Peter Huber Kältemaschinenbau Minichiller 600 Kompakter
Peter Huber Kältemaschinenbau Minichiller 600 Kompakter — Recirculating chillers.
The Minichiller 600 Kompakter is a compact recirculating chiller with a pressure-suction pump and Pilot ONE controller, covering -20 to 40 °C. It is the base 600 variant in the Minichiller series, distinguished from 1000-series siblings by its smaller cooling envelope. It serves as a point-of-use cooling source for laboratory and small process loops.
Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Minichiller 600 Kompakter Current | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 110-120V 1~ 60Hz |
| Minichiller 1000 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | — | — |
| Minichiller 1200w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 280 OLÉ Kompakter | 53 | OLÉ | I / NFL | -5–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300 OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w-H OLÉ Kompakter | 51 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 600-H OLÉ Kompakter | 56 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 600 OLÉ Kompakter | 56 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 60Hz |
| Minichiller 600w OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
Full specifications
| Maximum delivery flow rate | 24 l/min (pressure pump max; 0, 7 bar) |
| Maximum delivery pressure (suction) | 18 l/min; 0, 4 bar (suction pump max) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure- and suction pump |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 37 kg |
| Fuse rating | 16 A |
| Power supply | 110-120V 1~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Maximum return flow temperature | 80 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- The Minichiller 600 Kompakter is a compact, air-cooled recirculating chiller with an integrated pressure-suction pump and the Pilot ONE controller. It is the smallest capacity variant in the Minichiller 600 line; sibling models (Minichiller 1000-H, 1000-H OLÉ, 1000 OLÉ, 1000w-H) offer larger cooling capacities or high-pressure pump options. This model is intended for applications with moderate pressure drop and limited space.
- The pump connections are M16x1 male, and the unit is supplied with hose connection NW12 and M16x1 sleeve nuts. The pump is speed-controlled, allowing the flow or pressure to be matched to the application.
Behavior under load and media
- The temperature range is -20 to 40 °C, with a maximum permissible return-line temperature of 80 °C. The pressure-suction pump delivers up to 24 l/min at 0.7 bar pressure and 18 l/min at 0.4 bar suction, which bounds the external circuit resistance.
- The plastic pump is not resistant to acetone or polyglycols; these fluids must not be used. For freeze protection, a mixture of water with glysantine or ethylene glycol is recommended. The minimum filling capacity is 2.8 l and the expansion volume is 2.2 l.
- The refrigerant is R-290 (propane, A3 flammable), so ignition sources and ventilation must be considered at the installation location.
Installation, control and operation
- The unit requires a 110–120 V single-phase 60 Hz supply with a 16 A fuse; maximum current is 8.5 A. A country-specific plug must be specified at ordering. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%.
- The Pilot ONE controller provides Ethernet, USB (Host and Device) and RS232 interfaces, and a Pt100 external sensor connection. The unit has IP20 protection and must be installed in a dry indoor location. Ambient temperature must stay between 5 and 40 °C; cooling capacity is rated at 20 °C ambient and drops as ambient rises.
- A drip tray for condensation water and thermal fluid must be installed below the unit. Electrical lines directly below the unit must be protected against liquid ingress. Do not install valves or quick-release couplings in the feed/discharge lines to the application. The unit must not be installed in an ATEX zone.
Integration & Operating Notes
The unit has IP20 protection and must not be installed in an ATEX zone. Ambient temperature must remain between 5 and 40 °C. Electrical lines directly below the unit must be protected against liquid ingress because condensation can form at the cooling water connections.
A drip tray for condensation water and thermal fluid must be installed below the unit. Electrical lines directly below the unit must be protected against liquid ingress. Do not install valves or quick-release couplings in the feed/discharge lines to the application.
Selection & Integration Q&A
Choose the 600 when the heat load and external circuit pressure drop are moderate and space is limited. The 1000-series siblings offer larger cooling capacity or high-pressure pump options for higher loads or higher pressure-drop circuits.
The pump connections are M16x1 male; hose connection NW12 with M16x1 sleeve nuts are included. The pressure-suction pump delivers up to 24 l/min at 0.7 bar pressure and 18 l/min at 0.4 bar suction, so external circuit resistance must stay within these limits. Do not install valves or quick-release couplings in the feed/discharge lines.
A drip tray for condensation water and thermal fluid must be provided by the installer. A country-specific plug must be specified at ordering. For freeze protection, glysantine or ethylene glycol must be added to the water; anti-corrosion agents are mandatory if salts are present. The plastic pump is not resistant to acetone or polyglycols.