BINDER BD 23
BINDER ·Made in Germany · In Active Production

BINDER BD 23

BINDER BD 23 — Incubators with Temperature 5…100 °C.

The BD 23 is a 20 L natural-convection incubator in BINDER's BD series, configured as the 230 V/50 Hz standard variant (BD023-230V, Art. 9010-0187) with a CEE 7/7 plug. It covers +5 to 100 °C with 0.2 K fluctuation and 0.5 K uniformity, reaching 100 °C in 59 min. It serves as a compact benchtop unit for temperature-controlled incubation where forced-air circulation is not required.

Temperature
5…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BD Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kg
BD 23 Current 5–100525949.52012
BD 115 5–10060.59070.511230
BD 260 5–10076759425340
BD 400 5–10076.5120102.540035
BD 56 5–10056.57562.55730
BD 720 5–100879015973745

Full specifications

Core Performance & Functional Specifications
Temperature 5...100 °C
Heating-up time to 100 °C 59 min
Interior volume 20 L
Load per rack 12 kg
Nominal power 0.2 kW
Permitted load 25 kg (BD023-230V)
Temperature fluctuation 0.2 K
Temperature uniformity 0.5 K
Back wall clearance 100 mm (BD023-230V)
Side wall clearance 50 mm (BD023-230V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 20 min
Dimensions, Weight & Mechanical Design
Depth 520 mm (BD023-230V)
Height 495 mm (BD023-230V)
Interior width 222 mm (BD023-230V)
Width 435 mm (BD023-230V)
Net weight of the unit (empty) mfr 27 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 12.5 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load

Installation and operation

Integration & Operating Notes

Electrical
Match the 230 V variant to a 50/60 Hz supply with CEE 7/7 outlet

The BD023-230V is a single-phase 230 V unit drawing 0.2 kW nominal and protected by a 12.5 A fuse; the supply circuit and outlet must be rated accordingly. The 120 V/60 Hz UL variant (BD023UL-120V, NEMA 5-15) is a separate article and cannot be substituted on a 230 V installation.

Data Sheet Model BD 23 en.pdf, variant table
Mounting
Respect 100 mm rear and 50 mm side clearance for heat dissipation

The chamber relies on natural convection and passive heat exchange with the room; insufficient rear or side clearance compromises the 0.5 K uniformity figure and can overheat the housing. Plan bench depth for the 520 mm interior depth plus the 100 mm rear gap.

Factory Specification Sheet, wall clearance values
Process Limits
Loading pattern governs achievable uniformity and recovery time

With 0.2 K fluctuation and 0.5 K uniformity specified, the 25 kg total / 12 kg per-rack load limits and free air paths must be maintained; dense or floor-covering loads block gravity circulation and lengthen the 20 min recovery after a 30 s door opening at 37 °C.

Factory Specification Sheet, load and recovery values

Selection & Integration Q&A

When should I choose the BD 23 over the larger BD 56, BD 115, BD 260 or BD 400?

Pick BD 23 for small batches on a crowded bench: 20 L interior, 435 mm exterior width, 27 kg. Move to a sibling when sample count or rack area exceeds the 25 kg total / 12 kg per-rack load; all siblings share the +5 to 100 °C range, so the decision is purely chamber size and load capacity.

What supply and outlet does the 230 V BD 23 require, and can I use the UL variant instead?

The BD023-230V needs single-phase 230 V, 50/60 Hz, 0.2 kW nominal, on a circuit with 12.5 A fuse protection and a CEE 7/7 outlet. The BD023UL-120V is a distinct article (120 V, 60 Hz, NEMA 5-15) and is not interchangeable on a 230 V supply.

How much space must I leave around the unit, and how does loading affect temperature performance?

Allow 100 mm behind and 50 mm at each side for natural-convection heat exchange. Keep loads within 25 kg total and 12 kg per rack and avoid blocking air paths; otherwise the specified 0.5 K uniformity and 20 min recovery at 37 °C after a 30 s door opening will not be met.

What is not included with the BD 23 that I need to plan for at commissioning?

The standard scope covers the unit only; racks, stacking accessories and calibration certificates are not included and must be ordered separately. Size the rack order against the 12 kg per-rack limit and the 25 kg total load before installation.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model BD 23 en
Datasheet ·235,2 КБ ·7 pp.
View PDF