MK+MKT (E5) 10 2026
Technical Reference Summary
Covered Models & Document Scope
This operating manual (MK/MKT (E5), issue 10/2026, art. no. 7001-0364) covers the BINDER MK (E5) alternating climate chambers with program control and HIT Insulation® and the MKT (E5) alternating climate chambers with deep temperature and program control. Covered model versions: MK 56 (MK056-230V, MK056-240V), MK 115 (MK115-400V, MK115-400V-C), MK 240 (MK240-400V, MK240-400V-C), MK 400 (MK400-400V, MK400-400V-C), MK 720 (MK720-400V, MK720-400V-C), MK 1020 (MK1020-400V, MK1020-400V-C), MKT 115 (MKT115-400V, MKT115-400V-C), MKT 240 (MKT240-400V, MKT240-400V-C) and MKT 720 (MKT720-400V, MKT720-400V-C). The suffix -C denotes chambers equipped with a voltage and frequency changer. Options include RS485 interface, analog temperature outputs, flexible Pt 100 object temperature sensor, compressed air connection, compressed air dryer, water cooling (MK 56 and MK 1020 only), APT-COM™ 4 software, and over/under temperature safety device class 2.
Section-by-Section Technical Breakdown
Describes the chamber overview, the MB2 instrument panel, the lateral control panel, and the main and rear power switches. It identifies the main power switch on MK 56 and the rear power switch on MK/MKT from size 115 upward. The chamber is equipped with an Ethernet serial interface and can be networked with APT-COM™ 4 software.
- MK 56 uses a main power switch (1); MK/MKT from size 115 on additionally have a rear power switch (12).
- Instrument panel and lateral control panel are described in chapters 2.2 and 2.3.
- Ethernet serial interface is standard for networking and APT-COM™ 4 Multi Management Software.
- Rear chamber view is documented in chapter 2.6.
Covers cooling water connections for the water cooling option (MK 56 and MK 1020), installation of the voltage and frequency changer, and electrical connection of the chamber. It includes information on connecting the alternating climate chamber and on connecting the voltage and frequency changer where fitted.
- Water cooling option is available only for MK 56 and MK 1020; separate cooling water inlet and outlet connections are described.
- Voltage and frequency changer installation is described for chambers with the -C suffix.
- Electrical connection information is given in chapter 4.4, including connection of the voltage and frequency changer.
- Prescribed minimum distances for installation are referenced in chapter 3.4.
Describes the MB2 controller normal display, program display and chart-recorder display, the controller icon overview, operating modes, and the menu structure including main menu, Settings and Service submenus. It also covers controller entry principles and behavior during power failures and when the door is opened.
- Display views: normal display, program display, chart-recorder display.
- Operating modes and controller menu structure are described in chapters 5.4 and 5.5.
- Main menu, Settings submenu and Service submenu are documented in chapters 5.5.1 to 5.5.3.
- Performance during and after power failures is described in chapter 5.7; door-opening behavior in chapter 5.8.
Covers turning on the chamber, controller settings upon start up, and set-point entry in the Fixed Value operating mode. Setpoints can be entered through the Setpoints menu or directly via the normal display, and special controller functions are described.
- Turning on the chamber is described in chapter 6.1; controller settings upon start up in chapter 6.2.
- Set-point entry through the Setpoints menu is described in chapter 7.1.
- Direct setpoint entry via the normal display is described in chapter 7.2.
- Special controller functions are described in chapter 7.3.
Describes the timer stopwatch function and the creation, editing and execution of time programs. It covers program delay time, cancelling a running program, behavior after program end, program management, section management, section duration, set-point ramps and steps, tolerance range, and repetition of sections.
- Timer program start, cancellation and end behavior are described in chapters 8.1 to 8.3.
- Time program start, cancellation and end behavior are described in chapters 9.1 to 9.3.
- Program editor and section editor are described in chapters 9.5 and 9.6.
- Section value entry includes section duration, set-point ramp/step, tolerance range and repetition (chapters 9.7.1 to 9.7.7).
