Peter Huber Kältemaschinenbau Minichiller 600 Kompakter
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Minichiller 600 Kompakter

Peter Huber Kältemaschinenbau Minichiller 600 Kompakter — Recirculating chillers.

The Minichiller 600 Kompakter is a compact recirculating chiller with a pressure-suction pump and Pilot ONE controller, covering -20 to 40 °C. It is the base 600 variant in the Minichiller series, distinguished from 1000-series siblings by its smaller cooling envelope. It serves as a point-of-use cooling source for laboratory and small process loops.

Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Minichiller 600 Kompakter Current —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D110-120V 1~ 60Hz
Minichiller 1000 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H Kompakter —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1200-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200w-H Kompakter —Pilot ONEIII / FL-20–100——
Minichiller 1200w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 280 OLÉ Kompakter 53OLÉI / NFL-5–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300 OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w-H OLÉ Kompakter 51OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 600-H OLÉ Kompakter 56OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 600 OLÉ Kompakter 56OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 60Hz
Minichiller 600w OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz

Full specifications

Core Performance & Functional Specifications
Maximum delivery flow rate 24 l/min (pressure pump max; 0, 7 bar)
Maximum delivery pressure (suction) 18 l/min; 0, 4 bar (suction pump max)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure- and suction pump
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 37 kg
Electrical, Power & Drive Specifications
Fuse rating 16 A
Power supply 110-120V 1~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Maximum return flow temperature 80 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Mounting
Install only in a dry, non-ATEX indoor location

The unit has IP20 protection and must not be installed in an ATEX zone. Ambient temperature must remain between 5 and 40 °C. Electrical lines directly below the unit must be protected against liquid ingress because condensation can form at the cooling water connections.

Minichiller Operation Manual, pages 13, 15, 28; Datasheet Minichiller 600, page 1
Maintenance
Install a drip tray and protect electrical lines below the unit

A drip tray for condensation water and thermal fluid must be installed below the unit. Electrical lines directly below the unit must be protected against liquid ingress. Do not install valves or quick-release couplings in the feed/discharge lines to the application.

Minichiller Operation Manual, pages 15, 28, 31

Selection & Integration Q&A

When should I choose the Minichiller 600 Kompakter instead of a 1000-series sibling?

Choose the 600 when the heat load and external circuit pressure drop are moderate and space is limited. The 1000-series siblings offer larger cooling capacity or high-pressure pump options for higher loads or higher pressure-drop circuits.

What are the requirements for the external fluid circuit and pump connections?

The pump connections are M16x1 male; hose connection NW12 with M16x1 sleeve nuts are included. The pressure-suction pump delivers up to 24 l/min at 0.7 bar pressure and 18 l/min at 0.4 bar suction, so external circuit resistance must stay within these limits. Do not install valves or quick-release couplings in the feed/discharge lines.

What maintenance and accessories are required that are not included in the base delivery?

A drip tray for condensation water and thermal fluid must be provided by the installer. A country-specific plug must be specified at ordering. For freeze protection, glysantine or ethylene glycol must be added to the water; anti-corrosion agents are mandatory if salts are present. The plastic pump is not resistant to acetone or polyglycols.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Minichiller Operation Manual
Manual ·2,9 МБ ·84 pp.
View PDF
PDF
Minichiller 600
Datasheet ·30,6 КБ ·2 pp.
View PDF