BINDER KB PRO 1060
BINDER KB PRO 1060 — Incubators with Temperature -20…100 °C.
The KB PRO 1060 is a 1,070 L single-phase cooling incubator with a -20…100 °C working range and 1.8 kW nominal power. It is the second-largest chamber in the KB PRO series, distinguished by a 200 kg permitted load and 40 kg per rack, intended for high-volume temperature-stability processes at 0.1 K fluctuation and 0.2 K uniformity.
Direct drop-in replacements
BINDER KB PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Interior width, mm | Load per rack, kg |
|---|---|---|---|---|---|---|
| KB PRO 1060 Current | -20–100 | 117.5 | 195.1 | 1070 | 1011 | 40 |
| KB PRO 130 | -20–100 | 71 | 100.5 | 127 | 656 | 30 |
| KB PRO 1600 | -20–100 | 117.5 | 195.1 | 1610 | 1521 | 40 |
| KB PRO 260 | -20–100 | 90 | 139.6 | 262 | 656 | 40 |
| KB PRO 470 | -20–100 | 90 | 195.1 | 472 | 656 | 40 |
| KB PRO 720 | -20–100 | 90 | 195.1 | 727 | 1011 | 40 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 1070 L |
| Load per rack | 40 kg |
| Max. heat compensation | 1000 W |
| Nominal power | 1.8 kW |
| Permitted load | 200 kg |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm |
| Side wall clearance | 180 mm |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 4 min |
| Depth | 1175 mm |
| Height | 1951 mm |
| Interior width | 1011 mm |
| Width | 1285 mm |
| Net weight of the unit (empty) mfr | 353 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The KB PRO 1060 is a floor-standing cooling incubator with a 1,011 mm interior width, 1,175 mm depth, and 1,951 mm height, giving 1,070 L of usable volume. It uses the series' signal-triangle instrument panel and ergonomically adjustable control terminal (shared with the 470, 720, and 1600; the 130 and 260 use a triangular panel).
- Within the KB PRO series, the 1060 sits between the 720 and the 1600. All siblings share the -20…100 °C range; the 1060 is selected for its 200 kg total permitted load and 40 kg per-rack capacity, which exceed smaller models and support dense rack loading.
- The chamber is rated for 1.8 kW nominal power at 208…240 V, 1~, 50 Hz, with a 16.0 A unit fuse. The refrigeration circuit uses R290 (hydrocarbon) refrigerant; service access is via a machine-room cover removable without tools.
Behavior under load and in different modes
- Temperature performance is specified at 0.1 K fluctuation and 0.2 K uniformity. Recovery after a 30 s door opening at 37 °C is 4 minutes, which matters for frequent-access protocols.
- The MB2 controller supports fixed-value operation, timer/stopwatch programs, time programs with ramps and steps, and week programs. Hot-air disinfection is a dedicated procedure for chamber sanitation.
- Defrosting is required during refrigerating operation; anti-condensation protection is provided. The temperature safety device is class 2/3.3 per DIN 12880:2007, configurable as temperature limiter (class 2) or temperature protection (class 3.3), with optional class 3.2 undertemperature protection.
Installation, connection, and operation
- From size 260 upward, a wall spacer is required. For the 1060, maintain 100 mm rear clearance and 180 mm side clearance; do not install in unventilated recesses.
- Racks are available with U-rails or telescopic rails as accessories. Electrical connection is via CEE 7/7 plug; verify 208…240 V single-phase supply and 16 A branch protection.
- The controller provides Ethernet, E-Mail, and USB interfaces for data transfer and remote notification. Options include APT-COM 4 software, analog temperature outputs, zero-voltage alarm contacts, internal LED light, door lock, and a flexible Pt 100 object temperature sensor.
Integration & Operating Notes
The chamber requires a wall spacer from size 260 upward. Insufficient rear or side clearance prevents heat dispersal and can cause overheating; do not install in unventilated recesses. Verify clearance before positioning the 1,285 mm wide unit.
The unit draws 1.8 kW nominal at 208…240 V, 1~, 50 Hz and is protected by a 16.0 A fuse. Confirm supply voltage and frequency before connection; use a CEE 7/7 plug compatible outlet. Undersized branch protection will trip during heat-up.
The permitted load is 200 kg for the chamber and 40 kg per rack. Exceeding these limits risks rack deformation and invalidates temperature uniformity. Distribute loads evenly and use U-rails or telescopic rails rated for the application.
The temperature safety device can be set as class 2 (temperature limiter) or class 3.3 (temperature protection), with optional class 3.2 undertemperature protection. Select the class and limit/offset according to process risk; an incorrect class may fail to protect the sample or trip unnecessarily.
Regular maintenance is required at least once a year, and electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended before initial startup and after repair. A service message appears after 8,760 operating hours or two years and recurs every two weeks until reset by BINDER Service.
The condenser sits behind the machine-room cover, which is removable without tools. In dusty ambient air, suction-clean the condenser several times a year to maintain refrigeration performance. Refrigeration work must be performed by personnel trained per EN 13313:2010.
Selection & Integration Q&A
Choose the 1060 when you need 1,070 L and a 200 kg total load with 40 kg per rack, between the 720 and 1600. The 1600 offers more volume but the same temperature range; the 720 is smaller. All share -20…100 °C and 0.1 K fluctuation.
It requires 208…240 V, 1~, 50 Hz with a 16.0 A unit fuse and 1.8 kW nominal power. Provide a dedicated branch rated for at least 16 A and verify voltage/frequency before connection; the CEE 7/7 plug must match the outlet.
Maintain 100 mm rear and 180 mm side clearance; a wall spacer is required from size 260 upward. Do not install in unventilated recesses. Ambient dust increases condenser cleaning frequency. The unit weighs 353 kg empty, so floor loading must be verified.
Annual maintenance and electrical safety testing are required; a service message appears after 8,760 h or two years. Racks (U-rails or telescopic), APT-COM 4 software, Pt 100 sensor, alarm contacts, and door lock are options/accessories not included in base delivery.