Operating Instructions — C-VLR 401 OXY vacuum pump (BA 886-OXY, ed. 1.0 12.10.2023, EN)

Technical Reference Summary

Covered Models & Document Scope

Covers the contact-free running claw vacuum pumps C-VLR 401, C-VLR 501 and C-VLR 651 in the OXY variant (blue painted, for gases with increased oxygen content >21% by volume and other gaseous oxidising agents). Document BA 886-OXY-EN, edition 1.0, 12.10.2023.

Section-by-Section Technical Breakdown

pp. 14-15
Product overview and functioning

Describes the design of the C-VLR 401/501/651 OXY double-shaft rotary piston (claw) vacuum pumps with contact-free dry claw rotors synchronised by an oil-lubricated gear pair. Lists component positions (vacuum connection, exhaust silencer, condensate drain, cooling air inlet/outlet, oil filling point, oil sight glass, oil drain with magnet, direction-of-rotation arrow, hot surfaces >70 °C) and the data plate fields.

  • Double-shaft rotary piston vacuum pump with contact-free, dry-running claw rotors
  • Gearbox and bearings oil-lubricated; gearbox separated from compression chamber by special seals
  • Inlet-side connection flange and pressure-side exhaust silencer; intake air cleaned by mesh filter
  • Integral non-return valve prevents ventilation of the evacuated system after stop
  • Cooling by drum fan drawing fresh air through cooling air inlet and discharging heated air at cooling air outlet
  • Driven via coupling by flanged three-phase standard motor
  • OXY pumps are blue painted to distinguish them from standard variants
  • Data plate includes model/size, serial number, year of construction, data matrix barcode, article number, speed 50/60 Hz, motor output 50/60 Hz, operating mode, pumping capacity 50/60 Hz, final pressure (abs.) 50/60 Hz
pp. 16
Fields of application

Defines the permissible operating envelope: continuous operation between 700 mbar (abs.) and 200 mbar (abs.) inlet pressure, evacuation of closed systems limited to 2% of nominal pumping capacity, and free-intake pumping capacities at 50 Hz. Notes derating above 1000 m altitude and restrictions on switching frequency and ambient/inlet temperature.

  • Suitable for air or other gases with oxygen content by volume above 21%
  • Continuous operation permitted at any pressure between 700 mbar (abs.) and 200 mbar (abs.) inlet pressure
  • Volume of closed systems to be evacuated must be no more than 2% of nominal pumping capacity
  • Free-intake pumping capacity at 50 Hz: 380 m³/h (VLR 401), 495 m³/h (VLR 501), 650 m³/h (VLR 651)
  • Pumping capacity vs. intake pressure per data sheet D 886-OXY
  • Frequent switching (approx. 10 times per hour) or high ambient/inlet temperatures may exceed motor winding and bearing temperature limits — contact manufacturer
  • Observe motor protection class on motor data plate; protect outdoor installation from environmental influences
pp. 17-20
Installation

Covers preparation of installation, mechanical erection, pipe connection, lubricating oil check, electrical connection and pressure monitoring. Specifies ventilation to keep room oxygen ≤21% by volume, horizontal mounting with max. 8° inclination, minimum clearances, back-pressure limit, oil type and electrical protection requirements.

  • Sufficiently vent installation room; oxygen content by volume in room air may not exceed 21%
  • Machine freely accessible from all sides; do not close ventilation grids and holes
  • Provide at least 50 cm space around machine for maintenance
  • Operate only when set up horizontally; max. inclination 8° in all directions
  • Secure machine against tipping over and falling; floor must be plane and even
  • Cooling air inlets and outlets at least 30 cm from adjacent walls; cooling air must not be recirculated
  • Ambient temperature min. +5 °C, max. +40 °C
  • Maximum permissible back pressure +50 mbar; exhaust air opening must not be closed or restricted
  • Only screw in pipes by hand; use flexible connections/compensators if necessary
  • Use only oil type OXY-GEAR-LUBE 160
  • Electrical installation by qualified electrician per EN 60204; main switch and motor protection switch provided by user
  • Motor protection switch with overload-dependent turn-off delay recommended
  • Pressure monitoring system not included in scope of delivery; user must install it and switch off pump if minimum permissible pressure is undershot
pp. 21-23
Commissioning and decommissioning

Describes start-up checks, rotation-direction verification, operation including condensate discharge, and decommissioning/storage/recommissioning. Emphasises oil purity (OXY-GEAR-LUBE 160 only), restart only after standstill, and drying the pump before shutdown.

