Peter Huber Kältemaschinenbau Minichiller 1000w Kompakter
Peter Huber Kältemaschinenbau Minichiller 1000w Kompakter — Recirculating chillers.
The Minichiller 1000w Kompakter is a water-cooled recirculating chiller with a pressure/suction pump and Pilot ONE controller, covering -20 to 40 °C. It is the 'w' (water-cooled) variant of the 1000 series, requiring 3–6 bar cooling-water pressure. It serves as a compact cooling source for laboratory and pilot processes needing up to 24 l/min flow.
Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Minichiller 1000w Kompakter Current | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | — | — |
| Minichiller 1200w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 280 OLÉ Kompakter | 53 | OLÉ | I / NFL | -5–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300 OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w-H OLÉ Kompakter | 51 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 600-H OLÉ Kompakter | 56 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 600 Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 110-120V 1~ 60Hz |
| Minichiller 600 OLÉ Kompakter | 56 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 60Hz |
| Minichiller 600w OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
Full specifications
| Maximum delivery flow rate | 24 l/min (max.; pressure 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure- and suction pump |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Maximum return flow temperature | 80 °C (max.) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The Minichiller 1000w Kompakter is a water-cooled recirculating chiller in the Minichiller series. Unlike air-cooled siblings (e.g., Minichiller 1000-H, 1000 OLÉ), this 'w' modification uses a water-cooled refrigeration circuit, making it dependent on an external cooling-water supply. The pump is a pressure- and suction type with adjustable speed, connected via M16x1 male threads. Cooling-water connections are G1/2 male. The unit is controlled by the Pilot ONE controller and offers Ethernet, USB (host and device), and RS232 interfaces. Protection rating is IP20, safety class I/NFL.
Operating behavior and media
- Temperature range: -20 to 40 °C; maximum return-line temperature 80 °C.
- Pump delivery: up to 24 l/min at 0.7 bar in normal mode; whisper mode reduces to 14 l/min at 0.2 bar.
- Cooling water: differential pressure min. 3 bar, supply pressure max. 6 bar; intake temperature 5–20 °C. Cooling power drops as cooling-water temperature rises.
- Refrigerant: R-290 (A3 flammable).
- The plastic pump is not resistant to acetone or polyglycols. For freeze protection below 0 °C, mix water with glysantine or ethylene glycol.
Installation and operation
Install indoors, protected from water ingress (IP20), with a drip tray below the unit. Do not install in ATEX zones. Electrical lines below the unit must be protected against liquid ingress. Do not install valves or quick-release couplings in feed/discharge lines. The unit must not be moved while operating. Cooling water must be drinking water or filtered river/sea water; sea water is prohibited. Anti-corrosion agents are mandatory if salts are added.
Integration & Operating Notes
The water-cooled refrigeration circuit requires a minimum differential pressure of 3 bar to function; lower pressure will prevent proper cooling and may cause the compressor to trip. Verify the building chilled-water supply can sustain this differential at the required flow.
The cooling-water supply pressure must not exceed 6 bar. If the building water line is higher, install a pressure reducer upstream of the G1/2 male connection to avoid overpressure damage to the water-cooled condenser.
Connected piping, hoses, and components in the return line must be rated for at least 80 °C. Exceeding this limit can damage the chiller and pose a burn hazard. Ensure the application's heat load does not raise the return fluid above this value.
The pump's plastic materials are not resistant to acetone or polyglycols (depending on manufacturer). Using these fluids will degrade the pump and void reliable operation. Select an alternative fluid or a different pump material for such media.
The responsible body must install a condensation water / thermal fluid drip tray below the temperature control unit to contain leaks and condensation. This is required to protect the installation area and prevent slip hazards.
Operation within an ATEX zone can cause explosion due to the flammable R-290 refrigerant. The unit must be installed in a non-hazardous area with adequate ventilation per the refrigerant charge and room size.
The power supply must be within ±5% voltage and ±2% frequency simultaneously; certain combinations are not allowed. Check site power quality before installation to avoid erratic operation or damage to the controller and compressor.
If chlorides, bromide, or other salts are added to the water circuit, anti-corrosion agents are mandatory to protect the copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM materials from corrosion damage.
Selection & Integration Q&A
Choose the 'w' water-cooled version when a reliable chilled-water supply (3–6 bar differential) is available and air-cooled units would reject too much heat to the room. Air-cooled siblings are preferable where no cooling-water infrastructure exists.
Cooling water: G1/2 male, differential pressure min. 3 bar, supply pressure max. 6 bar, intake 5–20 °C. Process circuit: M16x1 male pump connections; do not install valves or quick-release couplings in feed/discharge lines. Return line must withstand 80 °C.
Ambient 5–40 °C, indoor dry location (IP20), not in ATEX zone. Install a drip tray below. Protect electrical lines below the unit from liquid ingress. Ensure room volume and refrigerant charge (R-290) comply with safety limits.
A drip tray must be installed by the responsible body. Anti-corrosion agents are required if salts are added. For freeze protection below 0 °C, mix water with glysantine or ethylene glycol. No cables or hoses are included; order M16x1 and G1/2 fittings separately.