Describes starting, cancelling and creating week programs, program and section management, and value entry for program sections including set-point ramp/step modes, weekday, start time, setpoint and special controller functions.
- Starting and cancelling week programs are described in chapters 10.1 and 10.2.
- Creating a new week program is described in chapter 10.3.
- Program editor and section editor are described in chapters 10.4 and 10.5.
- Value entry includes weekday, start time, setpoint and special controller functions (chapters 10.6.1 to 10.6.5).
Provides an overview of notification and alarm messages, the state of alarm, resetting an alarm and the list of active alarms, and activation/deactivation of the audible alarm buzzer.
- Notification and alarm messages overview is given in chapter 11.1.
- Alarm messages are listed in chapter 11.1.2.
- State of alarm and alarm reset are described in chapters 11.2 and 11.3.
- Audible alarm activation/deactivation is described in chapter 11.4.
Describes the over temperature protective device class 1, the overtemperature safety controller class 2 including safety controller modes, setting, alarm-state messages and measures, and function check. An optional over/under temperature safety device class 2 is also covered.
- Over temperature protective device class 1 is described in chapter 12.1.
- Overtemperature safety controller class 2 modes and setting are described in chapters 12.2.1 and 12.2.2.
- Messages and measures in the state of alarm are described in chapter 12.2.3.
- Function check is described in chapter 12.2.4; optional over/under temperature safety device class 2 in chapter 12.3.
Covers authorization levels and password protection, login/logout, user change, password assignment and change, deletion of passwords, new password assignment when the password function was deactivated, and activation codes.
- Authorization levels and password protection are described in chapter 13.1.
- Login, logout and user change are described in chapters 13.2 to 13.4.
- Password assignment and change are described in chapter 13.5.
- Activation code handling is described in chapter 13.6.
Describes controller menu language, date and time, temperature unit, display configuration and touchscreen calibration, network and communication settings (serial interfaces, Ethernet, e-mail), USB data transfer, and automatic interior lighting turn-off.
- Menu language, date/time and temperature unit are described in chapters 14.1 to 14.3.
- Display configuration and touchscreen calibration are described in chapters 14.4.1 and 14.4.2.
- Network and communication settings include serial interfaces, Ethernet configuration, MAC address display and e-mail (chapters 14.5.1 to 14.5.3).
- USB data transfer and automatic interior lighting turn-off are described in chapters 14.6 and 14.7.
Covers the service contact page, current operating parameters, event list, technical chamber information, and the self-test function for MK 56.
- Service contact page is described in chapter 15.1.
- Current operating parameters and event list are described in chapters 15.2 and 15.3.
- Technical chamber information is described in chapter 15.4.
- Self-test function is available for MK 56 (chapter 15.5).
Describes chart recorder views, showing and hiding the legend, switching legend pages, showing and hiding specific indications, history display, and setting chart recorder parameters.
- Views and legend handling are described in chapters 16.1.1 to 16.1.3.
- History display is described in chapter 16.1.4.
- Setting the parameters is described in chapter 16.2.
Provides notes on refrigerating operation and describes anti-condensation protection for the chamber.
- Notes on refrigerating operation are given in chapter 17.
- Anti-condensation protection is described in chapter 18.
Describes the zero-voltage switching outputs provided by the chamber controller.
- Zero-voltage switching outputs are described in chapter 19.
Describes available options and accessories: APT-COM™ 4 Multi Management Software including Basic Edition and registration, RS485 interface, analog temperature outputs, flexible Pt 100 object temperature sensor, compressed air connection, compressed air dryer, and water cooling for MK 56 and MK 1020.
- APT-COM™ 4 software and Basic Edition registration are described in chapters 20.1 and 20.1.1.
- RS485 interface and analog temperature outputs are described in chapters 20.2 and 20.3.
- Object temperature display with flexible Pt 100 sensor is described in chapter 20.4.
- Compressed air connection, compressed air dryer and water cooling are described in chapters 20.5 to 20.7.
