BINDER MKF 56
BINDER ·Made in Germany · In Active Production

BINDER MKF 56

BINDER MKF 56 — Climate chambers with Temperature -40…180 °C.

The BINDER MKF 56 is a 60-litre alternating climate chamber with a working range of -40 to 180 °C and 10 to 98 % RH, built for rapid temperature cycling. Its 5.0 K/min heating rate and 90-minute full-span cooling distinguish this smallest MKF size from the larger 400 V siblings. It serves as a compact test station for humidity-controlled temperature-change profiles in laboratory metrology.

Temperature
-40…180 °C

BINDER MKF Series — Model Comparison & Selection

Model Temperature, °CHeight, cmInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
MKF 56 Current -40–180144.560152502.8
MKF 1020 -40–180200.599140150012.5
MKF 1020 CO₂ -48–180200.599140140011.5
MKF 115 -40–180172.5115304004.8
MKF 240 -40–180171.5228304006.8
MKF 240 CO₂ -50...180171.3228309006.8
MKF 400 -40–180171417308008
MKF 400 CO₂ -50...180172.74173016007.9
MKF 720 -40–180200.573440100011
MKF 720 CO₂ -48–180200.573440110011

Full specifications

Core Performance & Functional Specifications
Temperature -40...180 °C
Interior volume 60 L (MKF056-230V, 50 Hz)
Load per rack 15 kg (MKF056-230V, 50 Hz)
Maximum heat compensation 250 W
Nominal power 2.8 kW (MKF056-230V, 50 Hz)
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 780 mm (MKF056-230V, 50 Hz)
Height 1445 mm (MKF056-230V, 50 Hz)
Width 720 mm (MKF056-230V, 50 Hz)
Electrical, Power & Drive Specifications
Power frequency 50 Hz (MKF056-230V)
Rated voltage 230 V
Unit fuse 16.0 A (MKF056-230V, 50 Hz)
Operating Conditions, Environment & Protection
Humidity range 10...98 % RH
Cooling down time from 180°C to -40°C mfr 90 min
Humidity fluctuation with humidity dependent on set value mfr 2.5 % r.F.
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.0 K/min

Description

Construction and series positioning

Behaviour across the working range

Control, installation and operation

Integration & Operating Notes

Process Limits
Humidity control is unavailable below 10 °C or above 95 °C

Although the chamber reaches -40 to 180 °C in dry operation, humidity control is specified only from 10 to 95 °C. Programming a humidified ramp outside this band will not produce valid humidity conditions and may stress the humidification system; split such profiles into dry temperature segments and humidified segments within the permitted band.

Factory Specification Sheet; Operating Manual MKF/MKFT (E5) 10/2026, Technical Data
Process Limits
Live load dissipation must stay within 250 W heat compensation

The maximum heat compensation of 250 W defines how much heat a powered test specimen may release while the chamber still holds its set point. Above this value the chamber cannot maintain the programmed temperature, so calculate specimen power before designing a test and reduce load or split specimens if necessary.

Factory Specification Sheet
Maintenance
Inspect the condenser fan grid weekly in dusty rooms

With increased dust in the ambient air the condenser fan must be cleaned several times a year, and the fan grid behind the left maintenance access flap checked weekly and cleaned by suction after disconnecting the chamber. Condensers themselves are recommended for cleaning every one to two years by a qualified technician.

Operating Manual MKF/MKFT (E5) 10/2026, Maintenance and Service
Process Limits
Ammonia and ozone in the process air attack the humidity sensor

The humidity sensor has stated resistance limits, for example ammonia at a maximum workplace threshold of 20 ppm and permanent load of 5500 ppm, and ozone at 0.1 ppm threshold and 0.5 ppm permanent load. Verify specimen outgassing before running humidified tests, because exceeding these concentrations degrades the sensor and invalidates humidity readings.

Operating Manual MKF/MKFT (E5) 10/2026, Humidity Sensor Limits

Selection & Integration Q&A

When should I choose the MKF 56 over the MKF 115 or MKF 240?

Pick the MKF 56 for 60 L loads on a 230 V single-phase supply, where its 5.0 K/min heating and 90-minute full-span cooling suffice. The 115 and 240 are 400 V machines with larger volumes; choose them when specimen size or throughput exceeds the 60 L interior and 15 kg per-rack limit.

What water and drain connections does the MKF 56 need for humidity testing?

Unlike larger MKF sizes, the MKF 56 uses a manual external freshwater can rather than an automatic water-pipe connection; wastewater must be routed to a drain or to the optional external wastewater can. Order the external can option if no fixed drain is available at the installation site.

Can I put the MKF 56 in a small room with other heat-producing equipment?

Allow clearance for the condenser fan grid behind the left maintenance access flap and keep ambient dust low; in dusty air the grid needs weekly suction cleaning. The 2.8 kW electrical draw and condenser heat rejection must be removed from the room, so size room ventilation accordingly.

What maintenance and accessories are not included with the base MKF 56?

Base delivery excludes the external freshwater/wastewater cans, water cooling, compressed air dryer, RS485 and analog outputs, Pt 100 object sensor and APT-COM 4 software. Plan annual maintenance with leak test, annual humidifier service, condenser cleaning every 1–2 years, and the BINDER Service reset at 8760 h.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model MKF 56 en
Datasheet ·794,1 КБ ·10 pp.
View PDF
PDF
MKF+MKFT (E5) 10 2026 en
Manual ·18,1 МБ ·179 pp.
View PDF