Peter Huber Kältemaschinenbau Minichiller 1200w-H Kompakter
Peter Huber Kältemaschinenbau Minichiller 1200w-H Kompakter — Recirculating chillers.
The Minichiller 1200w-H Kompakter is a water-cooled recirculating chiller with a plastic pump and Pilot ONE controller, covering -20 to 100 °C. It delivers 1.2 kW cooling at 15 °C, falling to 0.35 kW at -20 °C, and is distinguished from 1000-series siblings by its higher capacity. It serves as a compact cooling source for laboratory and pilot processes requiring precise temperature control.
Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Minichiller 1200w-H Kompakter Current | — | Pilot ONE | III / FL | -20–100 | — | — |
| Minichiller 1000 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 280 OLÉ Kompakter | 53 | OLÉ | I / NFL | -5–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300 OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w-H OLÉ Kompakter | 51 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 600-H OLÉ Kompakter | 56 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 600 Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 110-120V 1~ 60Hz |
| Minichiller 600 OLÉ Kompakter | 56 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 60Hz |
| Minichiller 600w OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
Full specifications
| Temperature range | -20...100 °C |
| Weight | 39 kg (net) |
| Controller | Pilot ONE |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The 1200w-H uses a water-cooled refrigeration circuit with R-290 (A3, H220) refrigerant and a plastic pump; the 'w' denotes water cooling, 'H' the heating function (safety class III/FL).
- Within the Minichiller series, the 1200w-H sits above the 1000w-H and 1000-H variants in cooling capacity; the 1000-series models are the smaller siblings, while the 1200w-H is selected when the load at the target temperature exceeds the 1000-series capability.
- The Pilot ONE controller provides the user interface; the unit is IP20 and intended for indoor, non-ATEX installation.
Performance and media behaviour
- Cooling capacity is strongly temperature-dependent: 1.2 kW at 15 °C, 0.9 kW at 0 °C, 0.7 kW at -10 °C and 0.35 kW at -20 °C (at +20 °C ambient). Selection must match the actual load at the intended set point.
- The speed-controlled pump allows either speed or pressure control to match the application; whisper mode reduces sound pressure to 51 dB(A) ±4 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.
- The plastic pump is not resistant to acetone or polyglycols; alternative media require a more resistant pump option.
Installation and operation
- Cooling water must be supplied at 5–20 °C with a differential pressure of at least 3 bar and no more than 6 bar; the circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) dictate water quality, and anti-corrosion agents are mandatory if salts such as chlorides or bromide are added.
- Electrical supply: 208–240 V 1~/2~ 50/60 Hz, max. current 12 A, 16 A fuse; voltage tolerance ±5% and frequency ±2%, but the combination of -5% voltage with +2% frequency is not allowed.
- Ambient temperature must be 5–40 °C; a drip tray for condensation water and thermal fluid must be installed below the unit, and electrical lines directly below must be protected against liquid ingress.
- Do not install valves or quick-release couplings in the feed/discharge lines to the application; the unit must not be moved while operating.
Integration & Operating Notes
Condensation water and thermal fluid can escape from the unit; the responsible body must install a drip tray below it, and any electrical lines directly underneath must be protected against liquid ingress to prevent shock hazard.
The facility water supply must provide a differential pressure of at least 3 bar and no more than 6 bar, with intake temperature between 5 °C and 20 °C. Insulate pipes if the water temperature can fall below the dew point.
The unit draws up to 12 A and requires a 16 A fuse. Voltage tolerance is ±5% and frequency ±2%, but the combination of -5% voltage with +2% frequency is not allowed; verify site supply quality against EN60034-1.
The plastic pump is not resistant to acetone or polyglycols; using these media will damage the pump. If such media are required, a more resistant pump option must be specified at purchase.
If chlorides, bromide or other salts are added to the cooling water circuit, anti-corrosion agents are mandatory to protect the copper, stainless steel and EPDM components from corrosion damage.
Whisper mode reduces sound pressure to 51 dB(A) ±4 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar; ensure the reduced flow and pressure are acceptable for the application before enabling it.
The unit must not be installed or started up within an ATEX zone due to explosion risk. Do not move the unit while it is operating because of burns, freezing and escaping thermal fluid.
Selection & Integration Q&A
Choose the 1200w-H when your cooling load at the target temperature exceeds the 1000-series capacity; it provides 1.2 kW at 15 °C and 0.35 kW at -20 °C. The 1000-series models are smaller and suit lower loads. The 'w' denotes water cooling, so a cooling water supply is required.
The unit needs cooling water at 5–20 °C with a differential pressure of 3–6 bar. Connections are G1/2 male for cooling water and M16x1 male for the pump; select hoses and fittings accordingly. The water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible.
Install indoors at 5–40 °C ambient, non-ATEX, with IP20 protection. A drip tray for condensation and thermal fluid is required below the unit, and electrical lines underneath must be protected from liquid. Do not install valves or quick-release couplings in the feed/discharge lines to the application.
The base unit does not include a drip tray, cooling water hoses or fittings, or anti-corrosion agents. If salts are added to the water circuit, anti-corrosion agents are mandatory. The minimum filling capacity is 2.8 l with 2.2 l expansion volume; ensure the external loop and expansion allowance are sized accordingly.