  • Installation check: no transport/assembly damage, safe horizontal standing, correct and leak-tight pipe connections, tight screw/flange connections, electrical installation per connection diagram, adequate ventilation, oil filled and level checked, pump and pipelines cleaned, optional accessories functional
  • Check rotation direction with phase sequence indicator (clockwise); brief start max. two seconds; when looking at motor blower it must rotate counter-clockwise
  • Keep a distance of 1 m to outlet and inlet connections during rotation check
  • Do not discharge condensate manually; allow machine to cool and wear protective gloves
  • Before shutdown >2 hours or decommissioning, run pump at 700 mbar (abs.) for at least 15–30 minutes with dry air to remove moisture and pollution
  • Restart only after machine has stopped completely
  • If lubricating oil may have been mixed with organic substances, replace immediately; if machine polluted with organic substances, put out of operation and clean by specialist personnel authorised by Elmo Rietschle
pp. 24-30
Maintenance and repair

Covers operational safety, maintenance table, cleaning, lubricating oil refilling and change, air filter, motor and coupling, repair/service and spare parts. Specifies oil type OXY-GEAR-LUBE 160 and references spare parts list E 886-OXY.

  • Only oil type OXY-GEAR-LUBE 160 permitted for lubrication
  • Check oil level through oil sight glass and top up if necessary
  • Air filter to be replaced when polluted (see Chapter 7.6)
  • Spare parts per spare parts list E 886-OXY
  • Repairs only by authorised workshops using original spare parts
pp. 31
Errors

Troubleshooting table (Tab. 4) listing malfunctions, causes and elimination measures, with references to chapters for electrical connection, pipe connection, air filter, intended use and cleaning. Warns never to restart a pump after automatic shut-off without determining and remedying the cause.

  • Never put pump back into operation after automatic shut-off without unequivocally finding out the reason and remedying it
  • Malfunctions covered: motor protection switch trip, insufficient suction capacity, final pressure not reached, machine overheating, strange noise
  • Contact Elmo Rietschle Service for other malfunctions or those that cannot be eliminated
pp. 33
Technical Data

Chapter 10 contains technical data; detailed values are in data sheet D 886-OXY. The manual references chapter 10 for weight, sound pressure level and other data.

  • Technical data and characteristic curves in data sheet D 886-OXY
  • Weight per chapter 10
  • Sound pressure level per chapter 10

