BINDER MK 240
BINDER MK 240 — Climate chambers with Temperature -40…180 °C.
The BINDER MK 240 is a 228 L alternating climate chamber (MK240-400V, Art. 9020-0376) for rapid temperature cycling from -40 to 180 °C, with average heating/cooling rates of 5.0/4.5 K/min per IEC 60068-3-5. It is the mid-size 400 V three-phase standard variant, distinguished from the CO₂-capable MK 240 CO₂ and the larger MK 1020 by its 228 L interior and 5.6 kW nominal power. It serves as a dynamic test station for thermal stress screening and material qualification in a laboratory infrastructure.
BINDER MK Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior height, mm | Interior volume, L | Interior width, mm |
|---|---|---|---|---|---|---|
| MK 240 Current | -40–180 | 92.5 | 171 | 700 | 228 | 735 |
| MK 1020 | -40–180 | 148 | 200.5 | 1020 | 991 | 1200 |
| MK 1020 CO₂ | -48–180 | 148 | 200.5 | 1020 | 991 | 1200 |
| MK 115 | -40–180 | 86.5 | 172.5 | 480 | 115 | 600 |
| MK 240 CO₂ | -50...180 | 92.5 | 171 | 700 | 228 | 735 |
| MK 400 | -40–180 | 140 | 171 | 700 | 417 | 735 |
| MK 400 CO₂ | -50...180 | 140 | 171 | 700 | 417 | 735 |
| MK 56 | -40–180 | 77.8 | 144.5 | 420 | 60 | 400 |
| MK 720 | -40–180 | 114 | 200.5 | 1020 | 734 | 1200 |
| MK 720 CO₂ | -48–180 | 114 | 200.5 | 1020 | 734 | 1200 |
Full specifications
| Temperature | -40...180 °C |
| Interior volume | 228 L |
| Load per rack | 30 kg (MK240-400V) |
| Nominal power | 5.6 kW |
| Article Number mfr | 9020 |
| Depth | 925 mm (MK240-400V) |
| Height | 1710 mm (MK240-400V) |
| Interior height | 700 mm (MK240-400V) |
| Interior width | 735 mm (MK240-400V) |
| Width | 1115 mm (MK240-400V) |
| Net weight of the unit (empty) mfr | 321 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
| Cooling down time from 180°C to -40°C mfr | 110 min |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.0 K/min |
Description
Construction and series positioning
- The MK 240 is a floor-standing alternating climate chamber with 228 L usable interior volume (735 mm W × 700 mm H × 925 mm D net), 321 kg empty weight, and a 400 V 3 N/PE 50 Hz supply at 5.6 kW nominal power, 16.0 A unit fuse.
- Within the MK (E5) series it sits between the MK 115 and MK 400; unlike the MK 240 CO₂ variant (which extends down to -50 °C), this standard model covers -40 to 180 °C. The -C suffix version (MK240-400V-C) adds a voltage and frequency changer for 480 V 60 Hz sites; the standard MK240-400V described here has no such converter.
- The chamber uses HIT Insulation® and program control via the MB2 controller; a rear power switch is fitted from size 115 upward.
Thermal behavior and control
- Factory data give temperature uniformity of 0.1...1.2 K and fluctuation of 0.1...0.5 K depending on set value, with a 110 min full cooling-down time from 180 °C to -40 °C.
- The MB2 controller supports fixed-value mode, time programs with ramps/steps and tolerance ranges, and week programs; behavior during power failure and door opening is defined in the controller documentation.
- Temperature safety is provided by an over-temperature protective device class 1 (trips at nominal +20 °C on chamber failure) and a DIN 12880 class 2.0 safety device; an over/under temperature safety device class 2 is optional.
Installation and operation
- The chamber must be installed with sufficient ventilation for heat dispersal and the prescribed minimum distances; it must not be placed in unventilated recesses, damp areas, or hazardous locations, and no explosive or flammable atmospheres may be present.
- Electrical connection is 400 V three-phase; the unit must be commissioned and tested by qualified personnel only. An Ethernet serial interface is standard for networking with APT-COM™ 4; RS485, analog temperature outputs, and a flexible Pt 100 object sensor are options.
- A maintenance message appears after 8760 operating hours or two years and recurs every two weeks until reset by BINDER Service.
Integration & Operating Notes
The MK240-400V draws 5.6 kW at 400 V 3 N/PE 50 Hz and is protected by a 16.0 A unit fuse. The site supply must be sized for this three-phase load and the chamber must not be connected to a single-phase or lower-voltage circuit; the -C variant is the only one that accepts 480 V 60 Hz via its converter.
Sufficient ventilation for heat dispersal and the prescribed minimum distances must be observed; installation in unventilated recesses risks overheating. The chamber must not become wet and must not be installed in damp areas or puddles, because water ingress creates a deadly electric shock hazard.
The chamber has no explosion protection measures and must not be operated in potentially explosive areas or with combustible dust or air-solvent mixtures near it. Charging materials and site classification must exclude flammable substances at working temperature up to 180 °C.
The over-temperature protective device class 1 responds if the nominal temperature is exceeded by 20 °C due to chamber failure. A class 1 trip must be treated as a failure condition: take the chamber out of operation and inform BINDER Service rather than simply resetting and resuming.
After 8760 operating hours or two years a maintenance message appears; after confirmation it reappears every two weeks until reset by BINDER Service. Schedule the service intervention at or before this interval to avoid recurring messages and to keep the maintenance counter valid.
An annual repeat electrical inspection is required, plus a test prior to initial startup and prior to resuming operation after maintenance or repair; testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. The chamber must not be returned to service without this test.
Selection & Integration Q&A
Pick the MK 240 standard for -40...180 °C cycling in a 228 L interior at 400 V 50 Hz. Choose the MK 240 CO₂ only if you need the extended -50 °C low end; choose the MK 1020 when 228 L is insufficient and a larger interior is required.
A 400 V 3 N/PE 50 Hz supply with 16.0 A fusing for the 5.6 kW load, plus an Ethernet connection for APT-COM™ 4 networking. RS485 and analog temperature outputs are options that must be ordered with the chamber.
Install with sufficient ventilation and minimum distances, never in unventilated recesses, damp areas, puddles, or hazardous locations. Explosive atmospheres and combustible dust or air-solvent mixtures near the chamber are prohibited because it has no explosion protection.
The base unit includes the chamber and MB2 controller; APT-COM™ 4 software, RS485, analog outputs, Pt 100 object sensor, compressed air connection/dryer, and class 2 safety device are options. Service is due at 8760 h or two years, with annual electrical safety testing.