BINDER KB PRO 1600
BINDER ·Made in Germany · In Active Production

BINDER KB PRO 1600

BINDER KB PRO 1600 — Incubators with Temperature -20…100 °C.

The KB PRO 1600 is a floor-standing cooling incubator with a 1,610 L interior, the largest chamber in the KB PRO series, operating from -20 to 100 °C with 0.1 K fluctuation and 0.2 K uniformity. Its 1,000 W heat compensation and 40 kg per-rack load capacity support high-density incubation protocols. It serves as a central, high-throughput temperature-controlled environment in pharmaceutical and biotechnological process chains.

Temperature
-20…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB PRO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB PRO 1600 Current -20–100117.5195.11610152140
KB PRO 1060 -20–100117.5195.11070101140
KB PRO 130 -20–10071100.512765630
KB PRO 260 -20–10090139.626265640
KB PRO 470 -20–10090195.147265640
KB PRO 720 -20–10090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 1610 L
Load per rack 40 kg
Max. heat compensation 1000 W
Nominal power 1.8 kW
Permitted load 200 kg
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm
Side wall clearance 180 mm
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 1175 mm
Height 1951 mm
Interior width 1521 mm
Width 1795 mm
Net weight of the unit (empty) mfr 443 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection and control

Integration & Operating Notes

Electrical
Branch circuit must accommodate 1.8 kW at 16 A fuse rating

The unit draws 1.8 kW nominal on a single-phase 208…240 V, 50 Hz supply and is protected by a 16.0 A unit fuse. The upstream branch protection and wiring must be sized for this load and for the inrush behavior of the refrigeration compressor; verify voltage and frequency before connection.

Data Sheet Model KB PRO 1600 en.pdf; Factory Specification Sheet
Mounting
Maintain 100 mm rear and 180 mm side clearance for heat dispersal

The chamber must not be installed in unventilated recesses. A wall spacer is required from size 260 upward, and the prescribed minimum distances must be observed to prevent overheating and to allow condenser access. Plan the floor position so that the machine room cover can be removed without tools.

Instruction Manual KB PRO (E7), Section 1.8 and Section 4.1
Process Limits
Total load limited to 200 kg with 40 kg per rack maximum

Rack loading must not exceed 40 kg per rack, and the total permitted load on the chamber is 200 kg. Exceeding these limits risks deformation of the racks and impaired temperature uniformity. Racks with U-rails or telescopic rails are available as accessories and must be selected according to loading and access requirements.

Factory Specification Sheet; Instruction Manual KB PRO (E7), Section 4.2
Maintenance
Replace door gasket only when the chamber is cold

The door gasket must be replaced only when the chamber is at ambient/cold condition; replacing it while warm may damage the gasket. Electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended prior to initial startup and after maintenance or repair.

Instruction Manual KB PRO (E7), Sections 22.1–22.2

Selection & Integration Q&A

When should I choose the KB PRO 1600 over the smaller KB PRO 1060 or 720?

Choose the 1600 when the 1,610 L interior and 200 kg total load capacity are needed for high-throughput protocols; the 1060 and 720 offer the same -20…100 °C range but smaller volumes and lower load limits. The 1600 also provides 1,000 W heat compensation for heavy internal loads.

What electrical supply and branch protection does the KB PRO 1600 require?

Single-phase 208…240 V, 50 Hz, nominal power 1.8 kW, unit fuse 16.0 A. The upstream branch circuit and wiring must be sized for this load and for compressor inrush; verify voltage and frequency before connection.

What installation clearances and environmental conditions must be respected?

Maintain at least 100 mm rear and 180 mm side clearance; a wall spacer is required from size 260 upward. Do not install in unventilated recesses. In dusty ambient air, clean the condenser several times a year by suction.

What maintenance schedule and accessories are required beyond the base delivery?

Annual maintenance is required, with a BINDER service reset at 8760 operating hours or two years. Racks (U-rails or telescopic rails), APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, and Pt 100 sensor are options/accessories not included in the base unit.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB PRO 1600 en
Datasheet ·644,8 КБ ·7 pp.
View PDF
PDF
KB PRO (E7) 09 2025 en
Manual ·12,7 МБ ·143 pp.
View PDF