BINDER MK 240
BINDER ·Made in Germany · In Active Production

BINDER MK 240

BINDER MK 240 — Climate chambers with Temperature -40…180 °C.

The BINDER MK 240 is a 228 L alternating climate chamber (MK240-400V, Art. 9020-0376) for rapid temperature cycling from -40 to 180 °C, with average heating/cooling rates of 5.0/4.5 K/min per IEC 60068-3-5. It is the mid-size 400 V three-phase standard variant, distinguished from the CO₂-capable MK 240 CO₂ and the larger MK 1020 by its 228 L interior and 5.6 kW nominal power. It serves as a dynamic test station for thermal stress screening and material qualification in a laboratory infrastructure.

Temperature
-40…180 °C

BINDER MK Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior height, mmInterior volume, LInterior width, mm
MK 240 Current -40–18092.5171700228735
MK 1020 -40–180148200.510209911200
MK 1020 CO₂ -48–180148200.510209911200
MK 115 -40–18086.5172.5480115600
MK 240 CO₂ -50...18092.5171700228735
MK 400 -40–180140171700417735
MK 400 CO₂ -50...180140171700417735
MK 56 -40–18077.8144.542060400
MK 720 -40–180114200.510207341200
MK 720 CO₂ -48–180114200.510207341200

Full specifications

Core Performance & Functional Specifications
Temperature -40...180 °C
Interior volume 228 L
Load per rack 30 kg (MK240-400V)
Nominal power 5.6 kW
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 925 mm (MK240-400V)
Height 1710 mm (MK240-400V)
Interior height 700 mm (MK240-400V)
Interior width 735 mm (MK240-400V)
Width 1115 mm (MK240-400V)
Net weight of the unit (empty) mfr 321 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 3 N/PE
Operating Conditions, Environment & Protection
Cooling down time from 180°C to -40°C mfr 110 min
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.0 K/min

Description

Construction and series positioning

Thermal behavior and control

Installation and operation

Integration & Operating Notes

Electrical
400 V three-phase supply with 16 A fusing required

The MK240-400V draws 5.6 kW at 400 V 3 N/PE 50 Hz and is protected by a 16.0 A unit fuse. The site supply must be sized for this three-phase load and the chamber must not be connected to a single-phase or lower-voltage circuit; the -C variant is the only one that accepts 480 V 60 Hz via its converter.

Data Sheet Model MK 240 en.pdf; Factory Specification Sheet
Mounting
Do not install in unventilated recesses or damp areas

Sufficient ventilation for heat dispersal and the prescribed minimum distances must be observed; installation in unventilated recesses risks overheating. The chamber must not become wet and must not be installed in damp areas or puddles, because water ingress creates a deadly electric shock hazard.

Operating Manual MK/MKT (E5), Chapter 1.8, pages 12-13
Process Limits
No explosive or flammable atmospheres permitted at the chamber

The chamber has no explosion protection measures and must not be operated in potentially explosive areas or with combustible dust or air-solvent mixtures near it. Charging materials and site classification must exclude flammable substances at working temperature up to 180 °C.

Operating Manual MK/MKT (E5), Chapter 1.8, page 13
Control & I/O
Class 1 over-temperature trip indicates chamber failure

The over-temperature protective device class 1 responds if the nominal temperature is exceeded by 20 °C due to chamber failure. A class 1 trip must be treated as a failure condition: take the chamber out of operation and inform BINDER Service rather than simply resetting and resuming.

Operating Manual MK/MKT (E5), Chapter 12.1, page 76; Chapter 22.3, page 120
Maintenance
Plan service at 8760 h or two years to reset counter

After 8760 operating hours or two years a maintenance message appears; after confirmation it reappears every two weeks until reset by BINDER Service. Schedule the service intervention at or before this interval to avoid recurring messages and to keep the maintenance counter valid.

Operating Manual MK/MKT (E5), Chapter 22.2, page 120
Maintenance
Annual electrical safety test and post-repair test mandatory

An annual repeat electrical inspection is required, plus a test prior to initial startup and prior to resuming operation after maintenance or repair; testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. The chamber must not be returned to service without this test.

Operating Manual MK/MKT (E5), Chapter 22.1, page 118

Selection & Integration Q&A

When should I choose the MK 240 standard over the MK 240 CO₂ or the larger MK 1020?

Pick the MK 240 standard for -40...180 °C cycling in a 228 L interior at 400 V 50 Hz. Choose the MK 240 CO₂ only if you need the extended -50 °C low end; choose the MK 1020 when 228 L is insufficient and a larger interior is required.

What supply and network infrastructure does the MK240-400V need?

A 400 V 3 N/PE 50 Hz supply with 16.0 A fusing for the 5.6 kW load, plus an Ethernet connection for APT-COM™ 4 networking. RS485 and analog temperature outputs are options that must be ordered with the chamber.

Where can the chamber be installed and what environment is prohibited?

Install with sufficient ventilation and minimum distances, never in unventilated recesses, damp areas, puddles, or hazardous locations. Explosive atmospheres and combustible dust or air-solvent mixtures near the chamber are prohibited because it has no explosion protection.

What maintenance and accessories are not included in the base delivery?

The base unit includes the chamber and MB2 controller; APT-COM™ 4 software, RS485, analog outputs, Pt 100 object sensor, compressed air connection/dryer, and class 2 safety device are options. Service is due at 8760 h or two years, with annual electrical safety testing.

Dimensional Drawings & Technical Documentation

Dimensional drawings

MK 240 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model MK 240 en
Datasheet ·924,7 КБ ·8 pp.
View PDF
PDF
MK+MKT (E5) 10 2026 en
Manual ·15,9 МБ ·169 pp.