BINDER MKF 115
BINDER ·Made in Germany · In Active Production

BINDER MKF 115

BINDER MKF 115 — Climate chambers with Temperature -40…180 °C.

The BINDER MKF 115 is a 115-litre alternating climate chamber with a working range of -40 to 180 °C and 10 to 98 % RH, driven by a 4.8 kW 400 V 3-phase supply. Its 400 W heat compensation and 5.5 K/min heating rate distinguish it within the MKF series for rapid, humidified temperature cycling. It serves as a programmable test station for material and component qualification under combined temperature-humidity profiles.

Temperature
-40…180 °C

BINDER MKF Series — Model Comparison & Selection

Model Temperature, °CHeight, cmInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
MKF 115 Current -40–180172.5115304004.8
MKF 1020 -40–180200.599140150012.5
MKF 1020 CO₂ -48–180200.599140140011.5
MKF 240 -40–180171.5228304006.8
MKF 240 CO₂ -50...180171.3228309006.8
MKF 400 -40–180171417308008
MKF 400 CO₂ -50...180172.74173016007.9
MKF 56 -40–180144.560152502.8
MKF 720 -40–180200.573440100011
MKF 720 CO₂ -48–180200.573440110011

Full specifications

Core Performance & Functional Specifications
Temperature -40...180 °C
Interior volume 115 L (MKF115-400V, 50 Hz)
Load per rack 30 kg (MKF115-400V, 50 Hz)
Maximum heat compensation 400 W
Nominal power 4.8 kW (MKF115-400V, 50 Hz)
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Height 1725 mm (MKF115-400V, 50 Hz)
Width 980 mm (MKF115-400V, 50 Hz)
Electrical, Power & Drive Specifications
Power frequency 50 Hz (MKF115-400V)
Rated voltage 400 V
Unit fuse 16.0 A (MKF115-400V, 50 Hz)
Operating Conditions, Environment & Protection
Humidity range 10...98 % RH
Cooling down time from 180°C to -40°C mfr 100 min
Humidity fluctuation with humidity dependent on set value mfr 2.5 % r.F.
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.5 K/min

Description

Construction and series positioning

Behaviour under load and in different media

Installation, control and operation

Integration & Operating Notes

Electrical
Provide a 16 A fused 400 V 3-phase supply with neutral and earth

The chamber draws 4.8 kW nominal on a 3 N/PE 400 V 50 Hz supply and requires a 16.0 A unit fuse. Confirm the site distribution provides a neutral conductor and protective earth; the -C variant with voltage and frequency changer must be specified where 480 V 60 Hz is the only available supply.

Data Sheet Model MKF 115 en.pdf; Operating Manual MKF/MKFT (E5) 10/2026, Technical data
Mounting
Allow rear access for the power switch and maintenance flap

From size 115 on the chamber has a rear power switch in addition to the main power switch, and the condenser fan grid sits behind the left maintenance access flap. Both the rear switch and the fan grid must remain reachable, so leave clearance behind and beside the unit for commissioning and weekly fan-grid checks in dusty rooms.

Operating Manual MKF/MKFT (E5) 10/2026, Chamber description and Maintenance
Process Limits
Humidity control is limited to 10–95 °C and 10–98 % RH

Although the chamber reaches -40 to 180 °C dry, humidity operation is restricted to a 10 to 95 °C window with 10 to 98 % RH. Program set points that combine humidity with temperatures outside this band will not be controllable; plan test profiles so that humidified segments stay within the stated envelope.

Factory Specification Sheet; Operating Manual MKF/MKFT (E5) 10/2026, Technical data
Maintenance
Schedule annual maintenance, leak test and electrical inspection

The manual requires maintenance at least once a year, a leak test per Regulation (EU) 517/2014, and repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial start-up and after repair. The steam humidifier also needs annual servicing whose interval depends on local water quality and steam output.

Operating Manual MKF/MKFT (E5) 10/2026, Maintenance and service
Maintenance
Plan BINDER Service to reset the 8760-hour service message

After 8760 operating hours or two years the controller displays a service message; once acknowledged it reappears every two weeks until reset by BINDER Service. Budget for a service visit at that interval, and include condenser cleaning every one to two years and weekly fan-grid checks where ambient dust is elevated.

Operating Manual MKF/MKFT (E5) 10/2026, p.135

Selection & Integration Q&A

When should I choose the MKF 115 rather than the MKF 240 or the MKFT 115?

Pick the MKF 115 when 115 L and 30 kg per rack cover the specimen load and the profile stays within -40 to 180 °C with humidity between 10 and 95 °C. The MKF 240 offers more chamber volume for larger batches; the MKFT 115 is the deep-temperature variant for profiles below the MKF range.

What supply and water connections must be in place before commissioning an MKF 115?

A 400 V 3 N/PE 50 Hz supply fused at 16.0 A for the 4.8 kW load, plus a freshwater and wastewater connection for the humidification system (manual internal can from size 115 on, or automatic via water pipe). The -C variant is required where only 480 V 60 Hz is available.

How do I integrate the chamber with an external monitoring or alarm system?

Use the zero-voltage switching outputs for hard-wired alarm forwarding, or the optional RS485 interface, analog outputs for temperature and humidity, and Ethernet for network integration. APT-COM 4 software is available as an option for multi-unit management; these interfaces are not part of the base delivery.

What maintenance and consumables are required, and what is not included in the base delivery?

Annual maintenance, leak test per (EU) 517/2014, electrical inspection per EN 50678/50699, annual steam-humidifier service and condenser cleaning every 1–2 years. Consumables such as disinfectant spray Art. No. 1002-0022, the water cooling option, compressed air dryer and interface cables are options or accessories not supplied with the standard unit.