BINDER MKF 56
BINDER MKF 56 — Climate chambers with Temperature -40…180 °C.
The BINDER MKF 56 is a 60-litre alternating climate chamber with a working range of -40 to 180 °C and 10 to 98 % RH, built for rapid temperature cycling. Its 5.0 K/min heating rate and 90-minute full-span cooling distinguish this smallest MKF size from the larger 400 V siblings. It serves as a compact test station for humidity-controlled temperature-change profiles in laboratory metrology.
BINDER MKF Series — Model Comparison & Selection
| Model | Temperature, °C | Height, cm | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW |
|---|---|---|---|---|---|---|
| MKF 56 Current | -40–180 | 144.5 | 60 | 15 | 250 | 2.8 |
| MKF 1020 | -40–180 | 200.5 | 991 | 40 | 1500 | 12.5 |
| MKF 1020 CO₂ | -48–180 | 200.5 | 991 | 40 | 1400 | 11.5 |
| MKF 115 | -40–180 | 172.5 | 115 | 30 | 400 | 4.8 |
| MKF 240 | -40–180 | 171.5 | 228 | 30 | 400 | 6.8 |
| MKF 240 CO₂ | -50...180 | 171.3 | 228 | 30 | 900 | 6.8 |
| MKF 400 | -40–180 | 171 | 417 | 30 | 800 | 8 |
| MKF 400 CO₂ | -50...180 | 172.7 | 417 | 30 | 1600 | 7.9 |
| MKF 720 | -40–180 | 200.5 | 734 | 40 | 1000 | 11 |
| MKF 720 CO₂ | -48–180 | 200.5 | 734 | 40 | 1100 | 11 |
Full specifications
| Temperature | -40...180 °C |
| Interior volume | 60 L (MKF056-230V, 50 Hz) |
| Load per rack | 15 kg (MKF056-230V, 50 Hz) |
| Maximum heat compensation | 250 W |
| Nominal power | 2.8 kW (MKF056-230V, 50 Hz) |
| Article Number mfr | 9020 |
| Depth | 780 mm (MKF056-230V, 50 Hz) |
| Height | 1445 mm (MKF056-230V, 50 Hz) |
| Width | 720 mm (MKF056-230V, 50 Hz) |
| Power frequency | 50 Hz (MKF056-230V) |
| Rated voltage | 230 V |
| Unit fuse | 16.0 A (MKF056-230V, 50 Hz) |
| Humidity range | 10...98 % RH |
| Cooling down time from 180°C to -40°C mfr | 90 min |
| Humidity fluctuation with humidity dependent on set value mfr | 2.5 % r.F. |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.0 K/min |
Description
Construction and series positioning
- The MKF 56 is the only MKF size supplied as a single-phase 230 V / 50 Hz unit (MKF056-230V, Art. 9020-0378); the 240 V / 60 Hz export version is MKF056-240V (Art. 9020-0389). All larger MKF models (115, 240, 400, 720, 1020) are 400 V machines.
- Net interior volume is 60 L; net cabinet dimensions are 720 mm wide, 780 mm deep and 1,445 mm high. The main power switch is on the front; unlike sizes 115 and above, there is no separate rear power switch.
- Freshwater supply is manual via an external can on the MKF 56; the internal freshwater can and the water-circle condensate lever are reserved for sizes 115 and up. External freshwater and wastewater cans are an MKF 56 option.
Behaviour across the working range
- Temperature spans -40 to 180 °C in dry operation; humidity control is limited to 10 to 95 °C and 10 to 98 % RH. Temperature uniformity is 0.5 to 1.5 K and fluctuation 0.1 to 0.5 K, both set-point dependent.
- Average heating rate per IEC 60068-3-5 between -40 and 180 °C is 5.0 K/min; cooling from 180 °C to -40 °C takes 90 minutes. Maximum heat compensation is 250 W, which caps the dissipating load that can be held at a set point.
- Humidity fluctuation is 2.5 % RH. The humidity sensor is sensitive to ammonia and ozone; process atmospheres must stay below the manual's concentration limits.
Control, installation and operation
- The MB2 chamber controller supports fixed-value, timer, time-program and week-program modes, with ramp/step set-point entry, tolerance ranges and section repetition. Zero-voltage switching outputs are available for external signalling.
- Temperature safety is provided by an over-temperature protective device (class 1) and a safety controller (class 2); an over/under temperature safety device class 2 is optional.
- Electrical supply is 230 V, 1~, 50 Hz with a 16.0 A unit fuse and 2.8 kW nominal power. Water connections for humidification and, if ordered, cooling water must be established before start-up.
Integration & Operating Notes
Although the chamber reaches -40 to 180 °C in dry operation, humidity control is specified only from 10 to 95 °C. Programming a humidified ramp outside this band will not produce valid humidity conditions and may stress the humidification system; split such profiles into dry temperature segments and humidified segments within the permitted band.
The maximum heat compensation of 250 W defines how much heat a powered test specimen may release while the chamber still holds its set point. Above this value the chamber cannot maintain the programmed temperature, so calculate specimen power before designing a test and reduce load or split specimens if necessary.
With increased dust in the ambient air the condenser fan must be cleaned several times a year, and the fan grid behind the left maintenance access flap checked weekly and cleaned by suction after disconnecting the chamber. Condensers themselves are recommended for cleaning every one to two years by a qualified technician.
The humidity sensor has stated resistance limits, for example ammonia at a maximum workplace threshold of 20 ppm and permanent load of 5500 ppm, and ozone at 0.1 ppm threshold and 0.5 ppm permanent load. Verify specimen outgassing before running humidified tests, because exceeding these concentrations degrades the sensor and invalidates humidity readings.
Selection & Integration Q&A
Pick the MKF 56 for 60 L loads on a 230 V single-phase supply, where its 5.0 K/min heating and 90-minute full-span cooling suffice. The 115 and 240 are 400 V machines with larger volumes; choose them when specimen size or throughput exceeds the 60 L interior and 15 kg per-rack limit.
Unlike larger MKF sizes, the MKF 56 uses a manual external freshwater can rather than an automatic water-pipe connection; wastewater must be routed to a drain or to the optional external wastewater can. Order the external can option if no fixed drain is available at the installation site.
Allow clearance for the condenser fan grid behind the left maintenance access flap and keep ambient dust low; in dusty air the grid needs weekly suction cleaning. The 2.8 kW electrical draw and condenser heat rejection must be removed from the room, so size room ventilation accordingly.
Base delivery excludes the external freshwater/wastewater cans, water cooling, compressed air dryer, RS485 and analog outputs, Pt 100 object sensor and APT-COM 4 software. Plan annual maintenance with leak test, annual humidifier service, condenser cleaning every 1–2 years, and the BINDER Service reset at 8760 h.