BINDER MKF 240 CO₂
BINDER ·Made in Germany · In Active Production

BINDER MKF 240 CO₂

BINDER MKF 240 CO₂ — Climate chambers.

The MKF 240 CO₂ is a 228 L alternating climate chamber with a working range of -50 to 180 °C and 10 to 98 % RH, distinguished within the MKF series by its extended low-temperature limit and CO₂ refrigerant circuit. It is intended for programmed temperature/humidity cycling of test specimens in a laboratory, with a 900 W heat compensation capacity for active loads.

BINDER MKF Series — Model Comparison & Selection

Model Temperature, °CHeight, cmInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
MKF 240 CO₂ Current -50...180171.3228309006.8
MKF 1020 -40–180200.599140150012.5
MKF 1020 CO₂ -48–180200.599140140011.5
MKF 115 -40–180172.5115304004.8
MKF 240 -40–180171.5228304006.8
MKF 400 -40–180171417308008
MKF 400 CO₂ -50...180172.74173016007.9
MKF 56 -40–180144.560152502.8
MKF 720 -40–180200.573440100011
MKF 720 CO₂ -48–180200.573440110011

Full specifications

Core Performance & Functional Specifications
Temperature -50...180
Interior volume 228 L (MKF240CO2-400V, 50 Hz)
Load per rack 30 kg (MKF240CO2-400V, 50 Hz)
Maximum heat compensation 900 W
Nominal power 6.8 kW (MKF240CO2-400V, 50 Hz)
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 925 mm (MKF240CO2-400V, 50 Hz)
Height 1713 mm (MKF240CO2-400V, 50 Hz)
Width 1115 mm (MKF240CO2-400V, 50 Hz)
Electrical, Power & Drive Specifications
Power frequency 50 Hz (MKF240CO2-400V)
Rated voltage 400 V
Unit fuse 16.0 A (MKF240CO2-400V, 50 Hz)
Operating Conditions, Environment & Protection
Humidity range 10...98 % RH
Cooling down time from 180°C to -40°C mfr 60 min
Humidity fluctuation with humidity dependent on set value mfr 2 % r.F.
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.5 K/min

Description

Construction and series positioning

Working behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must include a type B RCD and stay within ±10 %

The chamber is labeled for operation only with residual current circuit breakers type B, and the nominal voltage tolerance is ±10 % at the indicated power frequency. Size the feeder for 6.8 kW / 16 A at 400 V 3 N/PE 50 Hz and verify the actual type plate before connection.

Instruction Manual, Section 1.5 and Section 1.6 Type Plate
Process Limits
Humidity operation narrows the usable temperature window

With humidity control active the chamber is limited to 10…95 °C and 10…98 % RH, so programs that combine low-temperature phases below 10 °C with humidification are not permissible. Plan test profiles so that humidity sections stay inside this window.

Factory Specification Sheet; Instruction Manual, Section 1.14
Process Limits
Aggressive atmospheres can destroy the humidity sensor

The humidity sensor has tabulated maximum workplace concentrations for substances such as ammonia, acetone, chlorine, acetic acid, ethanol, formaldehyde, ozone, hydrochloric acid, hydrogen sulphide, sulphur dioxide and xylene. Permanent exposure above these limits damages the sensor, so specimens outgassing such substances must be evaluated before testing.

Instruction Manual, Section 1.14, p.20
Mounting
Room volume and ventilation must be verified before siting

Required room volume must be calculated per EN 378-1:2016+A1:2020 because of the CO₂ refrigerant charge, and installation in unventilated recesses is prohibited. Observe the minimum distances of chapter 3.4 so that heat from the refrigeration circuit can disperse.

Instruction Manual, Section 1.8, p.13
Maintenance
Service message appears at 8760 h or two years

After 8760 operating hours or two years the controller displays a service message; after confirmation it reappears every two weeks until reset by BINDER Service. Plan the annual maintenance and the BINDER Service reset together so the recurring message does not interrupt long test programs.

Instruction Manual, Section 23.2, p.121
Maintenance
Condenser and humidifier need fixed cleaning intervals

Condensers should be cleaned every 1 to 2 years by a qualified technician, and the steam humidifier requires maintenance at least once a year, with frequency depending on water quality and steam output. In dusty ambient air the condenser fan grid behind the left maintenance access flap must be checked weekly and cleaned by suction after disconnecting the chamber.

Instruction Manual, Section 23.2, p.121
Maintenance
Door gasket may only be replaced when cold

The door gasket must be replaced only when the chamber is cold; replacing it warm can damage the gasket. Allow the chamber to cool to ambient before any gasket service, and include this in the maintenance procedure.

Instruction Manual, Section 23.2, p.121
Electrical
Repeat electrical safety test before resuming after service

An annual repeat electrical inspection is required, plus a test prior to initial startup and prior to resuming operation after maintenance or repair, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. Build this test into the commissioning and service workflow.

Instruction Manual, Section 23.1, p.120

Selection & Integration Q&A

When should I choose the MKF 240 CO₂ instead of the standard MKF 240 or the larger MKF 1020 CO₂?

Pick the MKF 240 CO₂ when you need the -50 °C low limit in a 228 L cabinet; the standard MKF 240 stops at -40 °C. Choose the MKF 1020 CO₂ only if the larger cabinet volume is required, since it is a different size class, not an option of this model.

What water and drain connections must be provided for humidity operation?

Humidity control requires a freshwater supply, either automatic via a water pipe with the connection kit or manual via the internal freshwater can, plus a wastewater connection. A condensate recycling lever and a hose burst/reflux safety kit are available as options and should be specified at order time.

Can the chamber be installed in a small closed room or a recess?

No. Installation in unventilated recesses is prohibited, and the required room volume must be calculated per EN 378-1:2016+A1:2020 because of the CO₂ refrigerant charge. Observe the minimum distances of chapter 3.4 so refrigeration heat can disperse.

What maintenance and consumables are not included in the base delivery?

The base unit does not include the RS485 interface, analog temperature/humidity outputs, flexible Pt 100 object sensor, condensate recycling lever, hose burst/reflux safety kit, APT-COM 4 software or BINDER Pure Aqua Service. Plan annual maintenance, condenser cleaning every 1–2 years, yearly steam humidifier service and the 8760 h/two-year BINDER Service reset.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model MKF CO₂ 240 en
Datasheet ·836,9 КБ ·9 pp.
View PDF
PDF
0715 EN MKF CO2 E6 UM 10 2026
Manual ·11,3 МБ ·155 pp.
View PDF