BINDER BD 400
BINDER BD 400 — Incubators with Temperature 5…100 °C.
The BD 400 is a 400 L natural-convection incubator with a +5 to 100 °C working range, distinguished within the BD series by its large interior volume and 90 kg total load capacity. It provides stable temperature control (0.1 K fluctuation, 0.5 K uniformity) for high-volume incubation protocols. The unit operates on 230 V single-phase supply and requires 160 mm rear and 100 mm side wall clearances.
Direct drop-in replacements
BINDER BD Series — Model Comparison & Selection
Full specifications
| Temperature | 5...100 °C |
| Heating-up time to 100 °C | 120 min |
| Interior volume | 400 L |
| Load per rack | 35 kg |
| Nominal power | 0.85 kW |
| Permitted load | 90 kg (BD400-230V / BD400UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.5 K |
| Back wall clearance | 160 mm (BD400-230V / BD400UL-120V) |
| Side wall clearance | 100 mm (BD400-230V / BD400UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 37°C mfr | 10 min |
| Depth | 765 mm (BD400-230V / BD400UL-120V) |
| Height | 1025 mm (BD400-230V / BD400UL-120V) |
| Interior width | 1000 mm (BD400-230V / BD400UL-120V) |
| Width | 1235 mm (BD400-230V / BD400UL-120V) |
| Net weight of the unit (empty) mfr | 135 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 12.5 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and Series Positioning
The BD 400 is the largest model in BINDER's BD series of natural-convection incubators, offering 400 L of interior volume compared to smaller siblings such as the BD 260, BD 115, BD 56, and BD 23. All models share the same +5 to 100 °C temperature range, but the BD 400 is uniquely suited for high-throughput applications requiring a large chamber. Its interior dimensions (1,000 mm W × 1,025 mm H × 765 mm D) and 90 kg permitted load (35 kg per rack) accommodate multiple racks or bulky items. The unit is available in two voltage variants: BD400-230V (230 V, 50/60 Hz, CEE 7/7 plug) and BD400UL-120V (120 V, 60 Hz, NEMA 5-15 plug).
Thermal Behavior and Load Considerations
Natural convection ensures gentle, uniform heating without forced air, making the BD 400 suitable for applications sensitive to airflow. Temperature fluctuation is held to 0.1 K and uniformity to 0.5 K across the chamber. Heating from ambient to 100 °C takes 120 minutes, and recovery after a 30-second door opening at 37 °C is 10 minutes. These characteristics support protocols requiring stable long-term incubation. The load capacity (90 kg total, 35 kg per rack) must be respected to avoid compromising airflow and temperature distribution.
Installation and Operation
The unit requires a single-phase 230 V supply (or 120 V for UL variant) with a 12.5 A fuse and 0.85 kW nominal power. Wall clearances of 160 mm at the rear and 100 mm at the sides are mandatory for heat dissipation. The net weight is 135 kg, necessitating a sturdy floor or bench. No additional control interfaces or media connections are specified in the available documentation; operation is via the unit's integrated temperature controller.
Integration & Operating Notes
Insufficient clearance restricts natural convection airflow around the chamber, leading to overheating of components and potential temperature control errors. Ensure the installation location allows these minimum distances from walls or adjacent equipment.
The unit's 0.85 kW nominal power and 12.5 A fuse rating require a dedicated circuit capable of handling inrush current. Verify that the supply can deliver the required current without voltage drop that could affect temperature stability.
Exceeding the permitted load may damage the rack supports or compromise airflow, leading to temperature non-uniformity. Distribute loads evenly and verify rack capacity before loading.
The door seal affects temperature recovery and uniformity. Regular inspection for wear and cleaning with mild detergent helps maintain performance. No specific interval is provided; establish a schedule based on usage.
Selection & Integration Q&A
Choose the BD 400 when you need 400 L capacity and can handle 90 kg total load; it suits high-volume incubation. Smaller models like BD 260 (260 L) or BD 115 (115 L) are better for lower throughput or space-constrained labs, but all share the same temperature range.
The standard BD400-230V requires 230 V, 50/60 Hz single-phase supply with a 12.5 A fuse and 0.85 kW nominal power. The UL variant BD400UL-120V requires 120 V, 60 Hz. Ensure the circuit can handle the load and provide a reliable earth connection.
Maintain at least 160 mm clearance at the rear and 100 mm at the sides to allow adequate airflow for natural convection cooling. Failure to do so may cause overheating and affect temperature uniformity.
Regularly inspect and clean the door seal to maintain temperature recovery. No specific accessories are included beyond the standard unit; racks are optional and must not exceed 35 kg each. Refer to the manual for detailed maintenance intervals.