Covers cleaning and decontamination/chemical disinfection of the chamber, including a caution against eye contact with disinfectant spray and requirements to allow the chamber to dry and aerate after use.
- Cleaning is described in chapter 21.1.
- Decontamination/chemical disinfection is described in chapter 21.2.
- Do not empty disinfectant into drains; wear protective goggles.
- After using disinfectant spray, allow the chamber to dry thoroughly and aerate sufficiently.
Describes maintenance intervals and service, simple troubleshooting by operating personnel, detailed troubleshooting by BINDER Service, repair requirements, and electrical testing. It includes a fault table with fault description, possible cause and required measures, and a maintenance message that appears after 8760 operating hours or two years.
- Simple troubleshooting does not require technical intervention or disassembly of chamber parts (chapter 22.3).
- Detailed troubleshooting and repair must be performed by BINDER Service or BINDER-qualified service partners/technicians per the Service Manual.
- Annual repeat electrical inspection and testing before initial startup and after maintenance/repair are required; EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended.
- After 8760 operating hours or two years a maintenance message appears; after confirmation it reappears every two weeks until reset by BINDER Service.
Covers factory calibration and adjustment, over-current protection, definition of usable volume, MK and MKT technical data, equipment and options, accessories and spare parts, heating-up and cooling-down graphs, heat compensation, and dimensions.
- Factory calibration and adjustment are described in chapter 24.1.
- Over-current protection and usable volume definition are described in chapters 24.2 and 24.3.
- MK and MKT technical data are given in chapters 24.4 and 24.5.
- Equipment/options, accessories/spare parts, performance graphs, heat compensation and dimensions are covered in chapters 24.6 to 24.12.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Type plate example for MKT 240 shows nominal temperature 180 °C (356 °F), nominal power 6.50 kW, nominal voltage 400 V / 50 Hz ±10%, current type 3 N ~, nominal current 11.3 A, IP protection 20, temperature safety device DIN 12880 class 2.0, max operating pressure 29 bar, refrigerant stage 1 R452a 2.20 kg and stage 2 R23 0.38 kg. p.11 ↗ | Electrical supply, over-current protection and refrigerant handling must be sized for these values; the chamber requires a 400 V three-phase supply in this example. | MKT 240 (example type plate) |
| The chamber must not be installed in unventilated recesses; sufficient ventilation for heat dispersal and the prescribed minimum distances (chapter 3.4) must be observed. p.12 ↗ | Installation location must allow heat dissipation to avoid overheating damage to the chamber. | Whole series |
| The chamber must not be operated in potentially explosive areas and no combustible dust or air-solvent mixtures may be present near the chamber; the chamber has no explosion protection measures. p.13 ↗ | Site classification and charging-material selection must exclude explosive atmospheres and flammable substances at working temperature. | Whole series |
| The chamber must not become wet during operation, cleaning or maintenance and must not be installed in damp areas or puddles. p.13 ↗ | Water ingress creates a deadly electric shock hazard; installation and cleaning procedures must keep electrical parts dry. | Whole series |
| Water cooling option is available only for MK 56 and MK 1020, with separate cooling water inlet and outlet connections. p.28 ↗ | Water cooling plumbing must be planned only for these two model sizes. | MK 56 and MK 1020 |
| Chambers with the -C suffix are equipped with a voltage and frequency changer, which has its own installation and electrical connection instructions. p.29 ↗ | For -C variants, the voltage and frequency changer installation and connection procedure in chapters 4.3 and 4.4.2 must be followed in addition to the chamber connection. | MK115-400V-C, MK240-400V-C, MK400-400V-C, MK720-400V-C, MK1020-400V-C, MKT115-400V-C, MKT240-400V-C, MKT720-400V-C |
| MK 56 has a main power switch (1); MK/MKT from size 115 on have a rear power switch (12) in addition. p.22 ↗ | Troubleshooting a dead chamber must include checking the correct power switch for the model size. | MK 56 vs. MK/MKT from size 115 on |
