Peter Huber Kältemaschinenbau Minichiller 1200-H OLÉ Kompakter
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Minichiller 1200-H OLÉ Kompakter

Peter Huber Kältemaschinenbau Minichiller 1200-H OLÉ Kompakter — Recirculating chillers.

The Minichiller 1200-H OLÉ Kompakter is a compact recirculating chiller with a pressure/suction pump and OLÉ controller, covering -20 to 100 °C. It is the 1.2 kW-class air-cooled modification in the Minichiller series, distinguished by its R-1270 (A3) refrigerant and fixed-speed pump. It serves as a point-of-use cooling source for laboratory and pilot processes requiring stable temperature control.

Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Minichiller 1200-H OLÉ Kompakter Current —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H Kompakter —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1200 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200w-H Kompakter —Pilot ONEIII / FL-20–100——
Minichiller 1200w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 280 OLÉ Kompakter 53OLÉI / NFL-5–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300 OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w-H OLÉ Kompakter 51OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 600-H OLÉ Kompakter 56OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 600 Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D110-120V 1~ 60Hz
Minichiller 600 OLÉ Kompakter 56OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 60Hz
Minichiller 600w OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz

Full specifications

Core Performance & Functional Specifications
Maximum delivery flow rate 24 l/min (pressure pump; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure- and suction pump
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 38 kg
Electrical, Power & Drive Specifications
Current consumption 12 A
Fuse rating 16 A
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and capacity

Installation, control and operation

Integration & Operating Notes

Mounting
Install a drip tray and protect cables below the unit

Condensation and thermal fluid can escape at the pump connections and drain area. The responsible body must install a condensation water/thermal fluid drip tray below the unit, and any electrical lines routed directly underneath must be protected against liquid ingress to avoid shock hazard.

Minichiller Operation Manual, Installation (p.15, p.28
Cooling & Media
Use drinking or filtered water only; no sea water

The cooling water circuit accepts drinking water or filtered river/sea water, but sea water must not be used. If salts such as chlorides or bromide are added, anti-corrosion agents are mandatory to prevent corrosion damage to the circuit.

Minichiller Operation Manual, Installation (p.28
Electrical
Verify supply tolerance combination before commissioning

The unit accepts ±5% voltage and ±2% frequency, but the combination of -5% voltage with +2% frequency is explicitly not allowed. Check site supply quality against this rule; also provision a 16 A fuse and specify the country-specific plug at ordering.

Minichiller 1200-H OLÉ Technical Data Sheet, Specifications (p.2
Process Limits
Avoid acetone and polyglycols with the standard plastic pump

The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Select an alternative fluid or a more resistant pump option if the process requires these media, otherwise pump life and leak integrity are at risk.

Minichiller 1200-H OLÉ Technical Data Sheet, Specifications (p.2
Maintenance
Order the drain valve separately for fluid changes

A drain valve (#6839) is an optional accessory, not part of base delivery. Without it, gravity drainage for fluid replacement and maintenance is not possible, so plan the accessory into the initial order if routine fluid changes are expected.

Minichiller 1200-H OLÉ Technical Data Sheet, Specifications (p.2
Environment
Do not install in an ATEX zone or wet/hot location

Operation in an ATEX zone can cause explosion and is prohibited. IP20 protection and the 5-40 °C ambient limit mean the unit cannot be placed in wet or hot environments; also check refrigerant charge against room volume to stay below the limit value per m³.

Minichiller Operation Manual, Installation (p.13, p.16); Technical Data Sheet (p.2

Selection & Integration Q&A

When should I choose the 1200-H OLÉ over the 1000-series Minichiller siblings?

Pick the 1200-H OLÉ when your heat load needs the higher capacity class and you want the OLÉ controller; the 1000-series siblings (1000-H, 1000-H OLÉ, 1000 OLÉ, 1000w-H) cover lower loads. Size against actual set point: capacity falls from 1.2 kW at 15 °C to 0.35 kW at -20 °C.

What must the application loop and power supply provide to integrate this chiller?

Loop fittings must match M16x1 male pump connections, with no valves or quick-release couplings in the feed/discharge lines. Supply must be 208-240 V 1~/2~ 50/60 Hz, 16 A fuse, max 12 A, with country-specific plug specified at ordering; USB and RS232 are available for control.

What site and duty-cycle limits apply to mounting and operation?

Install only in 5-40 °C ambient, non-ATEX, dry locations (IP20); a drip tray is required below the unit and cables underneath must be protected from liquid. Do not move the unit while operating. Whisper mode cuts flow to 14 l/min at 0.2 bar, so verify heat load before enabling it.

What maintenance accessories and consumables are not included in base delivery?

The drain valve (#6839) is optional and must be ordered separately for gravity fluid changes. Freeze protection requires glysantine or ethylene glycol mixed with water; anti-corrosion agents are mandatory if chlorides or bromide are added. The power cable and country-specific plug must be specified at ordering.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Minichiller Operation Manual
Manual ·2,9 МБ ·84 pp.
View PDF
PDF
Minichiller 1200-H OLÉ Technical Data Sheet
Datasheet ·106,7 КБ ·2 pp.
View PDF