The unit must not be installed in an ATEX zone because operation there can cause explosion.
Determines whether the unit can be placed in a hazardous area at all.
Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
| Description | (A) | (B) | (C) | (D) | (E) | (F) |
|---|---|---|---|---|---|---|
| Acoustics | X | X | X | – | – | – |
| Display functions Warnings; Messages; Inactive Menu Items; Resolution Temperature | X | X | X | – | – | – |
| Display modes | X | X | X | – | – | – |
| AutoStart response | X | X | X | – | – | – |
| Limitations DeltaT limit; max. heating variable; max. cooling variable; max. permissible current consumption (for 230 V temperature control units) | X | X | X | – | – | – |
| Com.G@te (only if Com.G@te is connected) Analog interface (configuration input) Meaning of AIF input; response on cable break; measurement range setting | X | X | X | – | – | X |
| Calibration | – | – | – | – | – | X |
| In case of analog error; current / voltage switching; filter constant | X | X | X | – | – | X |
| Configuration output AIF output variable; output range setting | X | X | X | – | – | X |
| Calibration | – | – | – | – | – | X |
| RS232/RS485 (hardware RS, Baud rate, device address); ECS standby; POKO alarm | X | X | X | – | – | – |
| Properties (other) Change thermal fluid (thermal fluid, liter indicator, bypass usage); cooling bath / cooling (CC-E only) | X | X | X | – | – | – |
| Sensor calibration | – | – | – | – | – | – |
| Automated compressor | X | X | X | – | – | – |
| Enter program | X | – | – | X | – | – |
| Pump settings Pump speed setpoint; pump pressure setpoint; control mode | X | X | X | – | – | – |
| Fluid properties thermal fluid; liter indication; pressure application (only with XT models) | X | X | X | – | – | – |
| Configuration manual parameters |
Таблица напечатана как матрица: строки — характеристики (Description), колонки — варианты сброса (A)–(F). Транспонирование не требуется.
| Function | E-grade Basic | E-grade Exclusive | E-grade Professional |
|---|---|---|---|
| Temperature control | |||
| Controller parameterization: predefined1 / TAC2 | X/– | –/X | –/X |
| Find parameters: Fast Identification / With Preliminary Test / Estimate Control Parameters | –/–/X | X/X/– | X/X/– |
| Sensor calibration for control sensor3: x -point | 2 | 5 | 5 |
| Monitoring: Low level and overtemperature4 | X | X | X |
| Adjustable alarm thresholds | X | X | X |
| VPC5 (Variable Pressure Control) | X | X | X |
| Ventilation program | X | X | X |
| Automated compressor | X | X | X |
| Setpoint limitation | X | X | X |
| Program controller: x programs / max x steps | –/– | 3/15 | 10/100 |
| Ramp function: linear / non-linear | –/– | X/– | X/X |
| Temperature control mode: Internal and process | – | X | X |
X = standard equipment, O = optional, – = not possible. 1 TAC function available as a 30 day evaluation version. 2 True Adaptive Control. 3 Internal Pt100 and External Pt100. 4 For temperature control units with integrated overtemperature protection. 5 For units with variable-speed pump or external bypass.
| Designation | Specification |
|---|---|
| Calcium carbonate per liter | ≤ 1.5 mmol/l; corresponds to a water hardness of: ≤ 8.4 °dH (soft) |
| pH value | between 6.0 and 8.5 |
| Ultrapure water, distillates | Add 0.1 g of sodium carbonate (Na2CO3) per liter |
| Not approved water | Distilled, deionized, demineralized, chloric, ferruginous, ammoniacal, contaminated or untreated river water or sea water |
| Volume circulated (at least) | 3 l/min. |
| Thermal fluid: Water without ethylene glycol | |
| Use | ≥ +3 °C |
| Thermal fluid: Water-ethylene glycol mixture | |
| Use | < +3 °C |
| Thermal fluid composition | The mixture’s temperature must be 10 K below the permissible min. temperature. For the permissible temperature range, refer to the datasheet. → From page 82, section »Annex«. |
| Description | (A) | (B) | (C) | (D) | (E) | (F) | |
|---|---|---|---|---|---|---|---|
| Protective function | High limit alarm internal; low limit alarm internal; high limit alarm process; low limit alarm process; hydrostatic correction; warning time limit (CC-E only); min. level (for temperature control units with analogue level sensor); max. level (for temperature controllers with analogue level sensor); freeze protection (optional) | X | X | X | – | – | – |
| Setpoint | X | X | X | – | – | – | |
| Setpoint limitation | Min. setpoint; max. setpoint | X | X | X | – | – | – |
| Language | X | X | X | – | – | – | |
| Temperature format | X | X | X | – | – | – | |
| Temperature control mode | X | X | X | – | – | – | |
| Favorites menu | X | – | – | – | X | – | |
| Pilot ONE Ethernet | IP address; subnet mask; remote access | X | X | X | – | – | – |
| Time unit | X | X | X | – | – | – | |
| 2. Setpoint | X | X | X | – | – | – |
В строке Protective function значения (A)-(F) напечатаны как X, X, X, –, –, –; в исходном тексте они разнесены по строкам.
