BINDER KB PRO 1600
BINDER KB PRO 1600 — Incubators with Temperature -20…100 °C.
The KB PRO 1600 is a floor-standing cooling incubator with a 1,610 L interior, the largest chamber in the KB PRO series, operating from -20 to 100 °C with 0.1 K fluctuation and 0.2 K uniformity. Its 1,000 W heat compensation and 40 kg per-rack load capacity support high-density incubation protocols. It serves as a central, high-throughput temperature-controlled environment in pharmaceutical and biotechnological process chains.
Direct drop-in replacements
BINDER KB PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Interior width, mm | Load per rack, kg |
|---|---|---|---|---|---|---|
| KB PRO 1600 Current | -20–100 | 117.5 | 195.1 | 1610 | 1521 | 40 |
| KB PRO 1060 | -20–100 | 117.5 | 195.1 | 1070 | 1011 | 40 |
| KB PRO 130 | -20–100 | 71 | 100.5 | 127 | 656 | 30 |
| KB PRO 260 | -20–100 | 90 | 139.6 | 262 | 656 | 40 |
| KB PRO 470 | -20–100 | 90 | 195.1 | 472 | 656 | 40 |
| KB PRO 720 | -20–100 | 90 | 195.1 | 727 | 1011 | 40 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 1610 L |
| Load per rack | 40 kg |
| Max. heat compensation | 1000 W |
| Nominal power | 1.8 kW |
| Permitted load | 200 kg |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm |
| Side wall clearance | 180 mm |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 6 min |
| Depth | 1175 mm |
| Height | 1951 mm |
| Interior width | 1521 mm |
| Width | 1795 mm |
| Net weight of the unit (empty) mfr | 443 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The KB PRO 1600 (Art. No. 9020-0438) is the largest model in the KB PRO series, with a 1,610 L interior volume and net dimensions of 1,795 mm wide × 1,951 mm high × 1,175 mm deep; empty weight is 443 kg.
- Unlike the smaller KB PRO 130/260, which use a triangular instrument panel, the 1600 uses a signal triangle and an ergonomically adjustable control terminal, consistent with the 470/720/1060 models.
- All series models share the -20…100 °C working range; the 1600 is distinguished by its capacity and by a 1,000 W maximum heat compensation rating, enabling stable control under heavy internal loads.
- The chamber is equipped with a temperature safety device per DIN 12880:2007 class 2/3.3, with selectable class 2 (temperature limiter) or class 3.3 (temperature protection) modes; class 3.2 undertemperature protection is also available.
Behavior under load and in different media
- Temperature fluctuation is 0.1 K and uniformity 0.2 K, supporting reproducible incubation conditions across the large interior.
- Recovery time after a 30 s door opening at 37 °C is 6 minutes, reflecting the thermal inertia of the 1,610 L volume and the 1,000 W heat compensation capability.
- The permitted load is 200 kg total, with a maximum of 40 kg per rack; racks are available with U-rails or telescopic rails as accessories.
- The refrigeration system uses R290 refrigerant (hydrocarbon); any work on the refrigeration circuit must be performed by personnel trained to EN 13313:2010.
Installation, connection and control
- Electrical supply is single-phase 208…240 V, 50 Hz, with a nominal power of 1.8 kW and a 16.0 A unit fuse; verify branch protection accordingly.
- Wall clearance must be at least 100 mm at the rear and 180 mm at the sides; a wall spacer is required from size 260 upward, and the chamber must not be installed in unventilated recesses.
- The MB2 chamber controller provides fixed-value, timer, time-program and week-program operating modes, with Ethernet, E-Mail and USB interfaces for data transfer and remote notification.
- Options include APT-COM 4 software, analog temperature outputs, zero-voltage alarm contacts, internal LED light, door lock, and a flexible Pt 100 object temperature sensor with adjustable sensitivity and maximum deviation.
Integration & Operating Notes
The unit draws 1.8 kW nominal on a single-phase 208…240 V, 50 Hz supply and is protected by a 16.0 A unit fuse. The upstream branch protection and wiring must be sized for this load and for the inrush behavior of the refrigeration compressor; verify voltage and frequency before connection.
The chamber must not be installed in unventilated recesses. A wall spacer is required from size 260 upward, and the prescribed minimum distances must be observed to prevent overheating and to allow condenser access. Plan the floor position so that the machine room cover can be removed without tools.
Rack loading must not exceed 40 kg per rack, and the total permitted load on the chamber is 200 kg. Exceeding these limits risks deformation of the racks and impaired temperature uniformity. Racks with U-rails or telescopic rails are available as accessories and must be selected according to loading and access requirements.
The door gasket must be replaced only when the chamber is at ambient/cold condition; replacing it while warm may damage the gasket. Electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended prior to initial startup and after maintenance or repair.
Selection & Integration Q&A
Choose the 1600 when the 1,610 L interior and 200 kg total load capacity are needed for high-throughput protocols; the 1060 and 720 offer the same -20…100 °C range but smaller volumes and lower load limits. The 1600 also provides 1,000 W heat compensation for heavy internal loads.
Single-phase 208…240 V, 50 Hz, nominal power 1.8 kW, unit fuse 16.0 A. The upstream branch circuit and wiring must be sized for this load and for compressor inrush; verify voltage and frequency before connection.
Maintain at least 100 mm rear and 180 mm side clearance; a wall spacer is required from size 260 upward. Do not install in unventilated recesses. In dusty ambient air, clean the condenser several times a year by suction.
Annual maintenance is required, with a BINDER service reset at 8760 operating hours or two years. Racks (U-rails or telescopic rails), APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, and Pt 100 sensor are options/accessories not included in the base unit.