BINDER KB PRO 130
BINDER KB PRO 130 — Incubators with Temperature -20…100 °C.
The KB PRO 130 is a 127 L single-phase cooling incubator with a −20…100 °C working range, 0.1 K fluctuation and 0.2 K uniformity, housed in a 900 mm wide cabinet. It is the smallest KB PRO model and the only one that uses a triangular instrument panel and requires a flexible tilt protection kit instead of a wall spacer. It is intended for temperature-controlled incubation and storage where precise, reproducible thermal conditions are required.
Direct drop-in replacements
BINDER KB PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Interior width, mm | Load per rack, kg |
|---|---|---|---|---|---|---|
| KB PRO 130 Current | -20–100 | 71 | 100.5 | 127 | 656 | 30 |
| KB PRO 1060 | -20–100 | 117.5 | 195.1 | 1070 | 1011 | 40 |
| KB PRO 1600 | -20–100 | 117.5 | 195.1 | 1610 | 1521 | 40 |
| KB PRO 260 | -20–100 | 90 | 139.6 | 262 | 656 | 40 |
| KB PRO 470 | -20–100 | 90 | 195.1 | 472 | 656 | 40 |
| KB PRO 720 | -20–100 | 90 | 195.1 | 727 | 1011 | 40 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 127 L |
| Load per rack | 30 kg |
| Max. heat compensation | 450 W |
| Nominal power | 1 kW |
| Permitted load | 150 kg |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm |
| Side wall clearance | 100 mm |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 2 min |
| Depth | 710 mm |
| Height | 1005 mm |
| Interior width | 656 mm |
| Width | 900 mm |
| Net weight of the unit (empty) mfr | 135 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The KB PRO 130 (Art. No. 9020-0433) is the entry size of the KB PRO cooling incubator series; all siblings (260, 470, 720, 1060, 1600) share the same −20…100 °C range but differ in chamber volume and panel architecture.
- Unlike the larger models, the 130 uses a triangular instrument panel rather than a signal triangle with adjustable control terminal.
- It is the only model (together with the 470) that requires the flexible tilt protection kit; the wall spacer required from size 260 upward does not apply.
- Interior volume is 127 L with net dimensions 656 mm wide × 710 mm deep × 1,005 mm high; net weight 135 kg.
Thermal behavior and load
- Temperature fluctuation is 0.1 K and uniformity 0.2 K across the working range.
- Recovery time after a 30 s door opening at 37 °C is 2 min.
- Maximum heat compensation is 450 W; permitted load is 150 kg total, with 30 kg per rack.
- The chamber is rated for a nominal temperature of 100 °C and uses a temperature safety device per DIN 12880:2007 class 2/3.3, configurable as class 2 (limiter) or class 3.3 (protection); class 3.2 undertemperature protection is also available.
Installation, control and operation
- Electrical supply is 208…240 V, 1~, 50 Hz, nominal power 1 kW, unit fuse 16.0 A; a CEE 7/7 plug is fitted.
- The MB2 chamber controller provides fixed-value, timer, time-program and week-program modes, plus hot-air disinfection and defrosting routines.
- Controller supports Ethernet, e-mail notification and USB data transfer; options include APT-COM 4 software, analog temperature outputs, zero-voltage alarm contacts, internal LED light, door lock and a flexible Pt 100 object temperature sensor.
- Racks are available with U-rails or telescopic rails as accessories.
Integration & Operating Notes
The KB PRO 130 requires the flexible tilt protection kit to be mounted; operating without it leaves the cabinet without the stability safeguard specified for this model. The kit is supplied separately and must be installed during commissioning.
A minimum wall clearance of 100 mm at the back and sides is required for heat dispersal. Do not install the chamber in unventilated recesses; insufficient clearance can cause overheating and invalidate thermal performance.
The unit is protected by a 16.0 A fuse and draws 1 kW nominal at 208…240 V single-phase. The supply circuit and upstream protection must be sized accordingly; verify voltage and frequency (50 Hz) before connection.
Zero-voltage alarm contacts and analog temperature outputs are not part of the standard delivery. If remote alarm monitoring or external recording is required, these options must be specified at order; the controller otherwise provides Ethernet and USB interfaces for data transfer.
The permitted load is 150 kg for the chamber and 30 kg per rack. Exceeding these limits can damage the rack supports or affect temperature uniformity; distribute loads evenly across racks.
Regular maintenance must be performed at least once a year, and an electrical safety test per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after any repair. A service message appears after 8760 operating hours or two years and must be reset by BINDER Service.
With increased dust in ambient air, clean the condenser several times a year by suction. The condenser is behind the machine room cover, which can be removed without tools by lifting upwards and pulling forwards. Refrigeration system work must be performed by personnel trained to EN 13313:2010.
Selection & Integration Q&A
Choose the 130 when 127 L interior volume and a 150 kg load capacity suit your workload and you want the compact 900 mm wide cabinet. Larger models (260 and up) add volume but require a wall spacer; the 130 instead uses a tilt protection kit.
Annual maintenance and electrical safety testing are required but not included. Racks with U-rails or telescopic rails, APT-COM 4 software, analog outputs, alarm contacts, LED light, door lock and Pt 100 sensor are options/accessories. A service message appears after 8760 h or two years.