Peter Huber Kältemaschinenbau Minichiller 1000w-H OLÉ Kompakter
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Minichiller 1000w-H OLÉ Kompakter

Peter Huber Kältemaschinenbau Minichiller 1000w-H OLÉ Kompakter — Recirculating chillers.

The Minichiller 1000w-H OLÉ Kompakter is a compact recirculating chiller with a pressure-suction pump and OLÉ controller, covering -20 to 100 °C. Its 'w' suffix denotes a water-cooled condenser, distinguishing it from air-cooled siblings. It serves as a precise temperature-control source for laboratory and pilot processes requiring a stable fluid supply.

Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Minichiller 1000w-H OLÉ Kompakter Current —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H Kompakter —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1000w Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1200-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200w-H Kompakter —Pilot ONEIII / FL-20–100——
Minichiller 1200w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 280 OLÉ Kompakter 53OLÉI / NFL-5–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300 OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w-H OLÉ Kompakter 51OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 600-H OLÉ Kompakter 56OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 600 Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D110-120V 1~ 60Hz
Minichiller 600 OLÉ Kompakter 56OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 60Hz
Minichiller 600w OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz

Full specifications

Core Performance & Functional Specifications
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure- and suction pump
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 37 kg
Electrical, Power & Drive Specifications
Current consumption 12 A (max)
Fuse rating 16 A
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 15 kW (at 1 °C)
Cooling water connection G1/2 male
Cooling water consumption 24 л/час (at 15°C, flow -20°C)
at 15°C, flow -10°C: 30 л/час · at 15°C, flow 0°C: 36 l/h · at 15°C, flow 15°C: 39 l/h
Cooling water differential pressure 3 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and operation

Integration & Operating Notes

Mounting
Install a drip tray below the chiller

Condensation water and thermal fluid can escape; the responsible body must install a drip tray below the temperature control unit to contain leaks and prevent slip hazards.

Minichiller Operation Manual, page 15
Electrical
Protect electrical lines below the unit from liquid

Condensation can form at the cooling water connections; electrical lines directly below the temperature control unit must be protected against liquid ingress to avoid shock hazard.

Minichiller Operation Manual, page 28
Process Limits
Cooling water must be 5–20 °C and 3–6 bar differential

The chiller requires cooling water with a differential pressure between 3 and 6 bar and an inlet temperature between 5 °C and 20 °C; outside these limits cooling capacity drops or condensation and corrosion risk increases.

Minichiller 1000w-H OLÉ Technical Data, page 2
Process Limits
Avoid valves or quick-release couplings in application lines

Do not install valves or quick-release couplings in the feed/discharge lines to and from the application; they can cause overpressure damage, especially to glass apparatus.

Minichiller Operation Manual, page 31
Cooling & Media
Use only drinking water or filtered river/sea water

Sea water must not be used for water cooling; only drinking water or filtered river/sea water is permitted. Anti-corrosion agents are mandatory if salts such as chlorides or bromide are added.

Minichiller Operation Manual, page 28
Safety
Do not install in an ATEX zone

The unit must not be installed or started up within an ATEX zone because operation there can cause explosion; ensure the installation location is outside hazardous areas.

Minichiller Operation Manual, page 13
Safety
Flammable refrigerant R-290 requires ventilation check

The unit uses R-290 (A3, H220) refrigerant; observe the refrigerant charge and room size to avoid exceeding the limit value per m³ of ambient air, and avoid ignition sources.

Minichiller 1000w-H OLÉ Technical Data, page 1
Maintenance
Monitor fill level via sight glass

The unit has a level indicator with sight glass; low fill level can cause pump cavitation and poor temperature control, so check regularly during operation.

Minichiller 1000w-H OLÉ Technical Data, page 1

Selection & Integration Q&A

When should I choose the water-cooled 1000w-H OLÉ over the air-cooled 1000-H OLÉ?

Choose the 'w' variant when a facility cooling-water supply with 3–6 bar differential and 5–20 °C inlet is available; it rejects heat to water and avoids room heat load. The air-cooled sibling is preferable where no water supply exists.

What are the requirements for the cooling-water circuit and electrical supply?

Cooling water must be drinking or filtered river/sea water (no sea water) at 5–20 °C and 3–6 bar differential, with connections G1/2 male. Electrical supply is 208–240 V, 1~/2~, 50/60 Hz, 12 A max, 16 A fuse.

What ambient and installation limits apply from the manual?

Ambient temperature must be 5–40 °C, IP20 protection, no ATEX zone. Install a drip tray below, protect electrical lines below from liquid, and avoid moving the unit while operating.

What maintenance and accessories are required beyond base delivery?

A drip tray is mandatory. Anti-corrosion agents are required if salts are added to the water circuit. The plastic pump is not resistant to acetone or polyglycols. USB and RS232 cables are not included; order separately.