BINDER KB PRO 1060
BINDER ·Made in Germany · In Active Production

BINDER KB PRO 1060

BINDER KB PRO 1060 — Incubators with Temperature -20…100 °C.

The KB PRO 1060 is a 1,070 L single-phase cooling incubator with a -20…100 °C working range and 1.8 kW nominal power. It is the second-largest chamber in the KB PRO series, distinguished by a 200 kg permitted load and 40 kg per rack, intended for high-volume temperature-stability processes at 0.1 K fluctuation and 0.2 K uniformity.

Temperature
-20…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB PRO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB PRO 1060 Current -20–100117.5195.11070101140
KB PRO 130 -20–10071100.512765630
KB PRO 1600 -20–100117.5195.11610152140
KB PRO 260 -20–10090139.626265640
KB PRO 470 -20–10090195.147265640
KB PRO 720 -20–10090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 1070 L
Load per rack 40 kg
Max. heat compensation 1000 W
Nominal power 1.8 kW
Permitted load 200 kg
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm
Side wall clearance 180 mm
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 1175 mm
Height 1951 mm
Interior width 1011 mm
Width 1285 mm
Net weight of the unit (empty) mfr 353 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in different modes

Installation, connection, and operation

Integration & Operating Notes

Mounting
Maintain 100 mm rear and 180 mm side clearance

The chamber requires a wall spacer from size 260 upward. Insufficient rear or side clearance prevents heat dispersal and can cause overheating; do not install in unventilated recesses. Verify clearance before positioning the 1,285 mm wide unit.

Instruction Manual KB PRO (E7), Section 4.1 and Section 1.8
Electrical
Provide a 208–240 V single-phase branch rated for 16 A

The unit draws 1.8 kW nominal at 208…240 V, 1~, 50 Hz and is protected by a 16.0 A fuse. Confirm supply voltage and frequency before connection; use a CEE 7/7 plug compatible outlet. Undersized branch protection will trip during heat-up.

Data Sheet Model KB PRO 1060; Factory Specification Sheet
Process Limits
Respect 200 kg total and 40 kg per-rack load limits

The permitted load is 200 kg for the chamber and 40 kg per rack. Exceeding these limits risks rack deformation and invalidates temperature uniformity. Distribute loads evenly and use U-rails or telescopic rails rated for the application.

Factory Specification Sheet; Instruction Manual KB PRO (E7), Section 4.2
Control & I/O
Configure safety controller class before running risky processes

The temperature safety device can be set as class 2 (temperature limiter) or class 3.3 (temperature protection), with optional class 3.2 undertemperature protection. Select the class and limit/offset according to process risk; an incorrect class may fail to protect the sample or trip unnecessarily.

Instruction Manual KB PRO (E7), Sections 13.1–13.3
Maintenance
Schedule annual maintenance and electrical safety testing

Regular maintenance is required at least once a year, and electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended before initial startup and after repair. A service message appears after 8,760 operating hours or two years and recurs every two weeks until reset by BINDER Service.

Instruction Manual KB PRO (E7), Sections 22.1–22.2
Cooling & Media
Clean condenser more often in dusty environments

The condenser sits behind the machine-room cover, which is removable without tools. In dusty ambient air, suction-clean the condenser several times a year to maintain refrigeration performance. Refrigeration work must be performed by personnel trained per EN 13313:2010.

Instruction Manual KB PRO (E7), Section 22.2

Selection & Integration Q&A

When should I choose the KB PRO 1060 instead of the KB PRO 720 or 1600?

Choose the 1060 when you need 1,070 L and a 200 kg total load with 40 kg per rack, between the 720 and 1600. The 1600 offers more volume but the same temperature range; the 720 is smaller. All share -20…100 °C and 0.1 K fluctuation.

What are the power supply and branch protection requirements for the KB PRO 1060?

It requires 208…240 V, 1~, 50 Hz with a 16.0 A unit fuse and 1.8 kW nominal power. Provide a dedicated branch rated for at least 16 A and verify voltage/frequency before connection; the CEE 7/7 plug must match the outlet.

What mounting clearances and environmental conditions must be met?

Maintain 100 mm rear and 180 mm side clearance; a wall spacer is required from size 260 upward. Do not install in unventilated recesses. Ambient dust increases condenser cleaning frequency. The unit weighs 353 kg empty, so floor loading must be verified.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance and electrical safety testing are required; a service message appears after 8,760 h or two years. Racks (U-rails or telescopic), APT-COM 4 software, Pt 100 sensor, alarm contacts, and door lock are options/accessories not included in base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB PRO 1060 en
Datasheet ·630,6 КБ ·8 pp.
View PDF
PDF
KB PRO (E7) 09 2025 en
Manual ·12,7 МБ ·143 pp.
View PDF