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Ambient and inlet temperature must be between 5 °C and 40 °C. p.8 ↗ Operation outside this range requires consultation with the manufacturer; ensure installation room ventilation and cooling air supply. C-VLR 401 | 501 | 651 OXY
Maximum permissible back pressure is +50 mbar. p.8 ↗ Exhaust line must not be restricted or blocked; higher process back pressures require manufacturer consultation. C-VLR 401 | 501 | 651 OXY
Only oil type OXY-GEAR-LUBE 160 may be used for lubrication. p.8 ↗ Wrong oil creates acute fire risk in oxygen service; oil mixed with organic substances must be replaced immediately. C-VLR 401 | 501 | 651 OXY
Continuous operation permitted at any pressure between 700 mbar (abs.) and 200 mbar (abs.) inlet pressure. p.16 ↗ Process must remain within this inlet pressure band for continuous duty. C-VLR 401 | 501 | 651 OXY
When evacuating closed systems, the volume to be evacuated must be no more than 2% of the nominal pumping capacity. p.16 ↗ Size the evacuated volume relative to pump capacity to avoid inadmissible operating conditions. C-VLR 401 | 501 | 651 OXY
Free-intake pumping capacity at 50 Hz: 380 m³/h (VLR 401), 495 m³/h (VLR 501), 650 m³/h (VLR 651). p.16 ↗ Use these values for preliminary sizing; actual capacity depends on intake pressure per data sheet D 886-OXY. C-VLR 401, C-VLR 501, C-VLR 651
Maximum inclination during operation is 8° in all directions; machine must be set up horizontally. p.17 ↗ Mount on a plane, even floor and secure against tipping and falling. C-VLR 401 | 501 | 651 OXY
Cooling air inlets and outlets must be at least 30 cm away from adjacent walls; cooling air must not be recirculated. p.18 ↗ Maintain clearance and prevent hot exhaust air from other machines being drawn into the cooling system. C-VLR 401 | 501 | 651 OXY
Provide at least 50 cm space around the machine for maintenance work. p.17 ↗ Layout installation to allow access to oil filling point, oil sight glass and oil drain. C-VLR 401 | 501 | 651 OXY
Oxygen content by volume in the room air may not exceed 21% when the pump blows out freely. p.17 ↗ Install sufficient ventilation; the OXY pump is not gas-tight and can enrich room air with oxygen. C-VLR 401 | 501 | 651 OXY
Permissible voltage deviation ±5% of nominal value and frequency deviation ±2%. p.20 ↗ Verify mains conditions against motor data plate before connection. C-VLR 401 | 501 | 651 OXY
Motors comply with DIN EN 60034, protection class IP 55 and insulation class F. p.20 ↗ Suitable for the specified environment; observe motor data plate for actual motor data. C-VLR 401 | 501 | 651 OXY
Machines with AC motors >1.7 kVA, three-phase motors >5.2 kVA, or motors with starting currents >60 A are not designed for direct-on-line starting. p.20 ↗ Use star-delta starting, soft starter or frequency converter for such motors; reduce mechanical load during start. C-VLR 401 | 501 | 651 OXY
Pressure monitoring system is not included in the scope of delivery and must be installed by the user. p.20 ↗ Install a pressure monitoring device (e.g. pressure switch) that switches off the pump if minimum permissible pressure is undershot; permissible pressure is on the data plate. C-VLR 401 | 501 | 651 OXY
Check rotation direction by briefly starting motor (max. two seconds); when looking at the motor blower it must rotate counter-clockwise. p.22 ↗ Use a phase sequence indicator; keep 1 m distance from inlet and outlet connections during the check. C-VLR 401 | 501 | 651 OXY
Before shutdown for more than 2 hours or decommissioning, run the pump at 700 mbar (abs.) for at least 15–30 minutes with dry air. p.22 ↗ Removes moisture and pollution that could otherwise prevent restart or damage the machine. C-VLR 401 | 501 | 651 OXY
Storage temperature -20 °C to +70 °C, relative air humidity max. 80%, dust-free, dry, vibration-free, protected from sun radiation; storage not more than 6 months. p.13 ↗ Close openings air-tight during storage; observe storage guidelines I 100 and I 150. C-VLR 401 | 501 | 651 OXY
Output reduction noticeable when installed at more than 1000 m above sea level. p.18 ↗ Contact manufacturer for high-altitude installations. C-VLR 401 | 501 | 651 OXY
Surface temperatures on components may rise above 70 °C at operating temperature. p.21 ↗ Do not touch hot surfaces; wear protective gloves if necessary; provide guarding. C-VLR 401 | 501 | 651 OXY
The machine does not have its own Emergency Stop or Emergency Off; this must be implemented by the user. p.11 ↗ Integrate the machine into the user's safety concept. C-VLR 401 | 501 | 651 OXY

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Machine is switched off by the motor protection switch p.31 ↗ Mains voltage/ Frequency does not correspond with the motor data Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.31 ↗ Connection to motor terminal board is not correct Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.31 ↗ Motor protection switch is not set correctly Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.31 ↗ Motor protection switch is triggered too quickly Use a motor protection switch with an overload-dependent turn-off delay that considers the short excess current at start up (version with short circuit and overload trigger as per VDE 0660 Part 102 or IEC 60947-4-1)
Suction capacity is insufficient p.31 ↗ Air filter is polluted Replace air filter Chapter 7.6
Suction capacity is insufficient p.31 ↗ The inlet pipe is too long or too narrow Check the hose and/or the pipe Chapter 5.3
Suction capacity is insufficient p.31 ↗ Machine or system leaking Check the pipework and screw connections for leaks and check for tight fit Chapter 5.3
Final pressure (max. vacuum) is not reached p.31 ↗ Machine or system leaking Check the pipework and screw connections for leaks and check for tight fit Chapter 5.3
Final pressure (max. vacuum) is not reached p.31 ↗ Air filter is polluted Replace air filter Chapter 7.6
Machine gets too hot p.31 ↗ Ambient or inlet temperatures too high Ensure proper use Chapter 2.4
Machine gets too hot p.31 ↗ Cooling air supply is obstructed Check ambient conditions Chapter 5.1
Machine gets too hot p.31 ↗ Cooling air supply is obstructed Clean ventilation slots Chapter 7.4
The machine makes a strange noise p.31 ↗ Deposits on the rotary piston Clean the working space and the rotary piston Elmo Rietschle Service