BINDER KB PRO 720
BINDER ·Made in Germany · In Active Production

BINDER KB PRO 720

BINDER KB PRO 720 — Incubators with Temperature -20…100 °C.

The KB PRO 720 is a 727 L forced-convection cooling incubator with a -20 to 100 °C working range and 0.1 K fluctuation, positioned as the mid-large chamber in the KB PRO series. Its 1,000 W heat compensation and 200 kg permitted load distinguish it from smaller siblings, making it suited to high-density batch incubation and temperature cycling in process laboratories.

Temperature
-20…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB PRO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB PRO 720 Current -20–10090195.1727101140
KB PRO 1060 -20–100117.5195.11070101140
KB PRO 130 -20–10071100.512765630
KB PRO 1600 -20–100117.5195.11610152140
KB PRO 260 -20–10090139.626265640
KB PRO 470 -20–10090195.147265640

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 727 L
Load per rack 40 kg
Max. heat compensation 1000 W
Nominal power 1.7 kW
Permitted load 200 kg
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm
Side wall clearance 180 mm
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 900 mm
Height 1951 mm
Interior width 1011 mm
Width 1285 mm
Net weight of the unit (empty) mfr 296 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in different modes

Installation, connection and control

Integration & Operating Notes

Mounting
Wall spacer and clearances are mandatory for this size

From size 260 upward the wall spacer must be fitted, and the chamber needs 100 mm rear and 180 mm side clearance. Installing in an unventilated recess or ignoring these distances prevents heat dispersal and can cause overheating damage.

Instruction Manual, Section 4.1 and Section 1.8
Electrical
Branch circuit must be rated for 7.1 A at 208–240 V

The unit draws 1.7 kW at 7.1 A nominal on a single phase with a 16.0 A unit fuse. Verify supply voltage and frequency (50/60 Hz) before connection; an undersized or shared branch risks nuisance tripping and invalid temperature runs.

Data Sheet Model KB PRO 720; Instruction Manual, Section 1.6
Cooling & Media
Refrigeration work requires EN 13313-trained personnel

The R290 refrigerant circuit (0.095 kg, max 28 bar) may only be serviced by technicians certified per EN 13313:2010 / Regulation (EC) 303/2008. Plan service contracts accordingly; unauthorized intervention voids safety compliance and may be illegal.

Instruction Manual, Section 22.2
Maintenance
Annual maintenance and 8760-hour service reset

Maintenance is required at least once a year, and a service message appears after 8760 operating hours or two years, repeating every two weeks until reset by BINDER Service. In dusty environments, clean the condenser several times a year by suction.

Instruction Manual, Section 22.2
Process Limits
Door gasket must be replaced cold only

Replacing the door gasket while the chamber is warm will damage the gasket. Schedule gasket replacement with the chamber at ambient temperature, and include electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 at commissioning and after repair.

Instruction Manual, Section 22.2 and Section 22.1

Selection & Integration Q&A

When should I choose the KB PRO 720 rather than the KB PRO 470 or KB PRO 1060?

Pick the 720 when 727 L and a 200 kg load are needed but the 1060/1600 footprint is too large; the 470 suits smaller batches. All share -20…100 °C, but only the 720 balances 1,000 W heat compensation with a mid-size 1,285 mm width.

What supply and protection does the KB PRO 720 require, and can it share a circuit?

It needs 208–240 V single-phase, 50/60 Hz, 7.1 A nominal (1.7 kW) with a 16.0 A unit fuse and CEE 7/7 plug. Size the branch for the full 7.1 A; sharing with other high-draw equipment risks voltage dips that disturb the 0.1 K control stability.

What installation clearances and environment limits apply to this chamber?

Fit the wall spacer (mandatory from size 260) and keep 100 mm rear and 180 mm side clearance. Do not install in unventilated recesses; ensure heat dispersal. In dusty air, clean the condenser several times a year to protect refrigeration performance.

What maintenance and accessories are not included in base delivery?

Base delivery excludes APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock and the flexible Pt 100 sensor. Annual maintenance and electrical safety tests are required; a service message appears at 8760 h or two years until reset by BINDER Service.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB PRO 720 en
Datasheet ·551,3 КБ ·8 pp.
View PDF
PDF
KB PRO (E7) 09 2025 en
Manual ·12,7 МБ ·143 pp.
View PDF