BINDER KB 1060 — Incubators with Temperature 0…70 °C.

The BINDER KB 1060 is a 1,070 L cooling incubator with a 0–70 °C working range, temperature uniformity of 0.2 K and fluctuation of 0.1 K. As the largest KB series model, it offers 640 W heat compensation and 200 kg permitted load, distinguishing it from smaller siblings. It is used for temperature-controlled storage and incubation of large sample volumes in laboratory processes.

Temperature
0…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB 1060 Current 0–70117.5195.11070101140
KB 130 0–7071100.512765630
KB 260 0–7090139.626265640
KB 470 0–7090195.147265640
KB 65 0–707190.56340630
KB 720 0–7090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 1070 L
Load per rack 40 kg
Max. heat compensation 640 W
Nominal power 1.5 kW
Permitted load 200 kg (KB1060-230V / KB1060UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm (KB1060-230V / KB1060UL-120V)
Side wall clearance 180 mm (KB1060-230V / KB1060UL-120V)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 1175 mm (KB1060-230V / KB1060UL-120V)
Height 1951 mm (KB1060-230V / KB1060UL-120V)
Interior width 1011 mm (KB1060-230V / KB1060UL-120V)
Width 1285 mm (KB1060-230V / KB1060UL-120V)
Net weight of the unit (empty) mfr 315 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 220...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and positioning

Performance and load behavior

Installation and operation

Integration & Operating Notes

Mounting
Maintain 100 mm rear and 180 mm side clearance

Insufficient clearance restricts airflow and can cause overheating, leading to reduced performance or damage. Ensure the installation site allows these minimum distances from walls or other equipment.

Factory Specification Sheet
Electrical
Use a 16 A branch circuit protection

The unit requires a 16 A fuse to handle the 1.5 kW nominal power at 220–240 V. Verify that the supply can deliver the required current without nuisance tripping.

Factory Specification Sheet
Process Limits
Do not exceed 40 kg per rack

Exceeding the per-rack load limit may cause rack deformation or failure, potentially damaging samples or the interior. Distribute weight evenly and consider additional racks for heavy loads.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the KB 1060 over smaller KB series models like the KB 470?

Choose the KB 1060 when you need maximum capacity (1,070 L) and higher load capacity (200 kg total, 40 kg per rack). Smaller models are suitable for lower volume or when space is limited, but the KB 1060 offers the same temperature range and stability with greater throughput.

What ambient conditions and clearances are required for installation?

Install indoors with 100 mm rear and 180 mm side clearance for ventilation. The permissible ambient temperature range is not specified in the available documents; refer to the manual. Avoid unventilated recesses.

What maintenance and accessories are needed for the KB 1060?

Annual electrical safety inspection is recommended. No accessories are included in base delivery; racks and shelves must be ordered separately. The LED plant light module is an optional accessory for plant growth applications.