8 A and nominal power 1.3 kW; unit fuse is 16 A. Use a CEE 7/7 outlet. The UL variant (100–120 V, 7.5 A, NEMA 5-20) is a separate model.

What ambient and installation limits apply to the KB ECO 720?

Install with 100 mm clearance at back and sides, not in unventilated recesses. IP20 means splash-proof setup only. Ambient conditions per chap. 3.4 apply; dusty air requires frequent fan-grid cleaning. Do not operate in explosive areas.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance by BINDER-authorized personnel and electrical repeat inspection per EN 50678/50699 are mandatory. APT-COM™ 4 software (up to 100 chambers) and USB stick for data export are optional accessories not included. Service reminder is built-in.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB ECO 720 en
Datasheet ·626,5 КБ ·7 pp.
View PDF
PDF
KB ECO (E6.2) 07 2026 en
Manual ·9,3 МБ ·91 pp.
View PDF