BINDER KB 260
BINDER ·Made in Germany · In Active Production

BINDER KB 260

BINDER KB 260 — Incubators with Temperature 0…70 °C.

The BINDER KB 260 is a 262 L cooling incubator with a 0–70 °C working range, 0.1 K fluctuation and 0.2 K uniformity, housed in a 930 mm wide cabinet. It is the mid-series 260 L modification, distinguished from smaller KB 65/130 and larger KB 470/1060 siblings by its 656 mm interior width and 200 kg permitted load. It serves as a temperature-controlled culture and storage chamber in laboratory workflows requiring stable, uniform conditions.

Temperature
0…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB 260 Current 0–7090139.626265640
KB 1060 0–70117.5195.11070101140
KB 130 0–7071100.512765630
KB 470 0–7090195.147265640
KB 65 0–707190.56340630
KB 720 0–7090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 262 L
Load per rack 40 kg
Max. heat compensation 410 W
Nominal power 1.3 kW
Permitted load 200 kg (KB260-230V / KB260UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm (KB260-230V / KB260UL-120V)
Side wall clearance 180 mm (KB260-230V / KB260UL-120V)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 900 mm (KB260-230V / KB260UL-120V)
Height 1396 mm (KB260-230V / KB260UL-120V)
Interior width 656 mm (KB260-230V / KB260UL-120V)
Width 930 mm (KB260-230V / KB260UL-120V)
Net weight of the unit (empty) mfr 174 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 220...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior

Installation and operation

Integration & Operating Notes

Mounting
Keep 100 mm rear and 180 mm side clearance

The cabinet dissipates heat through its surfaces; violating the prescribed rear and side distances or installing it in an unventilated recess prevents heat dispersal and can cause overheating damage. Plan the bench or floor layout around the 930 mm width plus 180 mm on each side.

Data Sheet Model KB 260 en.pdf; Factory Specification Sheet
Electrical
Provide a 16 A branch circuit for the 1.3 kW load

The unit is protected internally by a 16.0 A fuse and draws 1.3 kW nominal at 220–240 V single-phase. The upstream circuit breaker and wiring must be sized for this continuous load; do not share the circuit with other high-draw equipment on the same phase.

Factory Specification Sheet
Process Limits
Active loads must stay within 410 W heat compensation

Any heat-generating load inside the chamber — including accessory lighting such as the LED plant light module, which contributes about 220 W at full output — counts against the 410 W maximum heat compensation. Exceeding it prevents the chamber from holding setpoint at the low end of the 0–70 °C range.

Factory Specification Sheet; BINDER LED Pflanzenlichtmodul (E7) 09-2025_en.pdf, Device description
Process Limits
Respect 40 kg per rack and 200 kg total load

Rack loading is limited to 40 kg each and the total permitted load is 200 kg. Distribute heavy items across multiple racks rather than concentrating mass on one shelf, and account for the weight of any accessory racks or light bars added to the chamber.

Factory Specification Sheet
Maintenance
Schedule annual electrical safety inspection

An electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required annually and after any maintenance or repair. Maintenance itself must be performed at least once a year by licensed electricians or BINDER-authorized experts; work by unauthorized personnel voids the warranty.

BINDER LED Pflanzenlichtmodul (E7) 09-2025_en.pdf, Maintenance intervals
Environment
Install indoors within 18–32 °C ambient

The chamber and any attached accessories are for indoor use only and must operate in an ambient temperature of +18 °C to +32 °C. After transport or storage at temperatures down to −10 °C, allow at least one hour at the installation site before start-up to avoid condensation.

BINDER LED Pflanzenlichtmodul (E7) 09-2025_en.pdf, Location of installation; Transport and storage

Selection & Integration Q&A

When should I choose the KB 260 rather than the KB 130 or KB 470?

All KB siblings share the 0–70 °C range, so select on volume and load: the KB 260 offers 262 L and 200 kg permitted load with a 656 mm interior width. Choose the KB 130 for smaller batches or the KB 470 when 262 L is insufficient; the 260 balances bench footprint (930 mm wide) against capacity.

What supply and protective device does the KB 260 require, and which plug variant should I order?

The 230 V variant (KB260-230V, Art.-No. 9020-0473) runs on 220–240 V, 1~, 50 Hz at 1.3 kW with a CEE 7/7 plug and a 16.0 A unit fuse; size the branch circuit accordingly. For 120 V sites order the KB260UL-120V (Art.-No. 9020-0501) with NEMA 5-20 plug.

Can I place heat-generating equipment or a light module inside the KB 260?

Yes, within the 410 W maximum heat compensation. The BINDER LED plant light module, for example, adds about 220 W at full output, leaving roughly 190 W of margin; beyond 410 W the chamber cannot hold setpoint at the low end of its 0–70 °C range.

What maintenance and consumables are needed beyond the base delivery?

Plan an annual electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 plus yearly maintenance by BINDER-authorized personnel. If the LED plant light accessory is used, its CR1225 controller battery and LED bars (replace after ~25,000 h or 6 years) are wear items not included with the base KB 260.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB 260 en
Datasheet ·589,3 КБ ·7 pp.
View PDF
PDF
BINDER LED Pflanzenlichtmodul (E7) 09 2025 en
Manual ·5,5 МБ ·55 pp.
View PDF