BINDER MK 240 CO₂
BINDER ·Made in Germany · In Active Production

BINDER MK 240 CO₂

BINDER MK 240 CO₂ — Climate chambers.

The MK 240 CO₂ is a 228 L alternating climate chamber with a CO₂ (R744) refrigeration circuit, extending the MK series range down to -50 °C. It is the CO₂-refrigerant modification of the MK 240, distinguished from the standard MK 240 (-40 °C) by its lower temperature limit and high-pressure refrigerant. It serves rapid temperature-change testing of components and materials under program control.

BINDER MK Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior height, mmInterior volume, LInterior width, mm
MK 240 CO₂ Current -50...18092.5171700228735
MK 1020 -40–180148200.510209911200
MK 1020 CO₂ -48–180148200.510209911200
MK 115 -40–18086.5172.5480115600
MK 240 -40–18092.5171700228735
MK 400 -40–180140171700417735
MK 400 CO₂ -50...180140171700417735
MK 56 -40–18077.8144.542060400
MK 720 -40–180114200.510207341200
MK 720 CO₂ -48–180114200.510207341200

Full specifications

Core Performance & Functional Specifications
Temperature -50...180
Interior volume 228 L
Load per rack 30 kg (MK240CO2-400V, 50 Hz)
Nominal power 5.2 kW
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 925 mm (MK240CO2-400V, 50 Hz)
Height 1710 mm (MK240CO2-400V, 50 Hz)
Interior height 700 mm (MK240CO2-400V, 50 Hz)
Interior width 735 mm (MK240CO2-400V, 50 Hz)
Width 1115 mm (MK240CO2-400V, 50 Hz)
Net weight of the unit (empty) mfr 311 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 3 N/PE
Operating Conditions, Environment & Protection
Cooling down time from 180°C to -40°C mfr 60 min
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.5 K/min

Description

Construction and series positioning

Thermal behavior

Control and operation

Electrical

Integration & Operating Notes

Electrical
Supply must be backed by a type B RCD

The chamber may only be operated on a circuit protected by a type B residual current device; a type A or AC RCD in the existing distribution board is not acceptable and must be replaced or supplemented before commissioning.

Instruction Manual, Safety Labels, chap. 1.5 p.9
Electrical
Size the feed for 8.9 A at 400 V / 50 Hz

The 50 Hz variant draws a nominal 8.9 A with 5.2 kW connected load and a 16.0 A unit fuse; the site breaker and cable cross-section must be coordinated with this rather than with the 16 A fuse alone.

Instruction Manual, Type Plate p.11
Mounting
Room volume must be verified per EN 378-1

Because the refrigerant is CO₂ at up to 120 bar, the installation room volume must be calculated according to EN 378-1:2016+A1:2020 before commissioning; unventilated recesses and damp areas are prohibited and minimum clearances must be kept for heat dispersal.

Instruction Manual, Safety chap. 1.8 p.12-13
Process Limits
Charging material must stay below its flash point

The chamber temperature must remain below the flash point or sublimation point of the load and no explosive mixture with air may form; verify the material data before setting any program that approaches the 180 °C upper limit.

Instruction Manual, Safety chap. 1.8 p.13
Maintenance
Annual service plus annual electrical repeat test

Maintenance is required at least once a year and an electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup, annually, and after any maintenance or repair; refrigeration work must be logged in a service log book.

Instruction Manual, Maintenance chap. 21.1-21.2 p.107-108
Maintenance
Service message at 8760 h is a reminder, not a fault

After 8760 operating hours or two years the controller shows a service message that reappears every two weeks once acknowledged; it can only be reset by BINDER Service, so plan the service visit rather than attempting a controller reset.

Instruction Manual, Maintenance chap. 21.2 p.108
Maintenance
Refrigeration circuit repairs only at BINDER

Work on the CO₂ circuit may currently only be carried out at BINDER, and any refrigeration maintenance must be performed by personnel trained to EN 13313:2010; on-site repair of the refrigerant side is therefore not an option and a return must be scheduled.

Instruction Manual, Maintenance chap. 21.1-21.2 p.107-108
Maintenance
Check condenser fan grid weekly in dusty rooms

With elevated ambient dust the condenser fan must be cleaned several times a year; inspect the fan grid behind the left maintenance access flap weekly and suction it clean when soiled, and never remove the rear panel or let the chamber become wet.

Instruction Manual, Maintenance chap. 21.2 p.107-108

Selection & Integration Q&A

When should I choose the MK 240 CO₂ rather than the standard MK 240 or the larger MK 1020 CO₂?

Pick the MK 240 CO₂ when you need the -50 °C lower limit in a 228 L interior; the standard MK 240 stops at -40 °C. Move to the MK 1020 CO₂ only if the 228 L volume and 30 kg/rack load are insufficient for your specimen batch.

What electrical supply and protective device does the 400 V variant require?

400 V 3 N/PE 50 Hz, 5.2 kW, 8.9 A nominal, 16.0 A unit fuse, and a type B residual current circuit breaker in the site installation; a 480 V/60 Hz variant (Art. 9020-0515) exists for other grids.

Where can the chamber be installed and what room conditions apply?

Not in unventilated recesses or damp areas; the room volume must be calculated per EN 378-1:2016+A1:2020 for the CO₂ charge, minimum distances must be kept for heat dispersal, and the unit is IP 20 so it must be set up splash-proof.

What maintenance and accessories are needed beyond the base unit?

Annual maintenance, annual electrical repeat inspection per EN 50678/50699, and a service log book are mandatory; RS485, zero-voltage switching outputs, analog temperature outputs, Pt 100 object sensor and APT-COM 4 are options not included in base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model MK CO₂ 240 en
Datasheet ·723,0 КБ ·9 pp.
View PDF
PDF
0714 EN MK CO2 E6 UM 10 2026
Manual ·10,0 МБ ·139 pp.
View PDF