| After 8760 operating hours or two years, a maintenance message appears; after confirmation with the Confirm icon, the message window reappears every two weeks until reset by BINDER Service. p.120 ↗ | Plan service intervention at or before 8760 h / two years to reset the maintenance counter and avoid recurring messages. | Whole series |
| An annual repeat electrical inspection is required, as well as a test prior to initial startup and prior to resuming operation after maintenance or repair; testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. p.118 ↗ | Electrical safety testing must be scheduled annually and after any maintenance or repair before the chamber is returned to service. | Whole series |
| Simple troubleshooting by operating personnel does not require technical intervention into the chamber or disassembly of chamber parts; detailed troubleshooting and repair must be performed by BINDER Service or BINDER-qualified service partners/technicians per the Service Manual. p.118 ↗ | On-site staff may only perform the checks in the simple troubleshooting table; deeper faults require BINDER service. | Whole series |
| After maintenance, the chamber must be tested prior to resuming operation. p.118 ↗ | A post-maintenance test is mandatory before the chamber is used again. | Whole series |
| The chamber is equipped with an Ethernet serial interface and can be networked with the BINDER APT-COM™ 4 Multi Management Software (accessory). p.19 ↗ | Network configuration (chapter 14.5) and APT-COM™ 4 registration (chapter 20.1.1.1) are required for networked monitoring and documentation. | Whole series |
| Available options include RS485 interface, analog outputs for temperature, object temperature display with flexible Pt 100 temperature sensor, compressed air connection, compressed air dryer, and water cooling (MK 56 and MK 1020). p.111 ↗ | Option selection must be fixed at ordering because several options are model-size dependent and require specific connections. | Whole series; water cooling only MK 56 and MK 1020 |
| The over temperature protective device class 1 responds if the nominal temperature is exceeded by 20 °C due to chamber failure. p.76 ↗ | A class 1 trip indicates a chamber failure condition and requires the chamber to be taken out of operation and BINDER Service informed. | Whole series |
| The chamber has a temperature safety device according to DIN 12880, class 2.0, as shown on the type plate. p.11 ↗ | The safety device classification must be considered when assessing chamber safety for the intended application. | MKT 240 example type plate; whole series per type plate indication |
| The chamber must only be installed, tested and started up by qualified personnel; operation is restricted to laboratory personnel trained for this purpose and familiar with laboratory precautions. p.6 ↗ | Personnel qualification must be verified before installation, commissioning and operation. | Whole series |
| The chamber must not be installed or operated in hazardous locations. p.12 ↗ | Site selection must exclude hazardous locations. | Whole series |
| The chamber must not be operated with a damaged housing or damaged power cord (listed under foreseeable misuse). p.16 ↗ | Visual inspection of housing and power cord is required before operation; damaged units must be taken out of service. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber without function. p.120 ↗ | No power supply. | Check connection to power supply. Check whether the chamber is turned on at the main power switch. | Chapter 22.3 |
| Chamber without function. p.120 ↗ | Wrong voltage. | Check power supply for correct voltage (chap. 4.4). | Chapter 22.3 |
| Chamber without function. p.120 ↗ | Main power switch (1) not turned on. | Turn on the main power switch (1). | Chapter 22.3 |
| Chamber without function. p.120 ↗ | MK/MKT from size 115 on: Rear power switch (12) not turned on. | Turn on the rear power switch (12). | Chapter 22.3 |
| Chamber without function. p.120 ↗ | Chamber fuse has responded. | Check chamber fuse and replace it if appropriate. If it responds again, contact BINDER service. | Chapter 22.3 |
| Chamber without function. p.120 ↗ | Controller defective. | Contact BINDER service. | Chapter 22.3 |
| Nominal temperature exceeded by 20 °C due to chamber failure. Over temperature protective device (class 1) responds. p.120 ↗ | Chamber failure. | Take the chamber out of operation and inform BINDER Service. | Chapter 22.3 |