BINDER MK 56
BINDER MK 56 — Climate chambers with Temperature -40…180 °C.
The BINDER MK 56 is a 60 L dynamic climate chamber with a -40 to 180 °C working range and 5 K/min average heating/cooling rates per IEC 60068-3-5. As the smallest MK series model, it is the only size besides the MK 1020 available with factory water cooling. It is used for rapid temperature-change testing of small specimens in laboratory and quality-assurance environments.
BINDER MK Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior height, mm | Interior volume, L | Interior width, mm |
|---|---|---|---|---|---|---|
| MK 56 Current | -40–180 | 77.8 | 144.5 | 420 | 60 | 400 |
| MK 1020 | -40–180 | 148 | 200.5 | 1020 | 991 | 1200 |
| MK 1020 CO₂ | -48–180 | 148 | 200.5 | 1020 | 991 | 1200 |
| MK 115 | -40–180 | 86.5 | 172.5 | 480 | 115 | 600 |
| MK 240 | -40–180 | 92.5 | 171 | 700 | 228 | 735 |
| MK 240 CO₂ | -50...180 | 92.5 | 171 | 700 | 228 | 735 |
| MK 400 | -40–180 | 140 | 171 | 700 | 417 | 735 |
| MK 400 CO₂ | -50...180 | 140 | 171 | 700 | 417 | 735 |
| MK 720 | -40–180 | 114 | 200.5 | 1020 | 734 | 1200 |
| MK 720 CO₂ | -48–180 | 114 | 200.5 | 1020 | 734 | 1200 |
Full specifications
| Temperature | -40...180 °C |
| Interior volume | 60 L |
| Load per rack | 15 kg (MK056-230V, 50 Hz) |
| Nominal power | 2.8 kW |
| Article Number mfr | 9020 |
| Depth | 778 mm (MK056-230V, 50 Hz) |
| Height | 1445 mm (MK056-230V, 50 Hz) |
| Interior height | 420 mm (MK056-230V, 50 Hz) |
| Interior width | 400 mm (MK056-230V, 50 Hz) |
| Width | 720 mm (MK056-230V, 50 Hz) |
| Net weight of the unit (empty) mfr | 165 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Cooling down time from 180°C to -40°C mfr | 90 min |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.0 K/min |
Description
Construction and series positioning
- The MK 56 is a single-phase 230 V / 50 Hz chamber (2.8 kW nominal, 16 A unit fuse) with a 60 L interior (400 × 420 mm W×H) and 15 kg load per rack.
- Within the MK series, it is the only model below 115 L and shares the -40…180 °C range with the MK 115, MK 240 and MK 1020; the MK 1020 CO₂ extends down to -48 °C.
- Unlike the larger MK 115 and above, the MK 56 has a main power switch on the front; from size 115 upward a rear power switch is added.
- The water cooling option is available only on the MK 56 and MK 1020, making this model suitable where air cooling is insufficient.
Thermal behavior and control
- Average heating and cooling rates are both 5.0 K/min between -40 °C and 180 °C; a full 180 → -40 °C cool-down takes 90 min.
- Temperature uniformity is 0.5…1.5 K and fluctuation 0.1…0.5 K, both dependent on set value.
- The MB2 controller supports fixed-value, time-program and week-program modes, with set-point ramps/steps, tolerance ranges and section repetition.
- Over temperature protective device class 1 trips if nominal temperature is exceeded by 20 °C due to chamber failure; an optional over/under temperature safety device class 2 is available.
Installation and interfaces
- The chamber must not be installed in unventilated recesses; sufficient ventilation for heat dispersal and prescribed minimum distances (chapter 3.4) are required.
- An Ethernet serial interface is standard for networking with APT-COM™ 4 software; RS485 and analog temperature outputs are options.
- Water cooling, if ordered, uses separate cooling water inlet and outlet connections.
- The chamber must not be operated in potentially explosive areas or with combustible dust/air-solvent mixtures; no explosion protection is provided.
Integration & Operating Notes
The chamber requires sufficient ventilation for heat dispersal and must observe prescribed minimum distances (chapter 3.4). Installation in unventilated recesses risks overheating damage; damp areas or puddles create a deadly electric shock hazard because water ingress into electrical parts is possible.
The MK056-230V variant draws 2.8 kW at 230 V / 50 Hz and requires a 16.0 A unit fuse. The supply circuit must be sized accordingly; the alternative MK056-240V variant uses 240 V / 60 Hz with a NEMA 6-20 plug. Verify the correct variant and plug type before installation.
After 8760 operating hours or two years, a maintenance message appears; after confirmation with the Confirm icon, the message window reappears every two weeks until reset by BINDER Service. Plan service intervention at or before this interval to reset the counter and avoid recurring messages.
An annual repeat electrical inspection is required, as well as a test prior to initial startup and prior to resuming operation after maintenance or repair. Testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Schedule this inspection to maintain compliance and safety.
The chamber is equipped with an Ethernet serial interface and can be networked with the BINDER APT-COM™ 4 Multi Management Software (accessory). Network configuration (chapter 14.5) and APT-COM™ 4 registration (chapter 20.1.1.1) are required for networked monitoring and documentation.
Selection & Integration Q&A
Choose the MK 56 for small specimens up to 60 L interior volume and 15 kg per rack, especially when single-phase 230 V power is available. Larger models (MK 115, MK 240) require 400 V three-phase supply and offer more space but no water cooling option (only MK 56 and MK 1020 offer water cooling).
Do not install in unventilated recesses; sufficient ventilation for heat dispersal and prescribed minimum distances (chapter 3.4) must be observed. Do not operate in potentially explosive areas or with combustible dust/air-solvent mixtures. Avoid damp areas or puddles; water ingress creates a deadly electric shock hazard.
A maintenance message appears after 8760 operating hours or two years; annual electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required. Base delivery excludes options like RS485 interface, analog temperature outputs, flexible Pt 100 sensor, compressed air connection/dryer, water cooling, APT-COM™ 4 software, and over/under temperature safety device class 2.