BINDER MKF 115
BINDER MKF 115 — Climate chambers with Temperature -40…180 °C.
The BINDER MKF 115 is a 115-litre alternating climate chamber with a working range of -40 to 180 °C and 10 to 98 % RH, driven by a 4.8 kW 400 V 3-phase supply. Its 400 W heat compensation and 5.5 K/min heating rate distinguish it within the MKF series for rapid, humidified temperature cycling. It serves as a programmable test station for material and component qualification under combined temperature-humidity profiles.
BINDER MKF Series — Model Comparison & Selection
| Model | Temperature, °C | Height, cm | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW |
|---|---|---|---|---|---|---|
| MKF 115 Current | -40–180 | 172.5 | 115 | 30 | 400 | 4.8 |
| MKF 1020 | -40–180 | 200.5 | 991 | 40 | 1500 | 12.5 |
| MKF 1020 CO₂ | -48–180 | 200.5 | 991 | 40 | 1400 | 11.5 |
| MKF 240 | -40–180 | 171.5 | 228 | 30 | 400 | 6.8 |
| MKF 240 CO₂ | -50...180 | 171.3 | 228 | 30 | 900 | 6.8 |
| MKF 400 | -40–180 | 171 | 417 | 30 | 800 | 8 |
| MKF 400 CO₂ | -50...180 | 172.7 | 417 | 30 | 1600 | 7.9 |
| MKF 56 | -40–180 | 144.5 | 60 | 15 | 250 | 2.8 |
| MKF 720 | -40–180 | 200.5 | 734 | 40 | 1000 | 11 |
| MKF 720 CO₂ | -48–180 | 200.5 | 734 | 40 | 1100 | 11 |
Full specifications
| Temperature | -40...180 °C |
| Interior volume | 115 L (MKF115-400V, 50 Hz) |
| Load per rack | 30 kg (MKF115-400V, 50 Hz) |
| Maximum heat compensation | 400 W |
| Nominal power | 4.8 kW (MKF115-400V, 50 Hz) |
| Article Number mfr | 9020 |
| Height | 1725 mm (MKF115-400V, 50 Hz) |
| Width | 980 mm (MKF115-400V, 50 Hz) |
| Power frequency | 50 Hz (MKF115-400V) |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A (MKF115-400V, 50 Hz) |
| Humidity range | 10...98 % RH |
| Cooling down time from 180°C to -40°C mfr | 100 min |
| Humidity fluctuation with humidity dependent on set value mfr | 2.5 % r.F. |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.5 K/min |
Description
Construction and series positioning
- The MKF 115 is the 115-litre member of the MKF (E5) alternating climate chamber series; larger siblings are the MKF 240 and MKF 1020, while the MKFT variants add deep-temperature capability.
- This model is supplied as standard in the 400 V 3~/N/PE 50 Hz configuration (Art. No. 9020-0379); the -C suffix version (Art. No. 9020-0357) adds a voltage and frequency changer for 480 V 3~ 60 Hz sites.
- From size 115 upward the chamber has both a main power switch and a rear power switch, unlike the smaller MKF 56.
- The interior provides 115 L usable volume with a maximum load of 30 kg per rack.
Behaviour under load and in different media
- Temperature range spans -40 to 180 °C; with humidity active the range narrows to 10 to 95 °C and 10 to 98 % RH.
- Temperature uniformity is 0.1 to 1.3 K and fluctuation 0.1 to 0.6 K depending on set value; humidity fluctuation is 2.5 % r.F.
- Average heating rate per IEC 60068-3-5 between -40 °C and 180 °C is 5.5 K/min; cooling from 180 °C to -40 °C takes 100 min.
- Maximum heat compensation is 400 W, allowing dissipation from active test specimens.
- The humidification/dehumidification system uses freshwater and wastewater paths; a water circle lever for condensate recycling is available from size 115 on.
Installation, control and operation
- The MB2 chamber controller supports fixed-value, timer, time-program and week-program modes, with ramp/step set-point entry, tolerance ranges and section repetition.
- Temperature safety devices include an over-temperature protective device (class 1) and a safety controller (class 2); an optional over/under temperature safety device class 2 is available.
- Water connection can be automatic via water pipe or manual via internal freshwater can from size 115 on; a water cooling option and controlled compressed air dryer are available for this size.
- Zero-voltage switching outputs, RS485, analog outputs and Ethernet are available for integration with external monitoring or alarm systems.
Integration & Operating Notes
The chamber draws 4.8 kW nominal on a 3 N/PE 400 V 50 Hz supply and requires a 16.0 A unit fuse. Confirm the site distribution provides a neutral conductor and protective earth; the -C variant with voltage and frequency changer must be specified where 480 V 60 Hz is the only available supply.
From size 115 on the chamber has a rear power switch in addition to the main power switch, and the condenser fan grid sits behind the left maintenance access flap. Both the rear switch and the fan grid must remain reachable, so leave clearance behind and beside the unit for commissioning and weekly fan-grid checks in dusty rooms.
Although the chamber reaches -40 to 180 °C dry, humidity operation is restricted to a 10 to 95 °C window with 10 to 98 % RH. Program set points that combine humidity with temperatures outside this band will not be controllable; plan test profiles so that humidified segments stay within the stated envelope.
The manual requires maintenance at least once a year, a leak test per Regulation (EU) 517/2014, and repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial start-up and after repair. The steam humidifier also needs annual servicing whose interval depends on local water quality and steam output.
After 8760 operating hours or two years the controller displays a service message; once acknowledged it reappears every two weeks until reset by BINDER Service. Budget for a service visit at that interval, and include condenser cleaning every one to two years and weekly fan-grid checks where ambient dust is elevated.
Selection & Integration Q&A
Pick the MKF 115 when 115 L and 30 kg per rack cover the specimen load and the profile stays within -40 to 180 °C with humidity between 10 and 95 °C. The MKF 240 offers more chamber volume for larger batches; the MKFT 115 is the deep-temperature variant for profiles below the MKF range.
A 400 V 3 N/PE 50 Hz supply fused at 16.0 A for the 4.8 kW load, plus a freshwater and wastewater connection for the humidification system (manual internal can from size 115 on, or automatic via water pipe). The -C variant is required where only 480 V 60 Hz is available.
Use the zero-voltage switching outputs for hard-wired alarm forwarding, or the optional RS485 interface, analog outputs for temperature and humidity, and Ethernet for network integration. APT-COM 4 software is available as an option for multi-unit management; these interfaces are not part of the base delivery.
Annual maintenance, leak test per (EU) 517/2014, electrical inspection per EN 50678/50699, annual steam-humidifier service and condenser cleaning every 1–2 years. Consumables such as disinfectant spray Art. No. 1002-0022, the water cooling option, compressed air dryer and interface cables are options or accessories not supplied with the standard unit.