BINDER FED 56
BINDER FED 56 — Drying ovens with Temperature 10…300 °C.
The BINDER FED 56 is a forced-convection drying and heating oven in the Avantgarde.Line, with a 60 L interior, +10 to 300 °C range, and 1.1 kW single-phase 230 V supply. Its 80 air changes per hour and 15 min heat-up to 150 °C distinguish it from natural-convection ED siblings, while the 56 size is the smallest of the FED series. It is intended for exact tempering, drying, and heat treatment of non-flammable solids and bulk materials in laboratory and production environments.
BINDER FED Series — Model Comparison & Selection
Full specifications
| Temperature | 10...300 °C |
| Air change rate | 80 x/h |
| Heating time to 150°C | 15 min |
| Interior volume | 60 L |
| Load per rack | 30 kg (FED056-230V / FED056UL-120V) |
| Nominal power | 1.1 kW |
| Permitted load | 70 kg (FED056-230V / FED056UL-120V) |
| Temperature fluctuation | 0.3 K |
| Temperature uniformity | 1.4 K |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 4 min |
| Depth | 565 mm (FED056-230V / FED056UL-120V) |
| Height | 625 mm (FED056-230V / FED056UL-120V) |
| Width | 560 mm (FED056-230V / FED056UL-120V) |
| Net weight of the unit (empty) mfr | 39 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The FED 56 uses forced convection (fan-assisted airflow) with an adjustable fan speed in 10 % steps, unlike the ED series (natural convection) and the FD series (fixed fan speed).
- Within the FED series, the 56 is the smallest chamber (60 L); siblings FED 115, 260, 400, and 720 offer larger volumes and, for the 720, a three-phase 400 V supply and two-door configuration.
- The interior is stainless steel; the unit has a triangular instrument panel with the R4 controller, an outer door, and a glass door is not listed for FED (glass door is on BD/BF incubators).
- Ethernet interface is standard on FED models, enabling network configuration and data logging; USB interface is also provided.
Behavior under load and in different modes
- Temperature uniformity is 1.4 K and fluctuation 0.3 K, supported by the 80 x/h air-change rate and adjustable fan speed.
- Heat-up time to 150 °C is 15 min; recovery after a 30 s door opening at 150 °C is 4 min.
- The safety controller (class 2 or 3.1) set-point range is 10 °C to 300 °C with 1 degree accuracy; it should be set 2–5 °C above the working set-point.
- The overtemperature protective device class 1 trips at 330 °C for FED 56, limiting maximum temperature and defining the cooling threshold for reset.
- Loading must leave at least 10 % of interior dimensions as wall clearance and must not exceed half the usable volume; maximum load per rack is 30 kg, total permitted load 70 kg.
Installation, connection, and operation
- Electrical supply: 230 V 1~ 50 Hz, nominal power 1.1 kW, external chamber fuse 6.3 A time-lag (T) 5x20 mm; the unit is supplied with an IEC connector plug and shockproof plug.
- Permissible ambient temperature is +18 to +40 °C, max 70 % r.h. non-condensing, max installation altitude 2000 m; supply voltage tolerance ±10 %.
- The R4 controller supports timer functions (Delayed Off, Temperature-dependent Delayed Off, Delayed On), ramp function, and data recording via USB.
- Cleaning: neutral cleaning agent (art. no. 1002-0016) or disinfectant spray (art. no. 1002-0022); hot-air sterilization at 190 °C for at least 30 min is possible for FED.
- Annual maintenance by licensed electricians or BINDER-authorized experts is required; electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended.
Integration & Operating Notes
The FED 56 is rated for 230 V 1~ 50 Hz with a nominal power of 1.1 kW and an external 6.3 A time-lag fuse. Voltage fluctuations beyond ±10 % can cause malfunction or damage; verify the supply quality and fuse rating before connection. The chamber uses an IEC connector plug with shockproof plug, so a compatible outlet and upstream circuit protection must be provided.
The chamber must be placed on a non-flammable surface in a well-ventilated indoor area, not in unventilated recesses or explosive atmospheres. Ambient temperature must remain between +18 °C and +40 °C, humidity ≤70 % r.h. non-condensing, and installation altitude ≤2000 m. Observe minimum wall distances to ensure airflow and access.
The safety controller acts as an absolute maximum temperature limit. After any change to the temperature set-point, re-adjust the safety controller set-point to approximately 2 °C to 5 °C above it. The permissible range is 10 °C to 300 °C with 1 degree accuracy. Failure to set it correctly may cause unnecessary tripping or inadequate protection.
The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized experts. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, and a test is required before initial startup and after any maintenance or repair. Warranty may be void if maintenance is performed by non-authorized personnel.
The chamber has no explosion protection. Do not introduce combustible or explosive substances, dust, or air-solvent mixtures. The operating temperature must remain below the flash point or sublimation point of the material. Corrosive ingredients such as acids and halides that may damage stainless steel, aluminum, or copper are also prohibited.
Selection & Integration Q&A
Choose FED 56 for small loads up to 60 L requiring forced convection and adjustable fan speed for uniform drying. For larger volumes, select FED 115/260/400/720. If natural convection is sufficient and fan adjustability is not needed, the ED series is an alternative, but FED offers enhanced timer functions and standard Ethernet.
Annual maintenance by authorized personnel and electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Base delivery excludes accessories such as APT-COM 4 software, Pt100 sensor, analog output, HEPA filter, and inert gas connection. Cleaning agents (art. 1002-0016, 1002-0022) must be ordered separately.