| Function | E-grade Basic | E-grade Exclusive | E-grade Professional |
|---|---|---|---|
| Communication Watchdog | – | – | X |
| Process data recording directly to a USB stick: Setpoint, Actual Value Internal and Actual Value Process / Heating Capacity %, Cooling Capacity %, and Pump Pressure / Pump Speed, and VPC Pressure | –/–/– | X/X/– | X/X/X |
| Calendar start | – | – | X |
| Раздел | Название | Страница |
|---|---|---|
| 2.9.1 | Connecting an externally closed application | 31 |
| 2.10 | Connecting to the power supply | 32 |
| 2.10.1 | Connection using socket with protective earth (PE) | 32 |
| 2.10.2 | Connection via hard wiring | 32 |
| 3 | Function description | 34 |
| 3.1 | Function description of the temperature control unit | 34 |
| 3.1.1 | General functions | 34 |
| 3.1.2 | Other functions | 34 |
| 3.2 | Information on the thermal fluids | 35 |
| 3.3 | To be noted when planning the test | 35 |
| 3.4 | “Pilot ONE®” controller | 36 |
| 3.4.1 | Functional overview of “Pilot ONE®” | 36 |
| 3.5 | Clock/event function | 39 |
| 3.5.1 | Rechargeable accumulator | 39 |
| 3.5.2 | Programmable event function | 39 |
| 3.5.2.1 | Event function “Alarm clock event” | 39 |
| 3.5.2.2 | Event function “Program event” | 39 |
| 3.6 | Operation via the touch screen | 39 |
| 3.7 | Display instruments | 40 |
| 3.7.1 | The touchscreen [88] | 40 |
| 3.8 | Control instruments | 40 |
| 3.8.1 | The touchbuttons | 40 |
| 3.8.2 | The categories | 41 |
| 3.8.3 | The sub-categories | 41 |
| 3.8.4 | The dialogs | 41 |
| 3.9 | Function examples | 41 |
| 3.9.1 | Display of software version | 41 |
| 3.9.2 | Start & Stop | 41 |
| 3.9.3 | Copying the settings to a data carrier | 42 |
| 3.9.3.1 | Saving to a USB flash drive | 42 |
| 3.9.3.2 | Loading from a USB flash drive | 42 |
| 3.9.4 | Restore factory settings | 43 |
| 3.9.4.1 | Restore to factory settings without overtemperature protection | 44 |
| 3.9.4.2 | Restore to factory settings including overtemperature protection | 44 |
| 4 | Setup mode | 46 |
| 4.1 | Setup mode | 46 |
| 4.1.1 | Turning on the temperature control unit | 46 |
| 4.1.2 | Turning off the temperature control unit | 46 |
| 4.1.3 | Setting the overtemperature protection | 46 |
| 4.1.3.1 | General information on the overtemperature protection | 47 |
The unit must not be installed in an ATEX zone because operation there can cause explosion.
Determines whether the unit can be placed in a hazardous area at all.
“Do NOT install or start up the temperature control unit within an ATEX zone.”
A drip tray for condensation water and thermal fluid must be installed below the unit.
Affects site preparation and containment of leaks.
“The responsible body must install a condensation water / thermal fluid drip tray below the temperature control unit.”
The refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit value per m³ of ambient air.
Room volume and charge determine whether additional safety measures are required.
“Observe the amount of refrigerant contained (see data sheet/rating plate of the temperature control unit) and the room size when installing the temperature control unit.”
Electrical lines directly below the unit must be protected against liquid ingress because condensation can form.
Prevents electrical shock hazard from condensation at cooling water connections.
“Electrical lines directly below the temperature control unit must be protected against liquid ingress.”
Only drinking water or filtered river/sea water may be used as cooling water; sea water must not be used.
Prevents corrosion and damage to the cooling water circuit.
“Sea water must not be used for water cooling.”
Do not install valves or quick-release couplings in the feed/discharge lines to the application.
Avoids overpressure damage to the application, especially glass apparatus.
“Do not install valves/quick-release couplings in the feed/discharge lines from the temperature control unit to the application and from the application to the temperature control unit.”
Models with 'P' in the name are intended for applications with high pressure drop.
Model suffix directly determines suitability for high-pressure-drop circuits.
“Temperature control units with “P” in model name: This temperature control unit is particularly suitable for applications requiring high pressure drops.”
The speed-controlled pump allows matching speed or pressure to the application.
Affects pump selection and achievable flow/pressure for the application.
“With the integrated speed-controlled pump it is possible to either control the speed or the pressure and hence optimally match it to the specified application.”
Do not move the unit while it is operating because of burns, freezing and escaping thermal fluid.
Prevents injury and thermal fluid spill during operation.
“Do not move temperature control units that are in operation.”
Units with heater are class III/FL and have adjustable overtemperature protection plus low-level protection.
Safety class determines required protective devices and permissible thermal fluids.
“Temperature control units with heating correspond to class number III/FL.”
Anti-corrosion agents are mandatory if salts such as chlorides or bromide are added to the water circuit.
Protects the cooling water circuit from corrosion damage.
“The addition of anti-corrosion agents is mandatory if salts (chlorides, bromide) have been added to the water circuit.”