BINDER KB PRO 470
BINDER KB PRO 470 — Incubators with Temperature -20…100 °C.
The KB PRO 470 is a 472 L cooling incubator with a -20 to +100 °C working range, part of the KB PRO series but distinguished by its 470-size cabinet, signal-triangle control terminal, and flexible tilt protection kit. It is intended for temperature-controlled incubation and storage where 0.1 K fluctuation and 0.2 K uniformity are required, and it supports rack loading up to 40 kg per rack and 200 kg total.
Direct drop-in replacements
BINDER KB PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Interior width, mm | Load per rack, kg |
|---|---|---|---|---|---|---|
| KB PRO 470 Current | -20–100 | 90 | 195.1 | 472 | 656 | 40 |
| KB PRO 1060 | -20–100 | 117.5 | 195.1 | 1070 | 1011 | 40 |
| KB PRO 130 | -20–100 | 71 | 100.5 | 127 | 656 | 30 |
| KB PRO 1600 | -20–100 | 117.5 | 195.1 | 1610 | 1521 | 40 |
| KB PRO 260 | -20–100 | 90 | 139.6 | 262 | 656 | 40 |
| KB PRO 720 | -20–100 | 90 | 195.1 | 727 | 1011 | 40 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 472 L |
| Load per rack | 40 kg |
| Max. heat compensation | 800 W |
| Nominal power | 1.5 kW |
| Permitted load | 200 kg |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm |
| Side wall clearance | 180 mm |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 4 min |
| Depth | 900 mm |
| Height | 1951 mm |
| Interior width | 656 mm |
| Width | 930 mm |
| Net weight of the unit (empty) mfr | 233 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The KB PRO 470 uses a 472 L interior (656 mm wide, 900 mm deep, 1,951 mm high) and is one of six KB PRO sizes (130, 260, 470, 720, 1060, 1600). Unlike the 130/260, which use a triangular instrument panel, the 470 and larger models use a signal triangle with an ergonomically adjustable control terminal.
- The 470 is specifically equipped with a flexible tilt protection kit (shared only with the 130) and, like all models from size 260 upward, requires a wall spacer to maintain the prescribed 100 mm rear and 180 mm side clearances.
- Net weight is 233 kg empty; the unit is single-phase (1 N~) and rated 208–240 V, 50 Hz, with a 16.0 A unit fuse and 1.5 kW nominal power.
Thermal behavior and load
- The chamber holds -20 to +100 °C with 0.1 K temperature fluctuation and 0.2 K uniformity. Recovery after a 30 s door opening at 37 °C is 4 minutes.
- Maximum heat compensation is 800 W, which sets the practical limit for how much self-heating load can be introduced without exceeding the control range. Rack loading is limited to 40 kg per rack and 200 kg total permitted load.
- No digitized curve points are available for this model; behavior under load must be validated against the 800 W heat-compensation ceiling and the 200 kg load limit.
Control, installation and operation
- The MB2 chamber controller provides fixed-value, timer, time-program and week-program modes, plus hot-air disinfection, defrosting, anti-condensation protection, alarm/notification functions, user management and a chart recorder display.
- Temperature safety is provided per DIN 12880:2007 class 2/3.3, with selectable class 2 (temperature limiter) or class 3.3 (temperature protection) and optional class 3.2 undertemperature protection.
- Installation requires ventilation and the prescribed wall clearances; the chamber must not be installed in unventilated recesses. Electrical connection is single-phase 208–240 V with branch protection sized for the 16.0 A unit fuse.
Integration & Operating Notes
The KB PRO 470 requires the wall spacer (from size 260 upward) to maintain 100 mm rear and 180 mm side clearance, and it must be fitted with the flexible tilt protection kit. Installing without these risks overheating from insufficient heat dispersal and loss of stability; the chamber must not be placed in unventilated recesses.
The unit is single-phase 1 N~, 208–240 V, 50 Hz, with a 16.0 A unit fuse and 1.5 kW nominal power. The supply circuit and upstream protection must be rated for the unit fuse current, and voltage/frequency must be verified before connection; the manual requires electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial startup and after maintenance.
Maximum heat compensation is 800 W, so the combined heat input from samples and any internal equipment must stay within this budget to hold setpoint across the -20 to +100 °C range. Rack loading is separately limited to 40 kg per rack and 200 kg total permitted load; exceeding either limit compromises temperature uniformity and mechanical safety.
The temperature safety device is class 2/3.3 per DIN 12880:2007; the safety controller can be set as class 2 (temperature limiter) or class 3.3 (temperature protection), with class 3.2 undertemperature protection available. The selected class and its limit/offset settings determine the alarm state and function check, so they must be defined before the chamber is released to the process.
A service message appears after 8760 operating hours or two years and repeats every two weeks until reset by BINDER Service. Annual maintenance is required, and all refrigeration work must be documented and performed by technicians trained per EN 13313:2010. In dusty ambient air, clean the condenser several times a year; replace the door gasket only when cold.
Selection & Integration Q&A
Pick the 470 for a 472 L working volume with 656 mm interior width and 900 mm depth. The 260 is smaller; the 720 and larger use the same signal-triangle control terminal but add volume. The 470 alone shares the flexible tilt protection kit with the 130, so plan for that kit and the wall spacer required from size 260 upward.
It needs single-phase 208–240 V, 50 Hz, with a 16.0 A unit fuse and 1.5 kW nominal power. Zero-voltage alarm contacts, analog temperature outputs, internal LED light, door lock, Pt 100 object sensor and APT-COM 4 software are options, not base delivery; specify them if remote alarm or recording is required.
Annual maintenance is required, with a service message at 8760 operating hours or two years that only BINDER Service can reset. Refrigeration work needs EN 13313:2010-trained personnel. Condenser cleaning is needed several times a year in dusty air; door gaskets, racks (U-rail or telescopic), APT-COM 4 and interface options are accessories, not base delivery.