Peter Huber Kältemaschinenbau Minichiller 800w-H OLÉ Kompakter
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Minichiller 800w-H OLÉ Kompakter

Peter Huber Kältemaschinenbau Minichiller 800w-H OLÉ Kompakter — Recirculating chillers.

The Minichiller 800w-H OLÉ Kompakter is a water-cooled recirculating chiller with a pressure-and-suction pump and OLÉ controller, covering -20 to 100 °C. Its G1/2 cooling-water connections distinguish it from the 1000-series siblings. It serves as a compact process cooling source for closed loops requiring both pressure and suction operation.

Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Minichiller 800w-H OLÉ Kompakter Current —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H Kompakter —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1200-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200w-H Kompakter —Pilot ONEIII / FL-20–100——
Minichiller 1200w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 280 OLÉ Kompakter 53OLÉI / NFL-5–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300 OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w-H OLÉ Kompakter 51OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 600-H OLÉ Kompakter 56OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 600 Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D110-120V 1~ 60Hz
Minichiller 600 OLÉ Kompakter 56OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 60Hz
Minichiller 600w OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz

Full specifications

Core Performance & Functional Specifications
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure- and suction pump
Temperature range -20...100 °C
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 24 l/h (at 15°C, flow -20°C)
at 15°C, flow -10°C: 30 l/h · at 15°C, flow 0°C: 36 l/h · at 15°C, flow 15°C: 39 l/h
Cooling water differential pressure 3 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Cooling water inlet must stay within 5–20 °C

Operating outside this range or below the dew point requires insulation of the cooling water pipes inside the unit; otherwise condensation and reduced performance occur. The datasheet explicitly designs the unit for 5–20 °C cooling water intake.

Minichiller 800w-H OLÉ Technical Data, page 2
Electrical
Voltage and frequency tolerances are interdependent

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed. Site power quality must be checked against this envelope before connection.

Minichiller 800w-H OLÉ Technical Data, page 2
Electrical
Three-phase units below 63 A arrive with cable but no plug

The unit requires 208–240 V single/two-phase 50/60 Hz supply; three-phase devices with current consumption less than 63 A are supplied with cable but without plug, so a suitable plug must be provided on site.

Minichiller 800w-H OLÉ Technical Data, page 2
Mounting
Do not install in an ATEX zone

Operation in a hazardous area can cause explosion. The unit must not be installed or started up within an ATEX zone, and the refrigerant R-290 (A3, flammable) requires attention to ignition sources and ventilation in the installation area.

Minichiller Operation Manual, page 13
Mounting
Provide a drip tray below the unit

The responsible body must install a condensation water / thermal fluid drip tray below the temperature control unit to contain leaks. Electrical lines directly below the unit must also be protected against liquid ingress because condensation can form.

Minichiller Operation Manual, page 15
Process Limits
Avoid valves and quick-release couplings in application lines

Do not install valves or quick-release couplings in the feed/discharge lines from the temperature control unit to the application and from the application to the temperature control unit, to avoid overpressure damage, especially to glass apparatus.

Minichiller Operation Manual, page 31
Cooling & Media
Sea water must not be used for cooling

Only drinking water or filtered river/sea water may be used as cooling water; sea water must not be used. Anti-corrosion agents are mandatory if salts such as chlorides or bromide are added to the water circuit.

Minichiller Operation Manual, page 28
Cooling & Media
Plastic pump is not resistant to acetone and polyglycols

The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer), so these media must be avoided or a more resistant pump specified. For freeze protection below 0 °C, water should be mixed with glysantine or ethylene glycol.

Minichiller 800w-H OLÉ Technical Data, page 2
Maintenance
Check refrigerant charge against room volume

Observe the amount of refrigerant contained (see data sheet/rating plate) and the room size when installing the unit to avoid exceeding the refrigerant limit value per m³ of ambient air; additional safety measures may be required.

Minichiller Operation Manual, page 16

Selection & Integration Q&A

When should I choose the Minichiller 800w-H OLÉ instead of a 1000-series sibling?

Choose the 800w-H OLÉ when the required cooling capacity fits the 800 W class and you need the pressure-and-suction pump with OLÉ controller. The 1000-series siblings offer a higher capacity class; select them if your load exceeds the 800-class capability. The 'w' denotes water cooling, so a cooling water supply is mandatory for all these variants.

What cooling water supply and process connections must be provided for this model?

Provide cooling water at 5–20 °C with at least 3 bar differential pressure (max 6 bar); connections are G1/2 male, so mating hoses/fittings must be G1/2 female. The process pump connection is M16x1 male, so process hoses must match this thread. Only drinking water or filtered river water may be used; sea water is prohibited.

What are the mounting environment and duty-cycle limits from the manual?

Install indoors at 5–40 °C ambient, not in an ATEX zone, with a drip tray below the unit and protection of electrical lines against condensation. The unit is IP20 and must not be exposed to water spray. Do not move it while operating. Power supply is 208–240 V 1~/2~ 50/60 Hz with interdependent voltage/frequency tolerances.

What maintenance and accessories are required beyond the base delivery?

The base delivery includes cable but no plug for three-phase units below 63 A. A drip tray must be provided on site. Anti-corrosion agents are mandatory if salts are added to the water circuit. For sub-zero operation, glysantine or ethylene glycol must be added for freeze protection. Check refrigerant charge against room volume per the manual.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Minichiller Operation Manual
Manual ·2,9 МБ ·84 pp.
View PDF
PDF
Minichiller 800w-H OLÉ Technical Data
Datasheet ·107,4 КБ ·2 pp.
View PDF