BINDER BF 260
BINDER BF 260 — Incubators with Temperature 7…100 °C.
The BINDER BF 260 is a forced-convection incubator with a 257 L interior, adjustable fan speed in 10% steps, and a working range of +7 to 100 °C. It sits between the smaller BF 115 and larger BF 400, offering 40 kg per rack and 270 kg total load for high-density culture work. Its primary role is precise incubation at 37 °C with 0.1 K display resolution and 0.2 K uniformity.
Direct drop-in replacements
BINDER BF Series — Model Comparison & Selection
Full specifications
| Temperature | 7...100 °C |
| Heating-up time to 100 °C | 30 min |
| Interior volume | 257 L (BF260-230V / BF260UL-120V) |
| Load per rack | 40 kg (BF260-230V / BF260UL-120V) |
| Nominal power | 0.9 kW |
| Permitted load | 270 kg (BF260-230V / BF260UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 160 mm (BF260-230V / BF260UL-120V) |
| Side wall clearance | 100 mm (BF260-230V / BF260UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 37°C mfr | 3 min |
| Height | 845 mm (BF260-230V / BF260UL-120V) |
| Width | 810 mm (BF260-230V / BF260UL-120V) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 8.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The BF 260 uses forced convection with a variable-speed fan, distinguishing it from natural-convection BD models and fixed-speed FD ovens. Within the BF series, it is the mid-size 257 L unit; the BF 115 offers 115 L and the BF 400 provides 400 L, while the BF 56 and BF 720 bracket the range at 56 L and 720 L respectively.
- The interior is stainless steel with a glass door and removable racks rated at 40 kg each; total permitted load is 270 kg. Net width is 810 mm and height 845 mm, requiring 100 mm side and 160 mm rear wall clearance.
Behavior under load and in operation
- Temperature control spans +7 to 100 °C with 0.1 K fluctuation and 0.2 K uniformity at 37 °C. Recovery after a 30 s door opening at 37 °C is 3 minutes, supported by 0.9 kW heating power and a 30-minute heat-up to 100 °C.
- Fan speed is adjustable in 10% increments, allowing airflow tuning to balance temperature homogeneity against sample desiccation. The safety controller set-point range is 10–100 °C with 0.1 K accuracy, and must be set 2–5 °C above the working set-point.
Installation, control and operation
- The R4 controller manages temperature, fan speed, timer functions (Delayed On/Off, temperature-dependent Delayed Off), ramp function, and USB data recording. Ethernet is optional for BF models.
- Electrical supply is 230 V 1~ 50/60 Hz with 0.9 kW nominal power and an external 8.0 A time-lag fuse. The chamber uses an IEC connector plug with shockproof plug; IP 20 protection applies.
- Loading must respect 10% wall clearance and not exceed half the usable volume to maintain airflow and uniformity.
Integration & Operating Notes
The BF 260 requires a 230 V 1~ 50/60 Hz supply with fluctuations not exceeding ±10%. Verify the type plate and supply quality before connection; wrong voltage is a listed cause of malfunction. The external chamber fuse is 8.0 A time-lag (T) 5x20 mm, 250 V, and must be sized accordingly.
The chamber must be placed on a non-flammable surface in a ventilated indoor room, not in unventilated recesses. Maintain at least 100 mm sidewise and 160 mm rear wall clearance. Ambient temperature must be +18 to +40 °C and humidity ≤70% r.h. non-condensing; installation altitude max. 2000 m.
Samples must remain inside the usable volume with at least 10% of interior dimensions as wall clearance, and the usable volume must not be loaded by more than half. Overloading or placing samples too close to walls prevents sufficient airflow and homogeneous temperature distribution, compromising uniformity and recovery time.
The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized experts; warranty is void if maintenance is performed by non-authorized personnel. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, plus a test before initial startup and after any maintenance or repair.
The door gasket must be replaced only when the chamber is cold. Replacing it while hot may damage the gasket. Use BINDER-recommended neutral cleaning agent (art. no. 1002-0016) for routine cleaning and disinfectant spray (art. no. 1002-0022) for chemical disinfection; do not use neutral cleaner on zinc-coated hinge parts or rear wall.
Selection & Integration Q&A
Choose the BF 260 when you need 257 L capacity and 270 kg total load for mid-scale culture work. The BF 115 offers only 115 L, while the BF 400 provides 400 L but requires more bench space. The BF 260 balances capacity with a 0.9 kW heating power and 30 min heat-up to 100 °C.
Annual maintenance by authorized personnel and electrical safety testing per EN 50678/VDE 0701 are required. Recommended cleaning agents (art. no. 1002-0016, 1002-0022) are not included. Optional Ethernet interface, APT-COM 4 software, Pt100 sensor, and analog output are available separately. Racks are included but additional racks can be ordered.