BINDER MK 1020
BINDER ·Made in Germany · In Active Production

BINDER MK 1020

BINDER MK 1020 — Climate chambers with Temperature -40…180 °C.

The BINDER MK 1020 is a 991 L dynamic climate chamber for rapid temperature cycling from -40 °C to 180 °C, with average heating/cooling rates of 5.5/5.0 K/min per IEC 60068-3-5. As the largest MK series model, it is distinguished by its 1200 mm interior width and 40 kg per rack load capacity, and is the only MK size besides the MK 56 to offer water cooling. It is intended for reproducible temperature stress testing of large components and batches in a laboratory or production environment.

Temperature
-40…180 °C

BINDER MK Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior height, mmInterior volume, LInterior width, mm
MK 1020 Current -40–180148200.510209911200
MK 1020 CO₂ -48–180148200.510209911200
MK 115 -40–18086.5172.5480115600
MK 240 -40–18092.5171700228735
MK 240 CO₂ -50...18092.5171700228735
MK 400 -40–180140171700417735
MK 400 CO₂ -50...180140171700417735
MK 56 -40–18077.8144.542060400
MK 720 -40–180114200.510207341200
MK 720 CO₂ -48–180114200.510207341200

Full specifications

Core Performance & Functional Specifications
Temperature -40...180 °C
Interior volume 991 L
Load per rack 40 kg (MK1020-400V)
Nominal power 11.5 kW
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 1480 mm (MK1020-400V)
Height 2005 mm (MK1020-400V)
Interior height 1020 mm (MK1020-400V)
Interior width 1200 mm (MK1020-400V)
Width 1580 mm (MK1020-400V)
Net weight of the unit (empty) mfr 621 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 32.0 A
Phase (Nominal voltage) mfr 3 N/PE
Operating Conditions, Environment & Protection
Cooling down time from 180°C to -40°C mfr 120 min
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.5 K/min

Description

Construction and series positioning

Thermal behavior and load

Installation, control and operation

Integration & Operating Notes

Electrical
Requires 400 V three-phase supply with 32 A fusing

The MK 1020 draws 11.5 kW at 400 V 3 N/PE 50 Hz and must be protected by a 32 A fuse. Verify that the site provides a three-phase supply with adequate short-circuit capacity and that the neutral and protective earth are correctly bonded; an undersized supply will cause tripping during heating or cooling ramps.

Data Sheet Model MK 1020 en.pdf; Technical data
Mounting
Do not install in unventilated recesses or damp areas

The chamber requires sufficient ventilation for heat dispersal and must observe minimum distances to walls and ceilings. Installation in unventilated recesses can cause overheating and damage. Damp areas or puddles are prohibited due to electric shock hazard; the chamber must remain dry during operation, cleaning and maintenance.

Operating Manual MK/MKT (E5), Chapter 1.8, pages 12-13
Process Limits
No explosive atmospheres or combustible dust near the chamber

The chamber has no explosion protection measures and must not be operated in potentially explosive areas or with combustible dust or air-solvent mixtures present. Site classification and the selection of charging materials must exclude flammable substances at working temperature to prevent ignition.

Operating Manual MK/MKT (E5), Chapter 1.8, page 13
Control & I/O
Over-temperature class 1 trip indicates chamber failure

The over-temperature protective device class 1 responds if the nominal temperature is exceeded by 20 °C due to chamber failure. A class 1 trip requires the chamber to be taken out of operation immediately and BINDER Service to be informed; it is not a normal process alarm and must not be reset without investigating the cause.

Operating Manual MK/MKT (E5), Chapter 12.1, page 76; Chapter 22.3, page 120
Maintenance
Annual electrical safety inspection required

An annual repeat electrical inspection is required, as well as a test prior to initial startup and prior to resuming operation after maintenance or repair. Testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Schedule these inspections to maintain compliance and ensure electrical safety.

Operating Manual MK/MKT (E5), Chapter 22.1, page 118
Cooling & Media
Water cooling option available only for MK 56 and MK 1020

The water cooling option, with separate cooling water inlet and outlet connections, is available only for the MK 56 and MK 1020 models. If water cooling is required, it must be specified at ordering and the plumbing connections planned accordingly; other MK sizes cannot be retrofitted with this option.

Operating Manual MK/MKT (E5), Chapters 4.1 and 4.2, page 28

Selection & Integration Q&A

When should I choose the MK 1020 over the MK 240 or MK 115?

Choose the MK 1020 when you need a large interior volume (991 L) and higher load capacity (40 kg per rack) for testing large components or batches. The MK 240 and MK 115 offer smaller volumes and are limited to 400 V or 230 V respectively; the MK 1020 is the only MK size besides the MK 56 to offer water cooling, making it suitable for high-ambient or continuous-duty installations.

What installation clearances and environmental conditions are required?

The chamber must not be installed in unventilated recesses; sufficient ventilation for heat dispersal and minimum distances per chapter 3.4 must be observed. It must not be placed in damp areas or puddles, and must not be operated in hazardous locations or potentially explosive atmospheres. Ambient conditions should be free of combustible dust and air-solvent mixtures.

What maintenance and service intervals apply, and what accessories are not included?

A maintenance message appears after 8760 operating hours or two years, recurring every two weeks until reset by BINDER Service. An annual electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required. Base delivery excludes options such as RS485 interface, analog temperature outputs, flexible Pt 100 sensor, compressed air connection/dryer, water cooling, APT-COM™ 4 software, and over/under temperature safety device class 2; these must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model MK 1020 en
Datasheet ·832,2 КБ ·7 pp.
View PDF
PDF
MK+MKT (E5) 10 2026 en
Manual ·15,9 МБ ·169 pp.
View PDF