Peter Huber Kältemaschinenbau Minichiller 300w OLÉ Kompakter
Peter Huber Kältemaschinenbau Minichiller 300w OLÉ Kompakter — Recirculating chillers with Noise level 53 dB.
The Minichiller 300w OLÉ Kompakter is a water-cooled recirculating chiller with a pressure/suction pump and OLÉ controller, operating from -20 to 40 °C with a maximum return temperature of 80 °C. It is distinguished from larger siblings (1000 series) by its compact 300-class capacity and fixed-speed pump. It serves as a point-of-use cooling source for laboratory and small process circuits requiring precise temperature control.
Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Minichiller 300w OLÉ Kompakter Current | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 1000 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | — | — |
| Minichiller 1200w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 280 OLÉ Kompakter | 53 | OLÉ | I / NFL | -5–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300 OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w-H OLÉ Kompakter | 51 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 600-H OLÉ Kompakter | 56 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 600 Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 110-120V 1~ 60Hz |
| Minichiller 600 OLÉ Kompakter | 56 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 60Hz |
| Minichiller 600w OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
Full specifications
| Noise level | 53 dB(A) |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure- and suction pump |
| Temperature range | -20...40 °C |
| Weight | 23 kg |
| Current consumption | 2 A (max) |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 15 l/h (flow 15 °C) |
| Cooling water differential pressure | 3 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Maximum return flow temperature | 80 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and Series Positioning
- Water-cooled refrigerating unit using R-290 (A3, H220) refrigerant; requires an external cooling water circuit with G1/2 male connections.
- Pressure- and suction pump with M16x1 AG male process connections; pump speed is not adjustable (fixed-speed).
- OLÉ controller with OLED display, 0.1 K resolution, USB (Device) and RS232 interfaces.
- Compact 23 kg unit, IP20 protection, designed for indoor installation at 5–40 °C ambient.
- Within the Minichiller series, this 300w OLÉ model sits below the 1000 and 1000-H variants in cooling capacity; the 'w' denotes water-cooled, and 'OLÉ' the controller variant. Siblings include 1000, 1000-H, 1000 OLÉ, and 1000w-H models.
Behavior Under Load and Media
- Temperature range -20 to 40 °C, but return line temperature must not exceed 80 °C.
- Cooling water inlet temperature must be 5–20 °C; higher inlet temperatures reduce cooling power and may increase required flow.
- Cooling water consumption is 15 l/h at 15 °C flow, with a minimum differential pressure of 3 bar and maximum pressure of 6 bar.
- The plastic pump is not resistant to acetone or polyglycols; freeze protection requires glycol or glysantine addition.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM; only drinking water or filtered river/sea water (no sea water) may be used.
Installation, Control, and Operation
- Must not be installed in ATEX zones; a drip tray for condensation/thermal fluid is required below the unit.
- Electrical lines below the unit must be protected against liquid ingress due to condensation.
- Do not install valves or quick-release couplings in feed/discharge lines to the application.
- Power supply: 220–240 V 1~/2~ 50/60 Hz, max current 2 A; voltage tolerance ±5% with simultaneous frequency tolerance ±2%.
- Country-specific plug must be specified at ordering; unit emits Class A, Group 1 disturbances per EN55011.
Integration & Operating Notes
The unit must not be placed in an ATEX zone due to explosion risk. A drip tray for condensation water and thermal fluid must be installed below the unit to contain leaks. Electrical lines directly below must be protected against liquid ingress.
Although the operating range is -20 to 40 °C, the permissible temperature in the return line is 80 °C. Exceeding this can overheat the unit. Also, cooling water inlet must be 5–20 °C; higher temperatures reduce cooling power and may require increased flow.
The water-cooled refrigerating unit requires a cooling water supply with differential pressure between 3 and 6 bar at G1/2 male connections. Consumption is 15 l/h at 15 °C flow. Only drinking water or filtered river/sea water may be used; sea water is prohibited. Anti-corrosion agents are mandatory if salts are added.
The power supply must be 220–240 V 1~/2~ 50/60 Hz with voltage deviation of ±5% and simultaneous frequency tolerance of ±2%. Opposite-sign deviations are not allowed. Max current consumption is 2 A. A country-specific plug must be specified at ordering.
The circulation pump is made of plastic and is not resistant to acetone or polyglycols (depending on manufacturer). Alternative fluids or a more resistant pump must be used. For freeze protection, water should be mixed with glysantine or ethylene glycol.
Do not install valves or quick-release couplings in the feed/discharge lines between the unit and the application, as this can cause overpressure damage, especially to glass apparatus. The pressure/suction pump has M16x1 AG male connections.
If salts such as chlorides or bromide are added to the water circuit, the addition of anti-corrosion agents is mandatory to prevent corrosion damage. Regular maintenance should include checking water quality and refrigerant charge against room size limits.
Selection & Integration Q&A
Choose the 300w OLÉ for lower cooling capacity and compact footprint where the 1000 series is oversized. The 1000 and 1000-H variants offer higher capacity; the 'H' may indicate heating. Select based on required cooling power and whether heating is needed.
It requires a cooling water circuit with G1/2 male connections, differential pressure 3–6 bar, and inlet temperature 5–20 °C. Consumption is 15 l/h at 15 °C. Only drinking or filtered river water; sea water prohibited. Anti-corrosion agents needed if salts added.
Install at 5–40 °C ambient, IP20, non-ATEX zone, with a drip tray below. Return line temperature must not exceed 80 °C. Do not move while operating. Electrical lines below must be protected from liquid ingress.
A drip tray for condensation/thermal fluid must be provided by the responsible body. Anti-corrosion agents are mandatory if salts are added. Country-specific plug must be specified at ordering. No valves or quick-release couplings in process lines.