BINDER BF 260
BINDER ·Made in Germany · In Active Production

BINDER BF 260

BINDER BF 260 — Incubators with Temperature 7…100 °C.

The BINDER BF 260 is a forced-convection incubator with a 257 L interior, adjustable fan speed in 10% steps, and a working range of +7 to 100 °C. It sits between the smaller BF 115 and larger BF 400, offering 40 kg per rack and 270 kg total load for high-density culture work. Its primary role is precise incubation at 37 °C with 0.1 K display resolution and 0.2 K uniformity.

Temperature
7…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BF Series — Model Comparison & Selection

Model Temperature, °CHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kgPermitted load, kg
BF 260 Current 7–1003084.525740270
BF 115 8–1006572.511430150
BF 400 5–10012102.54003590
BF 56 7–1003562.5593070
BF 720 10–1004515973445315

Full specifications

Core Performance & Functional Specifications
Temperature 7...100 °C
Heating-up time to 100 °C 30 min
Interior volume 257 L (BF260-230V / BF260UL-120V)
Load per rack 40 kg (BF260-230V / BF260UL-120V)
Nominal power 0.9 kW
Permitted load 270 kg (BF260-230V / BF260UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 160 mm (BF260-230V / BF260UL-120V)
Side wall clearance 100 mm (BF260-230V / BF260UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Height 845 mm (BF260-230V / BF260UL-120V)
Width 810 mm (BF260-230V / BF260UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 8.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must stay within ±10% of 230 V nominal

The BF 260 requires a 230 V 1~ 50/60 Hz supply with fluctuations not exceeding ±10%. Verify the type plate and supply quality before connection; wrong voltage is a listed cause of malfunction. The external chamber fuse is 8.0 A time-lag (T) 5x20 mm, 250 V, and must be sized accordingly.

Operating Manual 07/2025, Electrical connection data, p.18
Mounting
Install indoors with 100 mm side and 160 mm rear clearance

The chamber must be placed on a non-flammable surface in a ventilated indoor room, not in unventilated recesses. Maintain at least 100 mm sidewise and 160 mm rear wall clearance. Ambient temperature must be +18 to +40 °C and humidity ≤70% r.h. non-condensing; installation altitude max. 2000 m.

Operating Manual 07/2025, Installation location, p.20; Quick Start Guide, p.18
Process Limits
Load usable volume to max. half and keep 10% wall clearance

Samples must remain inside the usable volume with at least 10% of interior dimensions as wall clearance, and the usable volume must not be loaded by more than half. Overloading or placing samples too close to walls prevents sufficient airflow and homogeneous temperature distribution, compromising uniformity and recovery time.

Operating Manual 07/2025, Placing the charging material, p.19
Maintenance
Annual maintenance and electrical safety test required

The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized experts; warranty is void if maintenance is performed by non-authorized personnel. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, plus a test before initial startup and after any maintenance or repair.

Operating Manual 07/2025, Maintenance intervals, p.69
Maintenance
Replace door gasket only when cold

The door gasket must be replaced only when the chamber is cold. Replacing it while hot may damage the gasket. Use BINDER-recommended neutral cleaning agent (art. no. 1002-0016) for routine cleaning and disinfectant spray (art. no. 1002-0022) for chemical disinfection; do not use neutral cleaner on zinc-coated hinge parts or rear wall.

Operating Manual 07/2025, Maintenance intervals, p.69; Cleaning, p.23

Selection & Integration Q&A

When should I choose the BF 260 over the BF 115 or BF 400?

Choose the BF 260 when you need 257 L capacity and 270 kg total load for mid-scale culture work. The BF 115 offers only 115 L, while the BF 400 provides 400 L but requires more bench space. The BF 260 balances capacity with a 0.9 kW heating power and 30 min heat-up to 100 °C.

What maintenance and accessories are needed beyond the base delivery?

Annual maintenance by authorized personnel and electrical safety testing per EN 50678/VDE 0701 are required. Recommended cleaning agents (art. no. 1002-0016, 1002-0022) are not included. Optional Ethernet interface, APT-COM 4 software, Pt100 sensor, and analog output are available separately. Racks are included but additional racks can be ordered.